| Literature DB >> 23935701 |
Lin Li1, Ke Wang, Yan Yan, Xudong Song, Zhicheng Liu.
Abstract
The tilt and decentration of intraocular lens (IOL) result in defocussing, astigmatism, and wavefront aberration after operation. The objective is to give a method to estimate the tilt and decentration of IOL more accurately. Based on AS-OCT images of twelve eyes from eight cases with subluxation lens after operation, we fitted spherical equation to the data obtained from the images of the anterior and posterior surfaces of the IOL. By the established relationship between IOL tilt (decentration) and the scanned angle, at which a piece of AS-OCT image was taken by the instrument, the IOL tilt and decentration were calculated. IOL tilt angle and decentration of each subject were given. Moreover, the horizontal and vertical tilt was also obtained. Accordingly, the possible errors of IOL tilt and decentration existed in the method employed by AS-OCT instrument. Based on 6-12 pieces of AS-OCT images at different directions, the tilt angle and decentration values were shown, respectively. The method of the surface fitting to the IOL surface can accurately analyze the IOL's location, and six pieces of AS-OCT images at three pairs symmetrical directions are enough to get tilt angle and decentration value of IOL more precisely.Entities:
Mesh:
Year: 2013 PMID: 23935701 PMCID: PMC3725812 DOI: 10.1155/2013/572530
Source DB: PubMed Journal: Comput Math Methods Med ISSN: 1748-670X Impact factor: 2.238
Figure 1The AS-OCT image at 30 degrees scanned angle. Points A and B are scleral spurs. Point P is the intersection point of perpendicular bisector (L) and corneal anterior surface.
Figure 2The registered image at 30 degrees scanned angle.
Figure 3The geometrical definition of IOL tilt and decentration.
Figure 4The projection of IOL center point in the plane at different scanned angles.
Figure 5The image of IOL fitting surface and intersection of the anterior and posterior surfaces of IOL.
The IOL tilt and decentration obtained by the method of surface fitting.
| Case | The method of AS-OCT instrument | The method of surface fitting | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Tilt/° | Decentration/ | The IOL tilt (θ)/° | The tilt at horizontal direction/° | The tilt at vertical direction/° | Decentration/ | Optical diameter/ |
|
| |
| 1 | 7.90 | 1.01 | 7.52 | 86.19 | 83.53 | 0.88 | 5.82 | 0.67 | 0.81 |
| 2 | 4.30 | 0.83 | 3.35 | 87.82 | 87.46 | 0.71 | 5.91 | 0.89 | 0.91 |
| 3 | 3.30 | 0.73 | 2.13 | 89.30 | 87.98 | 0.29 | 5.87 | 0.87 | 0.93 |
| 4 | 4.90 | 0.85 | 2.79 | 87.80 | 89.13 | 0.80 | 5.81 | 0.85 | 0.86 |
| 5 | 6.90 | 0.67 | 5.86 | 88.04 | 84.47 | 0.28 | 5.77 | 0.86 | 0.86 |
| 6 | 4.20 | 2.33 | 3.92 | 86.23 | 88.94 | 1.08 | 5.85 | 0.86 | 0.83 |
| 7 | 3.60 | 0.77 | 1.76 | 88.59 | 88.94 | 0.67 | 6.12 | 0.93 | 0.88 |
| 8 | 3.30 | 0.51 | 2.2 | 88.18 | 88.77 | 0.45 | 5.91 | 0.83 | 0.89 |
| 9 | 5.40 | 0.78 | 2.13 | 88.24 | 88.80 | 0.76 | 5.96 | 0.90 | 0.85 |
| 10 | 5.20 | 0.73 | 4.08 | 87.85 | 86.54 | 0.27 | 6.15 | 0.97 | 0.87 |
| 11 | 6.80 | 0.71 | 5.85 | 89.07 | 84.21 | 0.44 | 6.21 | 0.91 | 0.85 |
| 12 | 5.40 | 0.74 | 4.55 | 86.40 | 87.22 | 0.63 | 6.06 | 0.93 | 0.83 |
Figure 6The relation with IOL tilts by the method of surface fitting and AS-OCT and scanned angles.
Figure 7The relation with IOL decentration by the method of surface fitting and AS-OCT and scanned angles.
IOL decentration and tilt obtained from the image with different scanned angle.
| Scanned angle/° | Tilt/° | Decentration/mm | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Case 6 | Case 4 | Case 12 | Case 6 | Case 4 | Case 12 | |||||||
| Surface fitting | AS-OCT | Surface fitting | AS-OCT | Surface fitting | AS-OCT | Surface fitting | AS-OCT | Surface fitting | AS-OCT | Surface fitting | AS-OCT | |
| 0 | 2.72 | 3.60 | 2.21 | 2.50 | 3.60 | 4.20 | 0.18 | 1.86 | 0.15 | 0.01 | 0.03 | — |
| 15 | 3.93 | 1.50 | 1.70 | 3.40 | 4.20 | 3.50 | 0.68 | 1.69 | 0.35 | 0.85 | 0.13 | 0.68 |
| 30 | 3.82 | 1.90 | 1.47 | 4.10 | 4.50 | 5.40 | 0.44 | 1.39 | 0.52 | 0.79 | 0.29 | 0.74 |
| 45 | 3.41 | 3.20 | 1.21 | 2.00 | 4.53 | 4.10 | 0.17 | 1.32 | 0.66 | 0.78 | 0.42 | 0.31 |
| 60 | 2.83 | 1.30 | 0.40 | 0.50 | 4.21 | 3.60 | 0.11 | 2.40 | 0.75 | 0.33 | 0.53 | 0.28 |
| 75 | 2.03 | 4.20 | 0.30 | 0.50 | 3.64 | 2.90 | 0.38 | 1.67 | 0.80 | 0.67 | 0.60 | 0.21 |
| 90 | 1.1 | 2.00 | 0.92 | 1.60 | 2.79 | 2.10 | 0.63 | 2.16 | 0.78 | 0.16 | 0.63 | 0.03 |
| 105 | 0.04 | 0.00 | 1.40 | 1.00 | 1.77 | 1.20 | 0.83 | 2.34 | 0.72 | 0.59 | 0.62 | 0.07 |
| 120 | 0.98 | 2.40 | 1.85 | 2.10 | 0.62 | 0.00 | 0.98 | 2.34 | 0.61 | 0.31 | 0.56 | 0.42 |
| 135 | 1.13 | 3.10 | 2.21 | 2.70 | 0.59 | 0.00 | 0.87 | — | 0.45 | 0.03 | 0.47 | 0.03 |
| 150 | 2.74 | 2.40 | 2.40 | 3.50 | 1.74 | 1.20 | 1.07 | 1.94 | 0.26 | 0.03 | 0.34 | 0.24 |
| 165 | 3.38 | 2.30 | 2.35 | 4.90 | 2.76 | 3.50 | 1.01 | 1.77 | 0.06 | 0.77 | 0.20 | 0.32 |
| IOL | 3.92 | 4.20 | 2.37 | 4.90 | 4.55 | 5.40 | 1.08 | 2.40 | 0.80 | 0.85 | 0.63 | 0.74 |
IOL tilt and decentration calculated by surface fitting method based on 6, 7, 8, 9, 10, and 11 pieces of AS-OCT images.
| AS-OCT image | Case 5 | Case 6 | Case 1 | Case 4 | ||||
|---|---|---|---|---|---|---|---|---|
| Tilt/° | Decentration/mm | Tilt/° | Decentration/mm | Tilt/° | Decentration/mm | Tilt/° | Decentration/mm | |
| 11 | 5.86 | 0.28 | 3.92 | 1.08 | — | — | — | — |
| 10 | 6.10 | 0.25 | 3.94 | 1.02 | 7.50 | 0.88 | 2.79 | 0.80 |
| 9 | 6.05 | 0.27 | 3.73 | 1.03 | 7.85 | 0.81 | 2.56 | 0.79 |
| 8 | 5.79 | 0.25 | 3.76 | 1.03 | 7.49 | 0.84 | 2.54 | 0.81 |
| 7 | 6.45 | 0.21 | 3.70 | 1.02 | 6.69 | 0.86 | 2.63 | 0.79 |
| 6 | 6.22 | 0.24 | 3.68 | 1.03 | 6.35 | 0.82 | 2.54 | 0.79 |