Literature DB >> 23929360

A review of the confirmation algorithm and related timing cycles of St. Jude ICDs.

S Serge Barold1, Roland X Stroobandt.   

Abstract

The confirmation algorithm of implantable cardioverter-defibrillators (ICDs) functioning in the uncommitted mode verifies the presence of a continuing tachyarrhythmia during or following charging of the capacitor. Confirmation is usually applied in relation to the first shock. The details of the normal reconfirmation process of St. Jude ICDs and its role in delayed shock delivery or other abnormalities have received relatively little attention. This review describes the confirmation process of St. Jude ICDs and provides an illustrative clinical example of delayed shock as a result of a pacing or sensing lead fracture. The fundamental timing cycles of St. Jude ICDs are also reviewed to facilitate understanding of the problems related to the confirmation process.

Entities:  

Mesh:

Year:  2013        PMID: 23929360     DOI: 10.1007/s00399-013-0276-4

Source DB:  PubMed          Journal:  Herzschrittmacherther Elektrophysiol        ISSN: 0938-7412


  12 in total

Review 1.  Advanced ICD troubleshooting: Part I.

Authors:  Charles D Swerdlow; Paul A Friedman
Journal:  Pacing Clin Electrophysiol       Date:  2005-12       Impact factor: 1.976

2.  ICD shock triggered by atrial bigeminy.

Authors:  S Serge Barold; Bengt Herweg
Journal:  Pacing Clin Electrophysiol       Date:  2006-09       Impact factor: 1.976

Review 3.  ICD monitoring zones: intricacies, pitfalls, and programming tips.

Authors:  Fadi Mansour; Paul Khairy
Journal:  J Cardiovasc Electrophysiol       Date:  2008-02-04

4.  Averaging detection algorithm of ICDs: not so shocking after all!

Authors:  S Serge Barold; Bengt Herweg; Bengt Herwerg
Journal:  Pacing Clin Electrophysiol       Date:  2008-01       Impact factor: 1.976

5.  Inappropriate shocks or inappropriate programming? A review of Guidant's reconfirmation algorithm.

Authors:  Fadi Mansour; Eduard Guash; Antonio Berruezo; Josep Brugada; Jose Maria Tolosana; Lluís Mont
Journal:  Europace       Date:  2009-05-13       Impact factor: 5.214

6.  Inappropriate shock despite successful termination of supraventricular tachycardia by antitachycardia pacing during charging.

Authors:  Martin Bernier; Vidal Essebag
Journal:  Pacing Clin Electrophysiol       Date:  2010-09       Impact factor: 1.976

7.  Near fatal arrhythmia induction by an implantable cardioverter defibrillator.

Authors:  Christian G Wollmann; Ulrich Göttker; Dirk Böcker; Rainer Gradaus
Journal:  J Cardiovasc Electrophysiol       Date:  2006-06-09

8.  Clinical performance of a specific algorithm to reconfirm self-terminating ventricular arrhythmias in current implantable cardioverter-defibrillators.

Authors:  T M Raedle-Hurst; J Wiecha; J O Schwab; H Schmitt; M Hinrichs; H Tillmanns; B Waldecker
Journal:  Am J Cardiol       Date:  2001-10-01       Impact factor: 2.778

9.  Shocking truths about implantable cardioverter defibrillator monitoring zones.

Authors:  Fadi Mansour; Bernard Thibault; Marc Dubuc; Peter G Guerra; Laurent Macle; Ramon Brugada; Denis Roy; Mario Talajic; Paul Khairy
Journal:  Pacing Clin Electrophysiol       Date:  2007-09       Impact factor: 1.976

10.  Defibrillator shock due to ventricular trigeminy.

Authors:  Krishnakumar Nair; Raja Selvaraj; Vijay S Chauhan
Journal:  Europace       Date:  2008-12-24       Impact factor: 5.214

View more