Near fatal arrhythmia induction by an implantable cardioverter defibrillator.

Christian G Wollmann1, Ulrich Göttker, Dirk Böcker, Rainer Gradaus.   

Abstract

Sudden death among implantable cardioverter defibrillator (ICD) recipients is well known. The underlying mechanism remains unclear in most cases. One possible mechanism is an accidentally induced arrhythmia by the ICD itself. This report describes for the first time a case of a life-threatening ventricular arrhythmia induced by a specific algorithm for ventricular arrhythmia reconfirmation before shock delivery in a fourth-generation ICD.

Entities:  

Mesh:

Year:  2006        PMID: 16764706     DOI: 10.1111/j.1540-8167.2006.00532.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  1 in total

Review 1.  A review of the confirmation algorithm and related timing cycles of St. Jude ICDs.

Authors:  S Serge Barold; Roland X Stroobandt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-08-09