Literature DB >> 20230474

Inappropriate shock despite successful termination of supraventricular tachycardia by antitachycardia pacing during charging.

Martin Bernier1, Vidal Essebag.   

Abstract

We present a case of a 70-year-old man who received an inappropriate implantable cardioverter-defibrillator shock for sinus tachycardia falling outside of the ventricular tachycardia zone. This occurred after termination of supraventricular tachycardia falling into the ventricular fibrillation zone by antitachycardia pacing. Particularities of the programming algorithms are reviewed. ©2010, The Authors. Journal compilation ©2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20230474     DOI: 10.1111/j.1540-8159.2010.02714.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

Review 1.  A review of the confirmation algorithm and related timing cycles of St. Jude ICDs.

Authors:  S Serge Barold; Roland X Stroobandt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-08-09