Dosing accuracy and insulin flow rate characteristics of a new disposable insulin pen, FlexTouch, compared with SoloSTAR.

Janine Bohnet1, Mario Schmitz, Stefan Kamlot, Mona Abdel-Tawab.   

Abstract

BACKGROUND: The introduction of the FlexTouch® (FT; Novo Nordisk; insulin aspart), a prefilled insulin pen with a spring-loaded mechanism, has created more insulin pen options. The present study compared the dosing accuracy of the FT with that of the manually operated SoloSTAR® (SS; Sanofi; insulin glulisine). The volumetric flow rate of insulin delivery with the FT was also evaluated.
METHODS: Thirty unused pens from one batch of each pen type were used to test dosing accuracy at minimum (1 U), mid (40 U), and maximum dose (80 U). Statistical analysis was performed using Student's t-test. Insulin flow was determined with 20 FT pens ejecting 80 U three times per pen using a mass flow meter.
RESULTS: Both insulin pens revealed excellent dosing accuracy, delivering all doses within the limits set by ISO 11608-1:2000. The average relative deviation of the actual dose from the target dose was +6.86% and +3.87% at the minimum, -0.72% and -1.01% at the mid, and -0.68% and -1.06% at the maximum dose for the SS and FT, respectively. The difference at maximum dose was statistically significant (p = .006) in favor of the SS. The FT showed a mean maximum flow rate of 15.61 U/s, with 80.52% of the total dose delivered at an injection speed exceeding 10 U/s.
CONCLUSIONS: This study demonstrated excellent dosing accuracy for the SS and FT at all tested dosage levels. The average maximum injection speed of the FT was considerably higher than the usual range of 6-10 U/s assumed for a smooth and painless injection. Further investigations should confirm the clinical relevance.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23911185      PMCID: PMC3879768          DOI: 10.1177/193229681300700426

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  15 in total

1.  Study on the dosing accuracy of commonly used disposable insulin pens.

Authors:  Meike Krzywon; Thomas van der Burg; Uwe Fuhr; Manfred Schubert-Zsilavecz; Mona Abdel-Tawab
Journal:  Diabetes Technol Ther       Date:  2012-06-13       Impact factor: 6.118

2.  Dosing accuracy of two insulin pre-filled pens.

Authors:  Toshinari Asakura; Hiroaki Seino; Miho Kageyama; Noriaki Yohkoh
Journal:  Curr Med Res Opin       Date:  2008-04-08       Impact factor: 2.580

3.  Comparison of the dose accuracy of prefilled insulin pens.

Authors:  Alexander Weise; Johannes W Pfützner; Julia Borig; Anna M Pfützner; Michael Safinowski; Heike Hänel; Petra B Musholt; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2009-01

4.  Engineering study comparing injection force and dose accuracy between two prefilled insulin injection pens.

Authors:  Debra A Ignaut; Michael R Opincar; Paula E Clark; Melanie K Palaisa; Sheila M Lenox
Journal:  Curr Med Res Opin       Date:  2009-12       Impact factor: 2.580

Review 5.  Reusable and disposable insulin pens for the treatment of diabetes: understanding the global differences in user preference and an evaluation of inpatient insulin pen use.

Authors:  Riccardo Perfetti
Journal:  Diabetes Technol Ther       Date:  2010-06       Impact factor: 6.118

6.  Comparative device assessments: Humalog KwikPen compared with vial and syringe and FlexPen.

Authors:  Debra A Ignaut; Sherwyn L Schwartz; Samiha Sarwat; Heather L Murphy
Journal:  Diabetes Educ       Date:  2009 Sep-Oct       Impact factor: 2.140

7.  Prefilled insulin device with reduced injection force: patient perception and accuracy.

Authors:  A Pfützner; T Reimer; C Hohberg; L P F Frøkjaer; C Jørgensen
Journal:  Curr Med Res Opin       Date:  2008-07-31       Impact factor: 2.580

Review 8.  Assessment of patient-reported outcomes of insulin pen devices versus conventional vial and syringe.

Authors:  Cliff Molife; Lauren J Lee; Lizheng Shi; Monika Sawhney; Sheila M Lenox
Journal:  Diabetes Technol Ther       Date:  2009-08       Impact factor: 6.118

Review 9.  Insulin delivery with FlexPen: dose accuracy, patient preference and adherence.

Authors:  A Pfützner; T Asakura; B Sommavilla; Wc Lee
Journal:  Expert Opin Drug Deliv       Date:  2008-08       Impact factor: 6.648

10.  A multicenter, randomized, open-label, comparative, two-period crossover trial of preference, efficacy, and safety profiles of a prefilled, disposable pen and conventional vial/syringe for insulin injection in patients with type 1 or 2 diabetes mellitus.

Authors:  Mary Korytkowski; David Bell; Carol Jacobsen; Rudee Suwannasari
Journal:  Clin Ther       Date:  2003-11       Impact factor: 3.393

View more
  5 in total

1.  FlexTouch: An Insulin Pen-Injector with a Low Activation Force Across Different Insulin Formulations, Needle Technologies, and Temperature Conditions.

Authors:  Niels Gudiksen; Thibaud Hofstätter; Birgitte B Rønn; Thomas Sparre
Journal:  Diabetes Technol Ther       Date:  2017-07-13       Impact factor: 6.118

2.  Analysis and perspective of Dosing accuracy and insulin flow rate characteristics of a new disposable insulin pen, FlexTouch, compared with SoloSTAR.

Authors:  Harold K Yeager; Debra A Ignaut
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

3.  Dosing Accuracy of Two Disposable Insulin Pens According to New ISO 11608-1: 2012 Requirements.

Authors:  Mona Abdel-Tawab; Mario Schmitz; Stefan Kamlot; Manfred Schubert-Zsilavecz
Journal:  J Diabetes Sci Technol       Date:  2015-07-17

4.  Do patients with diabetes use the insulin pen properly?

Authors:  Betul Tosun; Fatma Ilknur Cinar; Zeliha Topcu; Burcu Masatoglu; Nurten Ozen; Gulcan Bagcivan; Ozge Kilic; Canan Demirci; Ayten Altunbas; Alper Sonmez
Journal:  Afr Health Sci       Date:  2019-03       Impact factor: 0.927

5.  User-Centric Approach to Specifying Technical Attributes of Drug Delivery Devices: Empirical Study of Autoinjector-Cap Removal Forces.

Authors:  Andreas Schneider; Philipp Richard; Philippe Mueller; Christoph Jordi; Mary Yovanoff; Jakob Lange
Journal:  Patient Prefer Adherence       Date:  2021-02-02       Impact factor: 2.711

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.