Prefilled insulin device with reduced injection force: patient perception and accuracy.

A Pfützner1, T Reimer, C Hohberg, L P F Frøkjaer, C Jørgensen.   

Abstract

OBJECTIVE: The injection force and the patient perception of the Next Generation FlexPen (NGFP) with design modifications aimed at reducing injection force was assessed. The accuracy and precision of the NGFP was also tested under standard conditions. RESEARCH DESIGN AND METHODS: Dosing accuracy was tested (according to ISO 11608 requirements) at 1 IU, 30 IU and 60 IU doses (acceptable limits were 1 +/- 1 IU (0-2 IU), 30 +/- 1.5 IU (28.5-31.5 IU), and 60 +/- 3 IU (57-63 IU)). Pens were tested at reference conditions (18-28 degrees C and relative humidity 25-75%). Delivered doses were measured on a sensitive balance and corrected for the specific density of the insulin aspart used (according to ISO 11608-1). Precision was calculated from the variance around the mean delivered dose. The injection force of NGFP was measured, and user-preference of NGFP and FlexPen (FP) were compared in 50 patients with type 2 diabetes.
RESULTS: The mean injection force with NGFP and FP was 12.57 +/- 1.81 N and 17.90 +/- 1.51 N (p<0.001), respectively. Almost twice as many patients rated the injection force as 'good' or 'very good' with NGFP (80%, 72% and 38% when delivering 20 IU, 40 IU and 60 IU, respectively) compared with FP (48%, 32% and 20% when delivering 20 IU, 40 IU and 60 IU, respectively) and 76% of patients rated NGFP as superior, in terms of simplicity and comfort, to FP. NGFP accurately delivered the set doses (means [SD] were 0.98 [0.06] IU, 29.98 [0.18] IU, and 59.93 [0.24] IU for the 1 IU, 30 IU and 60 IU doses, respectively).
CONCLUSIONS: These results show that NGFP has a 30% reduction in injection force compared with FP and was rated as 'more simple and comfortable to use' by patients. Furthermore, NGFP was as accurate and as precise as FP.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18674410     DOI: 10.1185/03007990802329264

Source DB:  PubMed          Journal:  Curr Med Res Opin        ISSN: 0300-7995            Impact factor:   2.580


  14 in total

1.  FlexTouch: a prefilled insulin pen with a novel injection mechanism with consistent high accuracy at low- (1 U), medium- (40 U), and high- (80 U) dose settings.

Authors:  Jakob Oest Wielandt; Marcus Niemeyer; Marianne Rye Hansen; Ditlef Bucher; Niels Bjerrum Thomsen
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

2.  Study on the dosing accuracy of commonly used disposable insulin pens.

Authors:  Meike Krzywon; Thomas van der Burg; Uwe Fuhr; Manfred Schubert-Zsilavecz; Mona Abdel-Tawab
Journal:  Diabetes Technol Ther       Date:  2012-06-13       Impact factor: 6.118

3.  Comparison of the dose accuracy of prefilled insulin pens.

Authors:  Alexander Weise; Johannes W Pfützner; Julia Borig; Anna M Pfützner; Michael Safinowski; Heike Hänel; Petra B Musholt; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2009-01

4.  Practical experience with a new prefilled pen device for a long-acting insulin analog in Japan.

Authors:  Hitomi Fujii; Mikako Baba; Takayuki Mori; Yoshiko Asada; Akihito Nonaka; Yuko Watanabe; Yoshiaki Kawagoe; Miyuki Nogawa; Takaichi Miyakawa; Masafumi Kitaoka
Journal:  J Diabetes Sci Technol       Date:  2014-05-15

5.  A comparison of glide force characteristics between 2 prefilled insulin lispro pens.

Authors:  Tina M Rees; Amanda H Lennartz; Debra A Ignaut
Journal:  J Diabetes Sci Technol       Date:  2015-01-14

6.  Needle with a novel attachment versus conventional screw-thread needles: a preference and usability test among adults with diabetes and impaired manual dexterity.

Authors:  Birtha Hansen; Søren K Lilleøre; Gitte Ter-Borch
Journal:  Diabetes Technol Ther       Date:  2011-03-15       Impact factor: 6.118

Review 7.  Evolution of diabetes insulin delivery devices.

Authors:  Jean-Louis Selam
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

Review 8.  Practical aspects of insulin pen devices.

Authors:  Teresa L Pearson
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

9.  Dosing accuracy and insulin flow rate characteristics of a new disposable insulin pen, FlexTouch, compared with SoloSTAR.

Authors:  Janine Bohnet; Mario Schmitz; Stefan Kamlot; Mona Abdel-Tawab
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

10.  An evaluation of prefilled insulin pens: a focus on the Next Generation FlexPen(®).

Authors:  Estella M Davis; Emily L Sexson; Mikayla L Spangler; Pamela A Foral
Journal:  Med Devices (Auckl)       Date:  2010-08-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.