Dosing Accuracy of Two Disposable Insulin Pens According to New ISO 11608-1: 2012 Requirements.

Mona Abdel-Tawab1, Mario Schmitz2, Stefan Kamlot2, Manfred Schubert-Zsilavecz3.   

Abstract

OBJECTIVE: The aim was to compare 2 disposable insulin pens, FlexTouch® (Novo Nordisk, insulin aspart) and SoloSTAR® (Sanofi, insulin glulisine), according to new ISO 11608-1:2012 requirements for dosing accuracy.
METHODS: Sixty pens of each type were tested at 1, 40, and 80 U doses. Following the new ISO requirements, each dose was delivered from the front, middle, and rear one-third of the pen. Statistical analysis was performed using Student's t test.
RESULTS: Both pens delivered all doses within ISO limits. The difference between the average measured dose and the target dose was significantly smaller for SoloSTAR than FlexTouch at 40 U (P = .009) and 80 U (P = .008), but not at 1 U (P = .417).
CONCLUSION: Both insulin pens fulfilled the dosing accuracy requirements defined by ISO 11608-1:2012 at all 3 dosage levels.
© 2015 Diabetes Technology Society.

Entities:  

Keywords:  FlexTouch; SoloSTAR; disposable insulin pens; dosing accuracy

Mesh:

Substances:

Year:  2015        PMID: 26187635      PMCID: PMC4738205          DOI: 10.1177/1932296815595983

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


Patients with diabetes generally prefer insulin pens over vial and syringe because of the ease of use, discreetness of injection, and portability of the pens.[1-3] Insulin pens account for over 60% of insulin delivery worldwide.[4] Conversion from vial and syringe delivery to insulin pens is associated with improved medication adherence and reduced likelihood of hypoglycemic events, due in part to the greater dosing accuracy of pens.[4-6] To assure the accuracy of the injected dose, insulin pens need to fulfill ISO 11608-1 requirements for dosing accuracy, which they have demonstrated in several studies.[4,7-11] Substantial modifications have been made to the ISO requirements for needle-based injection systems; the new requirements differ from the old in defining the design of dose delivery, with each tested dose delivered from the front, middle, and rear one-third of the pen cartridge.[12] Recently, FlexTouch® (FT), a disposable insulin pen with a spring-loaded mechanism, has been introduced as another pen option. The aim of the present study was to compare the dosing accuracy of SoloSTAR® (SS) with FT according to the new ISO 11608-1:2012 requirements.[12] The 2 pens were previously compared according to the ISO 11608-1:2000[13] requirements at which both showed excellent dosing accuracy.[7,14]

Methods

SS pens (sanofi-aventis Deutschland GbmH, Frankfurt, Germany) were purchased from a German pharmacy, while FT pens (Novo Nordisk A/S, Bagsvaerd, Denmark) were obtained through an international pharmacy in the United Kingdom. The pens were shipped under cool conditions (36-46°F) and stored both unopened and while in use as indicated in the Information for Users for each pen type. An overview of the tested pens and corresponding injection needles is listed in Table 1. Needles were applied according to the manufacturers’ instruction leaflet.
Table 1.

Insulin Pens and Corresponding Needles Included in This Study.

Insulin penManufacturerBatchInsulinNeedles
SoloSTARSanofi2F082AGlulisine (Apidra®)BD Micro-Fine (0.25 mm [31G] × 5 mm)
FlexTouchNovo NordiskAP51446Aspart (NovoRapid®)NovoFine (0.25 mm [31G] × 6 mm)
Insulin Pens and Corresponding Needles Included in This Study. Dosing accuracy was determined as described previously by Bohnet et al,[7] except that the dosing scheme comprised 1 single measurement of each target dose per pen according to ISO 11608-1:2012 requirements.[12] Consequently, instead of 30 pens as per the former ISO standard,[13] 60 pens of each pen type were investigated. Furthermore, each target dose was delivered from the front one-third, middle one-third, and rear one-third of the pen cartridge with a corresponding transfer step in between the doses (Figure 1).
Figure 1.

Dosing scheme according new ISO 11608-1:2012.

Dosing scheme according new ISO 11608-1:2012. Target doses of 1 U (minimum), 40 U (middle), and 80 U (maximum) were ejected and measured for each pen. All measurements were performed by a single investigator to eliminate potential investigator variability. The individual insulin pens were operated according to each manufacturer’s instruction leaflet. The plunger was kept depressed after each dose for 10 seconds and 6 seconds for SS and FT, respectively, to ensure that the entire dialed dose had been expelled. Each dose was deposited in a beaker containing a 0.5-1 cm layer of liquid paraffin. The dose was weighed immediately using an analytical balance (XP205/M, Mettler Toledo AG, Gießen, Germany) with an accuracy of 0.00001 g. The balance was zeroed before each dose of insulin was deposited and weighed. Weights were corrected for the relative density of each insulin formulation (1.0066 for insulin aspart; 1.0072 for insulin glulisine). For each dose application, a new injection needle was used and an air shot included before the injection. The arithmetic average of the actual doses was calculated, as well as the standard deviation, the average deviation (percentage) from the target dose, and the statistical tolerance interval. The evaluation of dose accuracy was based on the guidelines of the ISO 11608-1:2012,[12] allowing a deviation not more than ± 1 U at the 1 U (0-2 U) dosage level, ± 5% (± 2 U) at the 40 U (38-42 U) dosage level, and ± 5% (± 4 U) at the 80 U (76-84 U) dosage level. In addition, the calculated statistical tolerance interval for each pen should lay within the specified upper and lower acceptance limits for each target dose by using a 95% confidence interval and probability content of P = .975 for 60 measurements.[12] To compare the dose accuracy of SS and FT, statistical analysis was performed using Student’s t test. A P value < .05 was considered to be statistically significant.

Results

Both insulin pens showed excellent dosing accuracy at all dosage levels (Figure 2). The average values of the actual measured doses were closer to the target dose for SS at all 3 dosage levels compared with FT (Table 2). Differences between average values of the actual dose from the target dose were statistically significant at the middle 40 U dose (SS: −0.12; FT: −0.34; P = .009) and the maximum 80 U dose (SS: −0.36; FT: −0.67; P = .008) in favor of SS; the difference was not significant at the minimum 1 U dose (SS: +0.03; FT: +0.04; P = .417). The average relative deviation of the actual dose from the target dose was +2.69% and +3.91% at the minimum, −0.31% and −0.85% at the middle, and −0.45% and −0.84% at the maximum dose for SS and FT, respectively. No single dose of SS or FT was detected outside the specified limits, and the statistical tolerance intervals defined by the ISO standards were met by both pens at all 3 dosage levels (Table 2).
Figure 2.

Distribution of the actual doses of SoloSTAR and FlexTouch at the 1 U (A), 40 U (B), and 80 U (C) dosage levels. The line near the middle of each box represents the median (50th percentile). The lower and upper ends of the box represent the 25th and 75th percentiles. The small square is the mean. The ends of the whiskers represent the minimum and maximum values, respectively. The horizontal lines at 0 and 2 in panel A, 38 and 42 in panel B, and 76 and 84 in panel C represent the ISO limits for the 1U, 40U, and 80U dosage levels. The dose tested and the corresponding ISO limits are represented by the horizontal lines in each plot.

Table 2.

Overview of the Average Actual Doses, Standard Deviation, and Statistical Tolerance Interval at Each Dosage Level for SoloSTAR and FlexTouch.

Actual dose
PenTarget dose (U)Average dose (U)Standard deviation (U)Statistical tolerance interval (U)[a]
SoloSTAR11.030.100.77-1.28
FlexTouch11.040.050.90-1.17
SoloSTAR4039.880.5038.53-41.22
FlexTouch4039.660.3638.69-40.63
SoloSTAR8079.640.6078.04-81.24
FlexTouch8079.330.7077.46-81.19

According to the International Organization for Standardization (DIN EN ISO 11608-1:2012), the acceptance range for the statistical tolerance interval is 0-2 U for the 1 U dosage level, 38-42 U for the 40 U dosage level, and 76-84 U for the 80 U dosage level.

Distribution of the actual doses of SoloSTAR and FlexTouch at the 1 U (A), 40 U (B), and 80 U (C) dosage levels. The line near the middle of each box represents the median (50th percentile). The lower and upper ends of the box represent the 25th and 75th percentiles. The small square is the mean. The ends of the whiskers represent the minimum and maximum values, respectively. The horizontal lines at 0 and 2 in panel A, 38 and 42 in panel B, and 76 and 84 in panel C represent the ISO limits for the 1U, 40U, and 80U dosage levels. The dose tested and the corresponding ISO limits are represented by the horizontal lines in each plot. Overview of the Average Actual Doses, Standard Deviation, and Statistical Tolerance Interval at Each Dosage Level for SoloSTAR and FlexTouch. According to the International Organization for Standardization (DIN EN ISO 11608-1:2012), the acceptance range for the statistical tolerance interval is 0-2 U for the 1 U dosage level, 38-42 U for the 40 U dosage level, and 76-84 U for the 80 U dosage level.

Discussion

The present study demonstrates that both SS and FT meet the new ISO 11608-1:2012[12] requirements for dosing accuracy at ambient temperature. Applying the new dosing scheme with single measurements of each target dose delivered from the front, middle, and the rear one-third of the pen and the higher sample size of 60 pens for each pen type, the average deviation between actual and target dose and the standard deviation remained very low for both pens. In addition to the significantly better dosing accuracy at maximum dose (80 U) with SS reported by Bohnet et al[7] when applying ISO 11608-1:2000,[13] the current study revealed significantly closer actual doses with SS at 40 U and 80 U compared with FT. However, the difference in dosing accuracy between SS and FT cannot be considered clinically relevant. Wielandt et al[14] reported significantly better dosing accuracy for FT than SS at 1 U. This could not be confirmed according to the new standard in the current study where equivalent results were obtained with the 2 pens. Götzche et al[15] also reported equivalent dosing accuracy according to the new ISO requirement at 1 U, 40 U, and 80 U for SS and FT. However, contrary to the new ISO 11608-1:2012[12] requirement, only 30 insulin pens with a dosing scheme comprising 2 single measurements of each target dose per pen were used. In the current study, all single doses, 180 with each pen, were delivered within the ISO limits, despite the larger pen sample size. The current results confirm the excellent dosing accuracy for SS[4,7-11] and FT[7,14] demonstrated in previous studies performed according to ISO 11608-1:2000.[13] Furthermore, the results demonstrate that the spring-loaded mechanism of FT, which is less dependent on the operator than SS, does not translate into a dosing accuracy advantage compared with the manually operated SS.
  13 in total

1.  FlexTouch: a prefilled insulin pen with a novel injection mechanism with consistent high accuracy at low- (1 U), medium- (40 U), and high- (80 U) dose settings.

Authors:  Jakob Oest Wielandt; Marcus Niemeyer; Marianne Rye Hansen; Ditlef Bucher; Niels Bjerrum Thomsen
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

2.  Study on the dosing accuracy of commonly used disposable insulin pens.

Authors:  Meike Krzywon; Thomas van der Burg; Uwe Fuhr; Manfred Schubert-Zsilavecz; Mona Abdel-Tawab
Journal:  Diabetes Technol Ther       Date:  2012-06-13       Impact factor: 6.118

3.  Injection force and dose accuracy of FlexTouch for the delivery of a new basal insulin.

Authors:  Dorte Götzche; Bo Ølgaard Rasmussen; Michael Torben Pedersen; Thomas Sparre; Ditlef Bucher; Marcus Niemeyer
Journal:  Expert Opin Drug Deliv       Date:  2013-12       Impact factor: 6.648

Review 4.  Dose accuracy and injection force dynamics of a novel disposable insulin pen.

Authors:  Alastair Clarke; Geralyn Spollett
Journal:  Expert Opin Drug Deliv       Date:  2007-03       Impact factor: 6.648

Review 5.  Reusable and disposable insulin pens for the treatment of diabetes: understanding the global differences in user preference and an evaluation of inpatient insulin pen use.

Authors:  Riccardo Perfetti
Journal:  Diabetes Technol Ther       Date:  2010-06       Impact factor: 6.118

6.  Comparative device assessments: Humalog KwikPen compared with vial and syringe and FlexPen.

Authors:  Debra A Ignaut; Sherwyn L Schwartz; Samiha Sarwat; Heather L Murphy
Journal:  Diabetes Educ       Date:  2009 Sep-Oct       Impact factor: 2.140

Review 7.  Assessment of patient-reported outcomes of insulin pen devices versus conventional vial and syringe.

Authors:  Cliff Molife; Lauren J Lee; Lizheng Shi; Monika Sawhney; Sheila M Lenox
Journal:  Diabetes Technol Ther       Date:  2009-08       Impact factor: 6.118

8.  Dosing accuracy and insulin flow rate characteristics of a new disposable insulin pen, FlexTouch, compared with SoloSTAR.

Authors:  Janine Bohnet; Mario Schmitz; Stefan Kamlot; Mona Abdel-Tawab
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

9.  Dose accuracy comparison between SoloSTAR and FlexPen at three different dose levels.

Authors:  Alfred Penfornis; Kristian Horvat
Journal:  Diabetes Technol Ther       Date:  2008-10       Impact factor: 6.118

10.  A multicenter, randomized, open-label, comparative, two-period crossover trial of preference, efficacy, and safety profiles of a prefilled, disposable pen and conventional vial/syringe for insulin injection in patients with type 1 or 2 diabetes mellitus.

Authors:  Mary Korytkowski; David Bell; Carol Jacobsen; Rudee Suwannasari
Journal:  Clin Ther       Date:  2003-11       Impact factor: 3.393

View more
  3 in total

1.  Clinical Use and Evaluation of Insulin Pens.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2015-08-31

Review 2.  Understanding the Clinical Profile of Insulin Degludec, the Latest Basal Insulin Approved for Use in Canada: a Narrative Review.

Authors:  Vincent Woo; Lori Berard; Robert Roscoe
Journal:  Diabetes Ther       Date:  2020-09-17       Impact factor: 2.945

3.  Open-Label, Randomized, Two-Way, Crossover Study Assessing the Bioequivalence of the Liquid Formulation versus the Freeze-Dried Formulation of Recombinant Human FSH and Recombinant Human LH in a Fixed 2:1 Combination (Pergoveris®) in Pituitary-Suppressed Healthy Women.

Authors:  Wilhelmina Bagchus; Özkan Yalkinoglu; Peter Wolna
Journal:  Front Endocrinol (Lausanne)       Date:  2018-01-11       Impact factor: 5.555

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.