Dosing accuracy of two insulin pre-filled pens.

Toshinari Asakura1, Hiroaki Seino, Miho Kageyama, Noriaki Yohkoh.   

Abstract

OBJECTIVE: This study was designed to determine the dose accuracy of two commonly available insulin pre-filled pens for use in diabetes - FlexPen (Novo Nordisk A/S, Bagsvaerd, Denmark) (FP) and SoloStar (Sanofi-Aventis, Paris, France) (SS). RESEARCH DESIGN AND METHODS: Dosing accuracy was tested at 5 U, 10 U and 30 U doses - three previously unused pens of both pre-filled pens were used for each dose. At the 5 U dose each pen was tested 42 times, at the 10 U dose each pen was tested 25 times and at the 30 U dose each pen was tested 9 times. The pre-filled pens were used strictly according to manufacturers' instructions and measurements were made on a sensitive balance and corrected for the specific density of the insulin formulations used. Specified limits were based on ISO standards (+/-1 U for the 5 U and 10 U doses and +/-1.5 U for the 30 U dose).
RESULTS: FP was more accurate for injecting 5 U, 10 U and 30 U doses (absolute mean 4.95 +/- 0.19 U, 9.61 +/- 0.27 U and 29.70 +/- 0.34 U, respectively) than SS (absolute mean 4.86 +/- 0.39 U, 9.27 +/- 0.52 U and 28.73 +/- 0.47 U, respectively). No doses were outside specified limits for 5 U and 30 U with FP and 1.3% of doses were outside these limits at the 10 U dose. For SS; 1.6%, 29.3% and 33.3% of doses were below the pre-specified threshold for the 5 U, 10 U and 30 U doses, respectively.
CONCLUSIONS: This non-blinded comparison indicates that FP is a more accurate insulin pre-filled pen than SS at three different insulin doses. In this single-user study, FP delivered consistent and accurate doses of insulin but SS had a high frequency of under dosing. No assessment was made of user variability in this study.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18400146     DOI: 10.1185/030079908x297394

Source DB:  PubMed          Journal:  Curr Med Res Opin        ISSN: 0300-7995            Impact factor:   2.580


  10 in total

1.  FlexTouch: a prefilled insulin pen with a novel injection mechanism with consistent high accuracy at low- (1 U), medium- (40 U), and high- (80 U) dose settings.

Authors:  Jakob Oest Wielandt; Marcus Niemeyer; Marianne Rye Hansen; Ditlef Bucher; Niels Bjerrum Thomsen
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

2.  Study on the dosing accuracy of commonly used disposable insulin pens.

Authors:  Meike Krzywon; Thomas van der Burg; Uwe Fuhr; Manfred Schubert-Zsilavecz; Mona Abdel-Tawab
Journal:  Diabetes Technol Ther       Date:  2012-06-13       Impact factor: 6.118

3.  Comparison of the dose accuracy of prefilled insulin pens.

Authors:  Alexander Weise; Johannes W Pfützner; Julia Borig; Anna M Pfützner; Michael Safinowski; Heike Hänel; Petra B Musholt; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2009-01

4.  Dose accuracy and injection force of different insulin glargine pens.

Authors:  Arnd Friedrichs; Janine Bohnet; Volker Korger; Steffen Adler; Manfred Schubert-Zsilavecz; Mona Abdel-Tawab
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

5.  Sources of glycemic variability--what type of technology is needed?

Authors:  Jonas Kildegaard; Toke Folke Christensen; Ole Kristian Hejlesen
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

6.  Comparison of intuitiveness, ease of use, and preference in two insulin pens.

Authors:  Toshinari Asakura; Klaus H Jensen
Journal:  J Diabetes Sci Technol       Date:  2009-03-01

7.  Injection force of reusable insulin pens: Novopen 4, Lilly Luxura, Berlipen, and ClikSTAR.

Authors:  Arnd Friedrichs; Volker Korger; Steffen Adler
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

8.  A comparison of dosing accuracy: visually impaired and sighted people using insulin pens.

Authors:  Ann S Williams; Patrick A Schnarrenberger
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

Review 9.  Practical aspects of insulin pen devices.

Authors:  Teresa L Pearson
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

10.  Dosing accuracy and insulin flow rate characteristics of a new disposable insulin pen, FlexTouch, compared with SoloSTAR.

Authors:  Janine Bohnet; Mario Schmitz; Stefan Kamlot; Mona Abdel-Tawab
Journal:  J Diabetes Sci Technol       Date:  2013-07-01
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.