Fatma Emel Koçak1, Mustafa Yöntem, Ozlem Yücel, Mustafa Cilo, Ozlem Genç, Ayfer Meral.
Abstract
INTRODUCTION: Today, the pneumatic tube transport system (PTS) is used frequently because of its advantages related to timing and speed. However, the impact of various types of PTS on blood components is unknown. The aim of this study was to examine the influence of PTS on the quality of routine blood cell counts, erythrocyte sedimentation, and certain blood coagulation tests.Entities:
Mesh:
Year: 2013 PMID: 23894866 PMCID: PMC3900063 DOI: 10.11613/bm.2013.024
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Blood analysis results (± SD) and statistical differences between blood samples delivered to laboratory by PTS and by laboratory staff.
| WBC (×103/μL) | 6.87 ± 2.09 | 6.84 ± 2.17 | 0.069 |
| RBC (×106/μL) | 4.79 ± 0.51 | 4.73 ± 0.51 | 0.076 |
| HBG (g/L) | 138.6 ± 1.43 | 138.7 ± 1.44 | 0.950 |
| HTC (%) | 42.55 ± 4.22 | 41.46 ± 5.60 | 0.133 |
| MCV (fL) | 89.12 ± 6.28 | 88.92 ± 6.32 | 0.321 |
| MCH (pg) | 29.08 ± 2.23 | 29.17 ± 2.73 | 0.578 |
| MCHC (g/L) | 326.0 ± 0.68 | 320.8 ± 0.74 | 0.458 |
| PLT (×103/μL) | 243.3 ± 68.9 | 248.3 ± 62.4 | 0.096 |
| MPV (fL) | 9.01 ± 0.98 | 8.85 ± 1.05 | 0.311 |
| PT (s) | 12.26 ± 1.05 | 12.25 ± 1.00 | 0.971 |
| aPTT (s) | 31.21 ± 3.19 | 31.22 ± 3.18 | 0.977 |
| ESR (mm/h) | 18.49 ± 12.68 | 18.04 ± 13.61 | 0.630 |
SD - standard deviation; SI - International System of Units; PTS - pneumatic tube system; WBC - white blood cell; RBC - red blood cell; HBG - haemoglobin; HTC - haematocrit; MCV - mean corpuscular volume; MCH - mean corpuscular haemoglobin; MCHC - mean corpuscular haemoglobin concentration; PLT - platelet; MPV - mean platelet volume; PT - prothrombin time; aPTT - activated partial thromboplastin time; ESR - erythrocyte sedimentation rate.