| Literature DB >> 30079089 |
Devi Subbarayan1, Chidambharam Choccalingam1, Chittode Kodumudi Anantha Lakshmi1.
Abstract
BACKGROUND: Automation helps improve laboratory operational efficiency and reduce the turnaround time. Pneumatic tube systems (PTS) automate specimen transport between the lab and other areas of the hospital. Its effect on complete blood count (CBC) and coagulation is still controversial. AIM: To study the effects of pneumatic tube system sample transport on complete blood count and coagulation parameters to compare them with hand delivered samples.Entities:
Year: 2018 PMID: 30079089 PMCID: PMC6051325 DOI: 10.1155/2018/6940152
Source DB: PubMed Journal: Adv Hematol
Summary of the differences in CBC between pneumatic tube samples and hand delivered samples.
| S. no. | Paired samples | Mean difference | Standard deviation of mean difference | 95% confidence interval of mean difference |
| |
|---|---|---|---|---|---|---|
| Lower | Upper | |||||
| 1 | RBC P-RBC M (x1012/l) | -.03 | .21 | -.08 | .01 | .11 |
| 2 | HB P-HB M (g/dl) | -.01 | .24 | -.06 | .05 | .84 |
| 3 | MCV P-MCV M (fl) | -.82 | 1.9 | -1.28 | -.37 | .001 |
| 4 | MCH P-MCH M (pg) | .03 | .41 | -.06 | .12 | .500 |
| 5 | MCHCP-MCHC M (g/dl) | .36 | .84 | .17 | .56 | <0.001 |
| 6 | RDW P-RDW M (%) | -.33 | .67 | -.49 | -.18 | <0.001 |
| 7 | WBC P-WBC M (x109/l) | .04 | 1.2 | -.24 | .32 | .759 |
| 8 | NE P-NE M (%) | -.45 | 2.94 | -1.13 | .21 | .182 |
| 9 | LY P-LY M (%) | .20 | 3.30 | -.55 | .96 | .590 |
| 10 | MO P-MO M (%) | .30 | 1.37 | -.01 | .62 | .057 |
| 11 | EO P-EO M (%) | .28 | 3.46 | -.50 | 1.08 | .472 |
| 12 | BA P-BA M (%) | .01 | .33 | -.06 | .08 | .754 |
| 13 | PLT P-PLT M (x109/l) | .13 | .26 | .07 | .19 | <0.001 |
| 14 | MPV P-MPV M (fl) | -.18 | .64 | -.32 | -.03 | .017 |
P: pneumatic tube samples; M: hand delivered samples; RBC: red blood cells count; HB: hemoglobin; MCV: mean corpuscular volume; MCH: mean corpuscular hemoglobin; RDW: red cell distribution width; WBC: white blood cell count; NE: neutrophils; LY: lymphocytes; MO: monocytes; EO: eosinophils; BA: basophils; PLT: platelets; MPV: mean platelet volume.
Summary of the differences in PT and APTT between pneumatic tube samples and hand delivered samples.
| S. no. | Paired samples | Mean difference | Standard deviation | 95% confidence interval of mean difference |
| |
|---|---|---|---|---|---|---|
| Lower | Upper | |||||
| 1 | PT P-PT M (secs) | -.38 | 1.3 | -.95 | .18 | .17 |
| 2 | APTT P-APTT M (secs) | -.33 | 3.5 | -1.7 | 1.1 | .63 |
P: pneumatic tube samples; M: hand delivered samples; PT: prothrombin time; APTT: activated partial thromboplastin time.