Maciej Kempa1, Szymon Budrejko, Lukasz Drelich, Tomasz Królak, Grzegorz Raczak, Dariusz Kozłowski.
Abstract
INTRODUCTION: Among the recipients of implantable cardioverter-defibrillators (ICDs), there is a group of patients in whom the defibrillation threshold (DFT) is too high to enable a sufficient safety margin between the DFT and the maximal available output of the device. The aim of the study was to investigate the ability of an additionally implanted single-coil subcutaneous array electrode to reduce the DFT in such patients.Entities:
Keywords: defibrillation threshold; implantable cardioverter-defibrillator; subcutaneous electrode
Year: 2013 PMID: 23847664 PMCID: PMC3701990 DOI: 10.5114/aoms.2013.35480
Source DB: PubMed Journal: Arch Med Sci ISSN: 1734-1922 Impact factor: 3.318
Figure 1An X-ray image of a patient with a cardiac resynchronization therapy defibrillator system with a subcutaneous array electrode. Antero-posterior view
Figure 2An X-ray image of a patient with a cardiac resynchronization therapy defibrillator system with a subcutaneous array electrode. Lateral view
Characteristics of the patients with a subcutaneous electrode included in the study
| Number of patients | Sex | Age | Type of device | Type of defibrillating electrode | Underlying heart disease |
|---|---|---|---|---|---|
| 15 | 13 male, | 56 ±10 | 6 – CRT-D; | 14 – single-coil; | 10 patients – CAD; |
| 2 female | years | 5 – ICD DR; | 1 – dual-coil | 4 patients – DCM; | |
| 4 – ICD VR | 1 patient – HCM |
CAD – coronary artery disease, CRT-D – cardiac resynchronization therapy cardioverter-defibrillator, DCM – dilative cardiomyopathy, HCM – hypertrophic cardiomyopathy, ICD-DR – dual-chamber cardioverter-defibrillator, ICD-VR single-chamber cardioverter-defibrillator
Figure 3Comparison of the defibrillation threshold before and after implantation of a subcutaneous array electrodeDFT – defibrillation threshold,
DFT-SQ – defibrillation threshold before the procedure, DFT + SQ – defibrillation threshold after the procedure
Figure 4Values of the defibrillation threshold in successive patients in the cohort.
DFT – defibrillation threshold, DFT-SQ – defibrillation threshold before the procedure, DFT + SQ – defibrillation threshold after the procedure
Specific values of the defibrillation thresholds for each patient
| Patient | DFT–SQ | DFT + SQ |
|---|---|---|
| 1 | > 30 | 30 |
| 2 | > 35 | 25 |
| 3 | > 35 | 20 |
| 4 | > 40 | 25 |
| 5 | > 25 | 20 |
| 6 | > 35 | No test |
| 7 | > 30 | 20 |
| 8 | > 30 | 25 |
| 9 | > 30 | 30 |
| 10 | > 34 | 30 |
| 11 | > 35 | 25 |
| 12 | > 35 | 25 |
| 13 | > 40 | 34 |
| 14 | > 30 | 20 |
| 15 | > 35 | 25 |
DFT–SQ – defibrillation threshold prior to implantation of the subcutaneous electrode, DFT + SQ – defibrillation threshold with the subcutaneous electrode implanted