High defibrillation threshold in patients with implantable defibrillator: how effective is the subcutaneous finger lead?

Brigitte R Osswald1, Raffaele De Simone, Sabine Most, Ursula Tochtermann, Ahmed Tanzeem, Matthias Karck.   

Abstract

OBJECTIVE: Even in the era of high output implantable cardioverter defibrillator (ICD) devices, a certain proportion of patients cannot be successfully defibrillated with 10 J safety margin. In practice, either the use of a single- or double-coil lead does not guarantee successful termination of induced ventricular fibrillation. Therefore, we investigated the effectiveness of the subcutaneous finger lead placed at the subcutaneous tissue dorsal to the left ventricle in terms of defibrillation threshold (DFT) lowering.
METHODS: Two thousand, eight hundred and three consecutive, unselected patients underwent first-time ICD implantation or ICD device exchange from 6/1999 through 3/2007. The mean age of the patients was 65.4 years. A total of 79.3% of the patients were male. The only implanted subcutaneous lead was the 6996 model by Medtronic Inc.
RESULTS: One hundred and seventy-seven patients (6.3%) received a subcutaneous finger lead implantation. According to the current institutional DFT testing protocol, any failure of the two standard DFT tests in first-time ICD implantation or a failure at the single test in ICD exchange operations was the trigger for subcutaneous finger lead implantation. The proportion of subcutaneous finger lead implantations increased parallel to a markedly larger amount of implantations. Since high output devices became standard, the implantation number of subcutaneous finger leads decreases. The mean of unsuccessful DFTs prior to subcutaneous finger lead implantation was 27.2+/-5.3 J. After subcutaneous finger lead implantation, the mean successful DFT was 17.9+/-3.3 J. No complication due to subcutaneous finger lead implantation occurred.
CONCLUSION: The subcutaneous finger lead is a quick, safe and effective method for DFT lowering.

Entities:  

Mesh:

Year:  2009        PMID: 19144533     DOI: 10.1016/j.ejcts.2008.10.021

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  7 in total

1.  Concept of defibrillation vector in the management of high defibrillation threshold.

Authors:  Kevin Hayes; Abhishek Deshmukh; Sadip Pant; Gareth Tobler; Hakan Paydak
Journal:  World J Cardiol       Date:  2013-04-26

2.  How to Manage a High Defibrillation Threshold in ICD Patients: and Does it Really Matter?

Authors:  Maria Vittoria Matassini; Jeff S Healey
Journal:  Curr Treat Options Cardiovasc Med       Date:  2013-08

3.  Intra-operative defibrillation testing and clinical shock efficacy in patients with implantable cardioverter-defibrillators: the NORDIC ICD randomized clinical trial.

Authors:  Dietmar Bänsch; Hendrik Bonnemeier; Johan Brandt; Frank Bode; Jesper Hastrup Svendsen; Miloš Táborský; Stefan Kuster; Carina Blomström-Lundqvist; Angelika Felk; Tino Hauser; Anna Suling; Karl Wegscheider
Journal:  Eur Heart J       Date:  2015-06-25       Impact factor: 29.983

4.  A Novel Solution for the High Defibrillation Threshold in Patients with a DF-4 Lead: Adding a High Voltage Adaptor/Splitter.

Authors:  Shmuel Inbar; Srikanth Seethala
Journal:  Indian Pacing Electrophysiol J       Date:  2014-05-25

5.  High defibrillation threshold: the science, signs and solutions.

Authors:  Sony Jacob; Victorio Pidlaoan; Jaspreet Singh; Aditya Bharadwaj; Mehul B Patel; Antonio Carrillo
Journal:  Indian Pacing Electrophysiol J       Date:  2010-01-07

6.  Implantation of additional subcutaneous array electrode reduces defibrillation threshold in ICD patients - preliminary results.

Authors:  Maciej Kempa; Szymon Budrejko; Lukasz Drelich; Tomasz Królak; Grzegorz Raczak; Dariusz Kozłowski
Journal:  Arch Med Sci       Date:  2013-06-21       Impact factor: 3.318

7.  Incidence of ineffective safety margin testing (<10 J) and efficacy of routine subcutaneous array insertion during implantable cardioverter defibrillator implantation.

Authors:  Marc-Alexander Ohlow; Marcus Roos; Bernward Lauer; J Christoph Geller
Journal:  Indian Pacing Electrophysiol J       Date:  2016-02-26