Defibrillation efficacy comparing a subcutaneous array electrode versus an "active can" implantable cardioverter defibrillator and a subcutaneous array electrode in addition to an "active can" implantable cardioverter defibrillator: results from active can versus array trials I and II.

R Gradaus1, M Block, K Seidl, J Brunn, F Isgro, D Hammel, B Hauer, G Breithardt, D Bocker.   

Abstract

INTRODUCTION: Placement of implantable cardioverter defibrillators (ICDs) has been simplified by using the shell of a pectorally implanted ICD as a defibrillation electrode in combination with an endocardial right ventricular defibrillation lead. However, a sufficiently low defibrillation threshold (DFT) cannot be obtained in a few patients. Therefore, alternative approaches were systematically tested in the Active Can versus Array Trial (ACAT). METHODS AND
RESULTS: In the first of two prospective randomized studies, the DFT of a subcutaneous left dorsolateral array anode introduced from a pectoral incision was compared to that of a standard active can anode in 68 patients. Intraoperatively, the DFT was determined twice in each patient using either the active can or, in patients with a subcutaneous array lead, once with all three fingers and once omitting the middle finger of the array. The second prospective randomized study included 40 patients. DFT also was determined twice in each patient using an active can in a left pectoral position as anode alone and combined with a left dorsolateral array electrode with two fingers. In ACAT I, stored energy at DFT decreased from 13.1+/-7.7 J (active can) to 9.6+/-6.1 J (three-finger array lead) (P = 0.04), impedance decreased from 53+/-8 ohms to 40+/-6 ohms (P < 0.0001). Omitting the middle finger of the array lead, stored energy at DFT increased by 0.9 J (P = 0.47) and impedance by 2 ohm (P < 0.0001). In ACAT II, DFT and impedance using an active can device were significantly lower when a two-finger array lead was added that decreased stored energy at DFT from 10.1+/-5.2 J to 6.9+/-3.9 J (P = 0.001) and impedance from 56+/-5 1 to 42+/-5 l (P < 0.0001).
CONCLUSION: In combination with a right ventricular defibrillation electrode, a left pectoral subcutaneous array lead improves defibrillation efficacy if used instead of, or in addition to, a left pectoral active can ICD device. Implantation of the array lead can be simplified by using two instead of three fingers, without a significant loss of defibrillation efficacy.

Entities:  

Mesh:

Year:  2001        PMID: 11513444     DOI: 10.1046/j.1540-8167.2001.00921.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  7 in total

1.  Case report of an anomalous single azygos venous coil insertion to reduce the defibrillation threshold in a patient with a right-sided deltopectoral ICD implant.

Authors:  Darragh Patrick Moran; Usman Bhutta; Ibrahim Yearoo; Edward Keelan; James O'Neill; Joseph Galvin
Journal:  Heart Asia       Date:  2013-03-05

2.  Incidence and clinical predictors of low defibrillation safety margin at time of implantable defibrillator implantation.

Authors:  Zhongwei Cheng; Mintu Turakhia; Ronald Lo; Anurag Gupta; Paul C Zei; Henry H Hsia; Amin Al-Ahmad; Paul J Wang
Journal:  J Interv Card Electrophysiol       Date:  2012-03-06       Impact factor: 1.900

3.  Implantation of implantable cardioverter-defibrillators from an ileofemoral approach.

Authors:  Christian Perzanowski; Pamela Timothy; Molly McAfee; Martin McDaniel; David Meyer; Vilma Torres
Journal:  J Interv Card Electrophysiol       Date:  2004-10       Impact factor: 1.900

4.  Higher defibrillation threshold in methamphetamine cardiomyopathy patients with implantable cardioverter-defibrillator.

Authors:  Rakesh Malhotra; Shyam Patel; Tekchand Ramchand; Omar Al Nimri
Journal:  Indian Pacing Electrophysiol J       Date:  2017-07-08

Review 5.  Non-traditional implantable cardioverter-defibrillator configurations and insertion techniques: a review of contemporary options.

Authors:  Johanna B Tonko; Christopher A Rinaldi
Journal:  Europace       Date:  2022-02-02       Impact factor: 5.214

6.  Implantation of additional subcutaneous array electrode reduces defibrillation threshold in ICD patients - preliminary results.

Authors:  Maciej Kempa; Szymon Budrejko; Lukasz Drelich; Tomasz Królak; Grzegorz Raczak; Dariusz Kozłowski
Journal:  Arch Med Sci       Date:  2013-06-21       Impact factor: 3.318

7.  Incidence of ineffective safety margin testing (<10 J) and efficacy of routine subcutaneous array insertion during implantable cardioverter defibrillator implantation.

Authors:  Marc-Alexander Ohlow; Marcus Roos; Bernward Lauer; J Christoph Geller
Journal:  Indian Pacing Electrophysiol J       Date:  2016-02-26
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.