| Literature DB >> 23845013 |
Seth J Seidman, Joshua W Guag.
Abstract
BACKGROUND: The use of radiofrequency identification (RFID) in healthcare is increasing and concerns for electromagnetic compatibility (EMC) pose one of the biggest obstacles for widespread adoption. Numerous studies have documented that RFID can interfere with medical devices. The majority of past studies have concentrated on implantable medical devices such as implantable pacemakers and implantable cardioverter defibrillators (ICDs). This study examined EMC between RFID systems and non-implantable medical devices.Entities:
Mesh:
Year: 2013 PMID: 23845013 PMCID: PMC3716957 DOI: 10.1186/1475-925X-12-71
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Figure 1RFID reader orientations tested: (a) horizontal orientation; (b) vertical orientation.
List of EMI observed
| Device takes unusually long time to turn on | |
| Minor malfunction of display. The display is still comprehensible. | probably not |
| Audible noise from the device (e.g. speaker hum or buzz) | clinically |
| Device inappropriately alarmed | significant |
| Unable to turn the device on | |
| Occasionally unable to turn the device on | |
| Device permanently damaged | |
| Unable to turn off the device once turned on | |
| Unable to program the device | device |
| Unable to pump | inoperable |
| Major malfunction of display. The display is incomprehensible. | |
| AED could not detect patient | |
| AED sensed false motion | |
| AED inappropriately did not advise shock for VFIB (Ventricular fibrillation) patient input | |
| Occasionally, AED inappropriately did not advise shock for VFIB patient input | probably |
| AED inappropriately advised shock for NSR (Normal Sinus Rhythm) patient input | clinically |
| AED advised shock for VFIB patient input, but failed to deliver. Returned to analysis mode | significant |
Test data for syringe pumps
| Syringe Pump #1 (1994) | Occurrences of EMI | 4 | 10 | 0 | 10 (50%) |
| Max Distance of EMI (cm) | 23 | 40 | - | 40 | |
| RFID Reader Code causing EMI | 5,8,10,11 | 5,7-11,15-18 | N/A | 5,7-11,15-18 | |
| Syringe Pump #2 (2004) | Occurrences of EMI | 3 | 2 | 0 | 3 (15%) |
| Max Distance of EMI (cm) | 59 | 59 | - | 59 | |
| RFID Reader Code causing EMI | 15-17 | 15,17 | N/A | 15-17 | |
| Syringe Pump #3 (2002) | Occurrences of EMI | 0 | 0 | 0 | 0 (0%) |
| Max Distance of EMI (cm) | - | - | - | - | |
| RFID Reader Code causing EMI | N/A | N/A | N/A | N/A |
The EMI Occurrences row represents the number of times EMI occurred during a test (n = 20). One experiment was defined as all testing performed on one medical device with one RFID system. Therefore the All Experiments column indicates whether any EMI occurred for a particular RFID system and does not indicate whether multiple effects were caused by the same RFID system. Production years for devices tested in parenthesis.
Test data for infusion pumps
| Infusion Pump #1 (1995) | Occurrences of EMI | 0 | 4 | 0 | 4 (20%) |
| Max Distance of EMI (cm) | - | 68 | - | 68 | |
| RFID Reader Code causing EMI | N/A | 8,15,17,18 | N/A | 8,15,17,18 | |
| Infusion Pump #2 (1996) | Occurrences of EMI | 1 | 4 | 0 | 5 (25%) |
| Max Distance of EMI (cm) | 9 | 136 | - | 136 | |
| RFID Reader Code causing EMI | 5 | 15-18 | N/A | 5,15-18 | |
| Infusion Pump #3 (2002) | Occurrences of EMI | 1 | 0 | 0 | 1 (5%) |
| Max Distance of EMI (cm) | 12 | - | - | 12 | |
| RFID Reader Code causing EMI | 5 | N/A | N/A | 5 |
The EMI Occurrences row represents the number of times EMI occurred during a test (n = 20). One experiment was defined as all testing performed on one medical device with one RFID system. Therefore the All Experiments column indicates whether any EMI occurred for a particular RFID system and does not indicate whether multiple effects were caused by the same RFID system. Production years for devices tested in parenthesis.
Test data for AEDs
| AED #1 (2008) | Occurrences of EMI | 3 | 0 | 0 | 3 (15%) |
| Max Distance of EMI (cm) | 21 | - | - | 21 | |
| RFID Reader Code causing EMI | 15,17,18 | N/A | N/A | 15,17,18 | |
| AED #2 (2002) | Occurrences of EMI | 1 | 3 | 2 | 3 (15%) |
| Max Distance of EMI (cm) | 11 | 10 | 18 | 18 | |
| RFID Reader Code causing EMI | 17 | 3,5,17 | 3,5 | 3,5,17 | |
| AED #3 (2009) | Occurrences of EMI | 2 | 4 | 4 | 8 (40%) |
| Max Distance of EMI (cm) | 14 | 51 | 18 | 51 | |
| RFID Reader Code causing EMI | 15,17 | 7,8,10,11 | 3,5,7,17 | 3,5,7,8,10,11,15,17 | |
| AED #4 (2004) | Occurrences of EMI | 2 | 1 | 1 | 4 (20%) |
| Max Distance of EMI (cm) | 38 | 0 | 1 | 38 | |
| RFID Reader Code causing EMI | 15,16 | 17 | 10 | 10,15-17 |
The EMI Occurrences row represents the number of times EMI occurred during a test (n = 20). One experiment was defined as all testing performed on one medical device with one RFID system. Therefore the All Experiments column indicates whether any EMI occurred for a particular RFID system and does not indicate whether multiple effects were caused by the same RFID system. Production years for devices tested in parenthesis.
RFID system characteristics
| 1 | Rectangular Loop 3.5 × 4.5 × 0.5 | 0.125 | CW | | -- | -- | 0 | ||
| 2 | Rectangular Loop 114 × 66 × 6.3 | 0.125 | NA | 13 | -- | -- | 0 | ||
| 3 | Rectangular Loop 85 × 50 × 5 | 0.134 | 14.3 | 0.72 | 49.9 | 93 | -- | 11785 | 2 |
| 4 | Rectangular Loop 85 × 50 × 5 | 0.134 | CW | 47 | -- | -- | 0 | ||
| 5 | Rectangular Loop 20 × 20 × 2.5 | 0.134 | 10.6 | 0.54 | 50.3 | 119 | -- | 11785 | 4 |
| 6 | Rectangular Loop 31 × 31 × 2.8 | 13.56 | 10.9 | 0.11 | 10.3 | 1 | 16 | 18000-3 mode 1 | 0 |
| 7 | Rectangular Loop 20 × 20 × 0.8 | 13.56 | 4.0 | 0.13 | 31.9 | 1 | 60 | 18000-3 mode 1 | 2 |
| 8 | Rectangular Loop 31 × 31 × 2.8 | 13.56 | 11.1 | 1.0 | 90.0 | 6 | 67 | 18000-3 mode 1 | 3 |
| 9 | Patch 2.3 × 2.5 × 0.1 | 13.56 | 2.6 | 0.95 | 362.6 | 2 | | 18000-3 | 1 |
| 10 | Patch 21 × 32 × 1.2 | 13.56 | 1.0 | 0.04 | 50.5 | 3 | 378 | 18000-3 mode 2 | 3 |
| 11 | Handheld 19 × 11 × 7.8 | 13.56 | 3.5 | 0.92 | 264.0 | 4 | 71 | 18000-3 mode 1 | 2 |
| 12 | Stick length 19.8 × diameter 1.4 | 433 | NA | -- | 0* | -- | 0 | ||
| 13 | Patch 38 × 36.5 × 1.5 | 433 | NA | -- | 0* | -- | 0 | ||
| 14 | Patch 15.7 × 5.5 × 3 | 433 | NA | -- | 0* | 18000-7 | 0 | ||
| 15 | Patch 31 × 31 × 4.8 | 915 | 56.1 k | 0.76 | 0.01 | -- | 59 | 18000-6B | 7 |
| 16 | Patch 48.5 × 16 × 5 | 915 | NA | -- | 33 | 18000-6B | 4 | ||
| 17 | Patch 21 × 21 × 3.5 | 915 | NA | -- | 89 | 18000-6 | 8 | ||
| 18 | Patch 22.5 × 21 × 5 | 915 | NA | -- | 73 | 18000-6C | 4 | ||
| 19 | Stick length 10.5 diameter 0.9 | 2450 | NA | -- | 1 | -- | 0 | ||
| 20 | Stick length 11 diameter 0.8 | 2450 | 897 | 0.16 | 0.18 | -- | 2 | -- | 0 |
This table describes all RFID equipment tested. It describes the carrier frequency, magnetic field strength, electric field strength, the applicable ISO RFID standard, and the number of occurrences of EMI. *RFID Readers 12 and 13 only receive 433 MHz (no transmit capability). RFID #14 is the active tag.
Figure 2Percentage of medical devices experiencing EMI for each device category at different RFID frequency ranges. LF = 134 kHz, HF = 13.56 MHz, UHF = 915 MHz. No EMI was observed at 433 MHz or 2.45 GHz. No EMI was observed for the one a ventilator tested.
Figure 3Maximum distances where EMI was observed for each device category at different RFID carrier frequencies. LF = 134 kHz, HF = 13.56 MHz, UHF = 915 MHz. No EMI was observed at 433 MHz or 2.45 GHz. No EMI was observed for the one ventilator tested.