Electromagnetic interference from radio frequency identification inducing potentially hazardous incidents in critical care medical equipment.

Remko van der Togt1, Erik Jan van Lieshout, Reinout Hensbroek, E Beinat, J M Binnekade, P J M Bakker.   

Abstract

CONTEXT: Health care applications of autoidentification technologies, such as radio frequency identification (RFID), have been proposed to improve patient safety and also the tracking and tracing of medical equipment. However, electromagnetic interference (EMI) by RFID on medical devices has never been reported.
OBJECTIVE: To assess and classify incidents of EMI by RFID on critical care equipment. DESIGN AND
SETTING: Without a patient being connected, EMI by 2 RFID systems (active 125 kHz and passive 868 MHz) was assessed under controlled conditions during May 2006, in the proximity of 41 medical devices (in 17 categories, 22 different manufacturers) at the Academic Medical Centre, University of Amsterdam, Amsterdam, The Netherlands. Assessment took place according to an international test protocol. Incidents of EMI were classified according to a critical care adverse events scale as hazardous, significant, or light.
RESULTS: In 123 EMI tests (3 per medical device), RFID induced 34 EMI incidents: 22 were classified as hazardous, 2 as significant, and 10 as light. The passive 868-MHz RFID signal induced a higher number of incidents (26 incidents in 41 EMI tests; 63%) compared with the active 125-kHz RFID signal (8 incidents in 41 EMI tests; 20%); difference 44% (95% confidence interval, 27%-53%; P < .001). The passive 868-MHz RFID signal induced EMI in 26 medical devices, including 8 that were also affected by the active 125-kHz RFID signal (26 in 41 devices; 63%). The median distance between the RFID reader and the medical device in all EMI incidents was 30 cm (range, 0.1-600 cm).
CONCLUSIONS: In a controlled nonclinical setting, RFID induced potentially hazardous incidents in medical devices. Implementation of RFID in the critical care environment should require on-site EMI tests and updates of international standards.

Entities:  

Mesh:

Year:  2008        PMID: 18577733     DOI: 10.1001/jama.299.24.2884

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  25 in total

1.  Deployment of RFID in healthcare facilities-experimental design in MRI department.

Authors:  Chen-Yang Cheng; Jyh-Wen Chai
Journal:  J Med Syst       Date:  2011-11-10       Impact factor: 4.460

2.  Online recognition of surgical instruments by information fusion.

Authors:  Thomas Neumuth; Christian Meissner
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-10-18       Impact factor: 2.924

3.  Towards a smart object network for clinical services.

Authors:  Martin Sedlmayr; Andreas Becker; Ulli Muench; Fritz Meier; Hans-Ulrich Prokosch; Thomas Ganslandt
Journal:  AMIA Annu Symp Proc       Date:  2009-11-14

4.  Evaluation of Electromagnetic Fields in a Hospital for Safe Use of Electronic Medical Equipment.

Authors:  Kai Ishida; Tomomi Fujioka; Tetsuo Endo; Ren Hosokawa; Tetsushi Fujisaki; Ryoji Yoshino; Minoru Hirose
Journal:  J Med Syst       Date:  2015-12-07       Impact factor: 4.460

Review 5.  The adoption and implementation of RFID technologies in healthcare: a literature review.

Authors:  Wen Yao; Chao-Hsien Chu; Zang Li
Journal:  J Med Syst       Date:  2011-10-19       Impact factor: 4.460

6.  Radiofrequency identification: exploiting an old technology for measuring nurse time and motion.

Authors:  Terry L Jones
Journal:  Comput Inform Nurs       Date:  2012-09       Impact factor: 1.985

7.  Adding intelligence to mobile asset management in hospitals: the true value of RFID.

Authors:  Linda Castro; Elisabeth Lefebvre; Louis A Lefebvre
Journal:  J Med Syst       Date:  2013-08-13       Impact factor: 4.460

8.  Development of a novel marking system for laparoscopic gastrectomy using endoclips with radio frequency identification tags: feasibility study in a canine model.

Authors:  Fumitsugu Kojima; Toshihiko Sato; Shigeru Tsunoda; Hiromi Takahata; Masatsugu Hamaji; Teruya Komatsu; Minoru Okada; Tadao Sugiura; Osamu Oshiro; Yoshiharu Sakai; Hiroshi Date; Tatsuo Nakamura
Journal:  Surg Endosc       Date:  2014-03-21       Impact factor: 4.584

Review 9.  Evaluation of current literature to determine the potential effects of radio frequency identification on technology used in diabetes care.

Authors:  Barbara Christe
Journal:  J Diabetes Sci Technol       Date:  2009-03-01

10.  Designing an artificial pancreas system to be compatible with other medical devices.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2008-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.