In vitro tests reveal sample radiofrequency identification readers inducing clinically significant electromagnetic interference to implantable pacemakers and implantable cardioverter-defibrillators.

Seth J Seidman1, Randall Brockman, Brian Marc Lewis, Joshua Guag, Mitchell J Shein, Wesley J Clement, James Kippola, Dennis Digby, Catherine Barber, Dan Huntwork.   

Abstract

BACKGROUND: The use of radiofrequency identification (RFID) systems is expanding and highlights the need to address electromagnetic interference (EMI) to implantable pacemakers and implantable cardioverter-defibrillators (ICDs).
OBJECTIVE: This study sought to examine the electromagnetic compatibility (EMC) between RFID readers and implantable pacemakers or ICDs.
METHODS: During in vitro testing, 15 implantable pacemakers and 15 ICDs were exposed to 13 passive RFID readers in 3 frequency bands: 134 kHz (low frequency [LF]), 13.56 MHz (high frequency [HF]), and 915 MHz (ultra high frequency [UHF]).
RESULTS: While being exposed to LF RFID, a reaction was observed for 67% of all pacemaker tests (maximum distance 60 cm) and 47% of all ICD tests (maximum distance 40 cm). During HF RFID exposure, a reaction was observed for 6% of all pacemaker tests (maximum distance 22.5 cm) and 1% of all ICD tests (maximum distance 7.5 cm). For both pacemakers and ICDs, no reactions were observed during exposure to UHF RFID or continuous-wave RFID. Pacemakers and ICDs were most susceptible to modulated LF RFID readers.
CONCLUSION: Although there is in vitro testing evidence for concern for implantable pacemaker and ICD EMI at LF and HF, the FDA has not received any incident reports of pacemaker or ICD EMI caused by any RFID system. We do not believe the current situation reveals an urgent public health risk.

Entities:  

Mesh:

Year:  2009        PMID: 20129290     DOI: 10.1016/j.hrthm.2009.09.071

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  10 in total

1.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

2.  Electromagnetic compatibility of implantable neurostimulators to RFID emitters.

Authors:  Oxana S Pantchenko; Seth J Seidman; Joshua W Guag; Donald M Witters; Curt L Sponberg
Journal:  Biomed Eng Online       Date:  2011-06-09       Impact factor: 2.819

3.  Analysis of induced electrical currents from magnetic field coupling inside implantable neurostimulator leads.

Authors:  Oxana S Pantchenko; Seth J Seidman; Joshua W Guag
Journal:  Biomed Eng Online       Date:  2011-10-21       Impact factor: 2.819

4.  Assessment of Electromagnetic Interference with Active Cardiovascular Implantable Electronic Devices (CIEDs) Caused by the Qi A13 Design Wireless Charging Board.

Authors:  Tobias Seckler; Kai Jagielski; Dominik Stunder
Journal:  Int J Environ Res Public Health       Date:  2015-05-27       Impact factor: 3.390

5.  Application safety evaluation of the radio frequency identification tag under magnetic resonance imaging.

Authors:  Xiaolu Fei; Shanshan Li; Shan Gao; Lan Wei; Lihong Wang
Journal:  Biomed Eng Online       Date:  2014-09-04       Impact factor: 2.819

6.  Personal medical electronic devices and walk-through metal detector security systems: assessing electromagnetic interference effects.

Authors:  Joshua Guag; Bisrat Addissie; Donald Witters
Journal:  Biomed Eng Online       Date:  2017-03-20       Impact factor: 2.819

7.  LBTM: Listen-before-Talk Protocol for Multiclass UHF RFID Networks.

Authors:  Pablo López-Matencio; Javier Vales-Alonso; Juan J Alcaraz
Journal:  Sensors (Basel)       Date:  2020-04-18       Impact factor: 3.576

8.  Adhoc electromagnetic compatibility testing of non-implantable medical devices and radio frequency identification.

Authors:  Seth J Seidman; Joshua W Guag
Journal:  Biomed Eng Online       Date:  2013-07-11       Impact factor: 2.819

9.  Feasibility results of an electromagnetic compatibility test protocol to evaluate medical devices to radio frequency identification exposure.

Authors:  Seth J Seidman; Omar Bekdash; Joshua Guag; Maryam Mehryar; Paul Booth; Paul Frisch
Journal:  Biomed Eng Online       Date:  2014-08-03       Impact factor: 2.819

10.  Electromagnetic interference in cardiac electronic implants caused by novel electrical appliances emitting electromagnetic fields in the intermediate frequency range: a systematic review.

Authors:  Sarah Driessen; Andreas Napp; Kristina Schmiedchen; Thomas Kraus; Dominik Stunder
Journal:  Europace       Date:  2019-02-01       Impact factor: 5.214

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.