Literature DB >> 23822423

Improving IMRT delivery efficiency with reweighted L1-minimization for inverse planning.

Hojin Kim1, Stephen Becker, Rena Lee, Soonhyouk Lee, Sukyoung Shin, Emmanuel Candès, Lei Xing, Ruijiang Li.   

Abstract

PURPOSE: This study presents an improved technique to further simplify the fluence-map in intensity modulated radiation therapy (IMRT) inverse planning, thereby reducing plan complexity and improving delivery efficiency, while maintaining the plan quality.
METHODS: First-order total-variation (TV) minimization (min.) based on L1-norm has been proposed to reduce the complexity of fluence-map in IMRT by generating sparse fluence-map variations. However, with stronger dose sparing to the critical structures, the inevitable increase in the fluence-map complexity can lead to inefficient dose delivery. Theoretically, L0-min. is the ideal solution for the sparse signal recovery problem, yet practically intractable due to its nonconvexity of the objective function. As an alternative, the authors use the iteratively reweighted L1-min. technique to incorporate the benefits of the L0-norm into the tractability of L1-min. The weight multiplied to each element is inversely related to the magnitude of the corresponding element, which is iteratively updated by the reweighting process. The proposed penalizing process combined with TV min. further improves sparsity in the fluence-map variations, hence ultimately enhancing the delivery efficiency. To validate the proposed method, this work compares three treatment plans obtained from quadratic min. (generally used in clinic IMRT), conventional TV min., and our proposed reweighted TV min. techniques, implemented by a large-scale L1-solver (template for first-order conic solver), for five patient clinical data. Criteria such as conformation number (CN), modulation index (MI), and estimated treatment time are employed to assess the relationship between the plan quality and delivery efficiency.
RESULTS: The proposed method yields simpler fluence-maps than the quadratic and conventional TV based techniques. To attain a given CN and dose sparing to the critical organs for 5 clinical cases, the proposed method reduces the number of segments by 10-15 and 30-35, relative to TV min. and quadratic min. based plans, while MIs decreases by about 20%-30% and 40%-60% over the plans by two existing techniques, respectively. With such conditions, the total treatment time of the plans obtained from our proposed method can be reduced by 12-30 s and 30-80 s mainly due to greatly shorter multileaf collimator (MLC) traveling time in IMRT step-and-shoot delivery.
CONCLUSIONS: The reweighted L1-minimization technique provides a promising solution to simplify the fluence-map variations in IMRT inverse planning. It improves the delivery efficiency by reducing the entire segments and treatment time, while maintaining the plan quality in terms of target conformity and critical structure sparing.

Entities:  

Mesh:

Year:  2013        PMID: 23822423      PMCID: PMC3702602          DOI: 10.1118/1.4811100

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  33 in total

1.  Intensity modulated photon beams subject to a minimal surface smoothing constraint.

Authors:  M Alber; F Nüsslin
Journal:  Phys Med Biol       Date:  2000-05       Impact factor: 3.609

2.  A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note.

Authors:  I Paddick
Journal:  J Neurosurg       Date:  2000-12       Impact factor: 5.115

3.  Leaf sequencing techniques for MLC-based IMRT.

Authors:  C B Saw; R A Siochi; K M Ayyangar; W Zhen; C A Enke
Journal:  Med Dosim       Date:  2001       Impact factor: 1.482

4.  Leaf sequencing with secondary beam blocking under leaf positioning constraints for continuously modulated radiotherapy beams.

Authors:  H G Kuterdem; P S Cho
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

5.  Smoothing intensity-modulated beam profiles to improve the efficiency of delivery.

Authors:  S V Spirou; N Fournier-Bidoz; J Yang; C S Chui; C C Ling
Journal:  Med Phys       Date:  2001-10       Impact factor: 4.071

6.  Direct aperture optimization: a turnkey solution for step-and-shoot IMRT.

Authors:  D M Shepard; M A Earl; X A Li; S Naqvi; C Yu
Journal:  Med Phys       Date:  2002-06       Impact factor: 4.071

7.  Smoothing intensity-modulated treatment delivery under hardware constraints.

Authors:  Lijun Ma
Journal:  Med Phys       Date:  2002-12       Impact factor: 4.071

8.  Use of a quantitative index of beam modulation to characterize dose conformality: illustration by a comparison of full beamlet IMRT, few-segment IMRT (fsIMRT) and conformal unmodulated radiotherapy.

Authors:  S Webb
Journal:  Phys Med Biol       Date:  2003-07-21       Impact factor: 3.609

9.  Optimization of intensity-modulated radiotherapy plans based on the equivalent uniform dose.

Authors:  Qiuwen Wu; Radhe Mohan; Andrzej Niemierko; Rupert Schmidt-Ullrich
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-01-01       Impact factor: 7.038

10.  Quantifying the degree of conformity in radiosurgery treatment planning.

Authors:  Nicoletta J Lomax; Stefan G Scheib
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-04-01       Impact factor: 7.038

View more
  2 in total

1.  Sequentially reweighted TV minimization for CT metal artifact reduction.

Authors:  Xiaomeng Zhang; Lei Xing
Journal:  Med Phys       Date:  2013-07       Impact factor: 4.071

2.  Nonisocentric treatment strategy for breast radiation therapy: a proof of concept study.

Authors:  Ruijiang Li; Lei Xing; Kathleen C Horst; Karl Bush
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-03-15       Impact factor: 7.038

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.