Abstract
Whether failure in unicompartmental knee arthroplasty (UKA) is related to implant design remains unclear. We hypothesize that current available UKAs fit within 2 mm. Forty-eight CTs of cadaveric knees were compared to current available UKA brands. Overall no-fit compared to at least one component within 2 mm is high (91.7%) and worse for males (100%) compared to females (833%). Good fit was observed for the medial but not for the lateral tibia plateau. Seven males (29.2%) had larger dimensions of more than 2 mm. For the widest UKA brand, 12 (57%) males and 2 females (8. 3%) had lateral femoral condyles 3 mm larger. Current UKA's in our sample population fit less on the lateral tibia and on femoral condyles.Entities:
Mesh:
Year: 2013 PMID: 23821971 PMCID: PMC4058334