W W Hack1, N Westerhof, T Leenhoven, A Okken.
Abstract
A technique is described for accurate measurement of intraarterial pressure through radial artery catheters in neonates. The technique, which can be used for short-term monitoring, uses cannulation of the radial artery with a 24-gauge Teflon catheter, connected by a Luer-Lok fitting to a three-way stopcock and a high-fidelity tip transducer. In vitro studies showed that the system is linear and the frequency response is flat (+/- 3 dB) up to 50 Hz. The technique permits gathering of high-quality pressure data and can be used in the area of neonatal clinical research for short-term monitoring. It needs to be developed further before routine application in clinical practice can be recommended.Entities:
Mesh:
Year: 1990 PMID: 2380752 DOI: 10.1007/bf02832149
Source DB: PubMed Journal: J Clin Monit ISSN: 0748-1977