Petra S Kroon, Sandra Hol, Marion Essers.
Abstract
INTRODUCTION: The main aim of the current study was to assess the dosimetric accuracy and clinical quality of volumetric modulated arc therapy (VMAT) plans for stereotactic (stage I) and conventional (stage III) lung cancer treatments planned with Eclipse version 10.0 Anisotropic Analytical Algorithm (AAA) and Acuros XB (AXB) algorithm.Entities:
Mesh:
Year: 2013 PMID: 23800024 PMCID: PMC3723919 DOI: 10.1186/1748-717X-8-149
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Measured and calculated percentage depth dose curves in a slab-phantom. Predicted percentage depth dose curves by AAA and Acuros XB grid sizes 2.5 and 1.0 mm compared to measured data using a slab-phantom with (top) 1.0x1.0 cm2 and (bottom) 4.0x4.0 cm2 6 MV AP photon beams.
Relative differences between algorithms and grid sizes for lung cancer patients
| | ||||
|---|---|---|---|---|
| | | |||
| PTV | ||||
| | −3.2% ± 4.0% (−12.3% – 0.5%) | 0.06 | 0.7% ± 1.3% (−0.8% – 3.1%) | 0.19 |
| | 0.2% ± 1.2% (−2.1% – 1.3%) | 0.69 | 0.9% ± 0.4% (0.3% – 1.4%) | <0.01* |
| | −0.6% ± 2.2% (−4.9% – 1.6%) | 0.46 | 0.2% ± 0.5% (−0.6% – 1.2%) | 0.45 |
| Total lung | ||||
| | 1.3% ± 1.9% (−1.9% –3.2%) | 0.12 | 0.1% ± 0.6% (−0.8% – 1.1%) | 0.73 |
| | 2.0% ± 2.3% (0.0% – 5.7%) | 0.05* | −0.7% ± 1.2% (−2.7% – 0.0%) | 0.17 |
| MLD | 0.0% ± 0.5% (−0.7% – 0.7%) | 0.60 | 0.1% ± 0.2% (0.0% – 0.4%) | 0.08 |
| Time | ||||
| Calculation | −39.2% ± 13.5% (−50.3% – –8.2%) | <0.01* | 304.7% ± 48.7% (226.9% – 345.8%) | <0.01* |
| PTV | ||||
| | −0.3% ± 0.7% (−0.8% – 1.3%) | 0.33 | 0.2% ± 0.8% (−1.4% – 0.9%) | 0.57 |
| | −0.6% ± 2.0% (−2.1% – 3.0%) | 0.47 | 0.6% ± 0.4% (−0.1% – 1.0%) | <0.01* |
| | −0.8% ± 0.7% (−1.6% – 0.3%) | 0.02* | 0.1% ± 0.3% (−0.5% – 0.4%) | 0.40 |
| Total lung | ||||
| | −1.8% ± 1.7% (−4.8% – 0.3%) | 0.03* | −0.4% ± 0.7% (−1.5% – 0.4%) | 0.16 |
| | 0.2% ± 0.7% (−0.5% – 1.5%) | 0.69 | 0.0% ± 0.0% (0.0% –0.0%) | 1.00 |
| MLD | −0.5% ± 0.6% (−1.3% – 0.0%) | 0.09 | 0.2% ± 0.3% (0.0% – 0.6% ) | 0.17 |
| Time | ||||
| Calculation | −31.4% ± 9.9% (−44.4% – –17.9%) | <0.01* | 556.9% ± 91.1% (412.3% – 646.0%) | <0.01* |
*Significant relative difference tested by two-tailed student t-test.
Figure 2Dose volume histogram comparisons of PTV doses between algorithms and grid sizes. Comparison of dose volume histograms of the planning target volume (PTV) between AAA grid size 2.5 mm, Acuros XB grid size 2.5 mm and Acuros XB grid size 1.0 mm of (top) a stage I and (bottom) a stage III lung patient. The black vertical lines indicate the prescription: The dose that has to cover 99% of the PTV, respectively, 54.0 Gy for stage I and 59.4 Gy for stage III.
Comparison of stage III VMAT and hybrid-VMAT plans
| De Bree-Balk et al. [ | VMAT | 20 | 66 | 8385 | 20.0 | 36.6 | NA | 69.6 |
| Verbakel et al. [ | VMAT | 14 | 66 | 779 | NA | 30.3 ± 5.7 | NA | 44.6 ± 9.0 |
| | H-VMAT | 14 | 66 | 779 | NA | 30.1 ± 5.8 | NA | 36.2 ± 15.0 |
| Chan et al. [ | VMAT | 24 | 60 | 508 | 14.4 ± 2.9 | 25.4 ± 6.0 | 64.0 ± 15.4 | NA |
| | H-VMAT | 24 | 60 | 508 | 14.0 ± 2.9 | 23.3 ± 5.3 | 59.5 ± 16.7 | NA |
| This study4 | VMAT | 73 | 66 | 344 | 12.4 ± 3.5 | 19.3 ± 6.8 | 46.6 ± 10.6 | 35.6 ± 7.1 |
| This study4 | VMAT | 13 | 66 | 678 | 14.5 ± 2.1 | 22.1 ± 6.4 | 52.1 ± 10.3 | 38.7 ± 10.0 |
Average values and standard deviations are given.
1[14] Use the volume of both lungs minus CTV for MLD and V20Gy; PTV V95% > 99% and PTV V107% < 1%.
2[15] Use the volume of both lungs minus PTV for V20Gy; PTV V95% > 97% and PTV V107% < 5%.
3[16] Use the volume of both lungs minus PTV for MLD, V20Gy and V5Gy; D98% > 57.0 Gy and D2% < 64.2 Gy.
4 This study uses the volume of both lungs minus ITV for MLD, and volume of both lungs minus PTV for V20Gy and V5Gy.
PTV V90% > 99% and PTV V110% ≤ 1%.
5 Median instead of average value is given.