| Literature DB >> 23767816 |
Simon Finfer, Jan Wernerman, Jean-Charles Preiser, Tony Cass, Thomas Desaive, Roman Hovorka, Jeffrey I Joseph, Mikhail Kosiborod, James Krinsley, Iain Mackenzie, Dieter Mesotten, Marcus J Schultz, Mitchell G Scott, Robbert Slingerland, Greet Van den Berghe, Tom Van Herpe.
Abstract
The management reporting and assessment of glycemic control lacks standardization. The use of different methods to measure the blood glucose concentration and to report the performance of insulin treatment yields major disparities and complicates the interpretation and comparison of clinical trials. We convened a meeting of 16 experts plus invited observers from industry to discuss and where possible reach consensus on the most appropriate methods to measure and monitor blood glucose in critically ill patients and on how glycemic control should be assessed and reported. Where consensus could not be reached, recommendations on further research and data needed to reach consensus in the future were suggested. Recognizing their clear conflict of interest, industry observers played no role in developing the consensus or recommendations from the meeting. Consensus recommendations were agreed for the measurement and reporting of glycemic control in clinical trials and for the measurement of blood glucose in clinical practice. Recommendations covered the following areas: How should we measure and report glucose control when intermittent blood glucose measurements are used? What are the appropriate performance standards for intermittent blood glucose monitors in the ICU? Continuous or automated intermittent glucose monitoring - methods and technology: can we use the same measures for assessment of glucose control with continuous and intermittent monitoring? What is acceptable performance for continuous glucose monitoring systems? If implemented, these recommendations have the potential to minimize the discrepancies in the conduct and reporting of clinical trials and to improve glucose control in clinical practice. Furthermore, to be fit for use, glucose meters and continuous monitoring systems must match their performance to fit the needs of patients and clinicians in the intensive care setting.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23767816 PMCID: PMC3706766 DOI: 10.1186/cc12537
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Figure 1Dendrogram of measures of central tendency, dispersion and hypoglycemia. See [3] for a full explanation. Reproduced with permission.
Figure 2Schematic representation of the performance of continuous blood glucose concentration. Area A, possible measure of hyperglycemia (when defined as blood glucose concentration >10.0 mmol/l). Area B, possible measure of moderate hypoglycemia (when defined as blood glucose concentration <4.0 mmol/l).
Metrics to report glycemic control using continuous glucose monitoring systems
| 1. Measures of central tendency |
| a. Mean blood glucose per patient, per day of ICU stay |
| b. For populations of patients, the median and interquartile range of individual patient means should be reported |
| c. For comparison of glycemic control achieved with intermittent measurement versus continuous measurement, two groups should be studied in a randomized controlled trial, both groups should have blood glucose concentration recorded by a CGM but the output of the CGM should be masked in the control group where blood glucose concentration is managed using intermittent measurement of blood glucose concentration |
| 2. Measures of variability and dispersion |
| a. Standard deviation and coefficient of variation of blood glucose (for comparison of glycemic control with intermittent measurement, two groups should be studied - both groups should have blood glucose concentration recorded by a CGM but the output of the CGM be masked in the group where the blood glucose concentration is managed using intermittent measurement of blood glucose concentration) |
| b. Peak blood glucose concentration reached within a set time period after correction of hypoglycemia |
| c. Analysis of the rate of change of the blood glucose concentration |
| 3. Hypoglycemia and lowest blood glucose concentration |
| a. Lowest blood glucose concentration recorded per patient |
| b. Number of times a blood glucose concentration defining hypoglycemia or severe hypoglycemia was recorded |
| c. Area above the curve under the target for mild and severe hypoglycemia |
| d. Duration of time with blood glucose below the concentration defining hypoglycemia or severe hypoglycemia |
| 4. Range and exposure measures |
| a. Percentage of time the blood glucose concentration is in the target range |
| b. Percentage of time the blood glucose concentration is outside a nominated target range |
| c. Area under the curve above the upper target for hyperglycemia |
| d. Area above the curve but under the target for mild and severe hypoglycemia |
| Desirable performance standards for continuous glucose monitoring systems - the meeting accepted that these were desirable performance standards that might not be currently achievable |
| There was acceptance but not complete agreement that desirable point accuracy might vary depending on whether CGMs were intended to operate alone and be used to adjust insulin dosing or be used to only as a warning system to advise clinicians when to check blood glucose on a reference system |
| 1. Desirable features with regard to set-up, calibration and integration with standard ICU care |
| a. <20 minutes to set by a nurse, technician, or physician |
| b. <10 minutes required for initial and subsequent calibration |
| c. <20 minutes required for sensor insertion |
| d. Calibration against a reference standard to maintain point accuracy is required no more frequently than every 8 hours and preferably no more than twice per 24 hours |
| e. CGM requires few nursing interventions per 24 hours to produce a near-continuous glucose datastream |
| f. CGM sensor inserted into an arterial or central venous catheter should not adversely affect blood sampling or monitoring of cardiovascular system or increase the frequency with which such monitoring lines occlude |
| g. Incidence infection, hematoma, tissue ischemia, and/or thromboembolism should not exceed that which occurs in usual clinical practice in the absence of a CGM |
| 2. Desirable reliability |
| a. CGM should continuously measure glucose and display in real time >95% of the time for the duration of time specified in the product label (2 to 7 days) |
| b. Skips in data acquisition due to system faults or failures should not exceed 30 minutes at a time (data may be missing for other reasons; for example, patient transports) |
| c. The CGM should have an internal mechanism that prevents the display or reporting of erroneous or spurious data |
| 3, Desirable point accuracy - this should be the same as for intermittent monitors if the CGM is being used alone to guide clinical management and administration of insulin |
| a. 98% of readings should be within 12.5% of a reference standard (or within 0.55 mmol/l for readings <5.5 mmol/l; the remaining 2% of readings should be within 20% of a reference standard |
| b. CGM sensors inserted in arterial and central venous catheters may transiently indicate false low readings when those catheters are flushed with saline or other glucose-free solutions. The CGM should alert the treating clinicians that rapid reductions in the measured blood glucose concentration may be due to flushing of the arterial or central venous catheter |
| c. CGM sensors inserted in peripheral and central veins may transiently indicate false high readings if venous blood in those veins is contaminated with glucose-containing solutions. The CGM should alert the treating clinicians to rapid increases in the measured blood glucose concentration that may be due to contamination by glucose-containing solutions |
| 4. Rate or trend accuracy - CGMs should be carefully characterized in the intended-use critical care patient population to ensure point accuracy over a wide range of blood glucose rates of change. Rate or trend metrics have not been tested sufficiently to provide definitive guidance |
| 5. Alarms and alerts for hypoglycemia and hyperglycemia alerts |
| a. Real-time blood glucose concentration and trend data should be displayed at the bedside; with visual and audible alerts and alarms for hypoglycemia, hyperglycemia, and rapid rates of change |
| b. The CGM data should be sufficiently frequent, reliable, and accurate for alarm algorithms to detect and/or predict hypoglycemia and hyperglycemia with high sensitivity and specificity |
The meeting recognized that further research is needed to determine which of these metrics are closely associated with mortality and major morbidity when continuous or automated intermittent glucose monitoring systems (CGMs) are used.
Summary of the recommendations
| Blood sampling |
| 1. Patients whose severity of illness justifies invasive vascular monitoring |
| a. All blood samples should be drawn from an arterial line |
| b. If an arterial line is temporarily or permanently unavailable, sample from a venous line |
| c. Capillary (needle stick) samples are inaccurate and should not be used |
| 2. Patients whose severity of illness |
| a. Capillary (needle stick) samples may be used |
| 3. Clinical research papers should report the number and percentage of blood samples obtained from arterial catheters, central and peripheral venous catheters and capillary (needle stick) samples |
| Choice of blood glucose analyzer in clinical research in critical care units |
| a. Samples taken from arterial or central venous catheters should be analyzed in a central laboratory or blood gas analyzer; a blood gas analyzer should be the default analyzer |
| b. Only when capillary samples are taken from patients without invasive vascular monitoring is analysis using a glucose meter acceptable |
| c. Clinical research papers should report the number and percentage of samples analyzed using central laboratory or blood gas analyzers or glucose meters. In all cases, the make and model of the analyzer used should be reported along with routine calibration and quality assurance measures |
| Reporting glycemic control - trials or observational studies should report |
| 1. Central tendency - for blood glucose concentration measurements from a population of patients, the median and interquartile range of individual patient means should be reported |
| 2. Dispersion - calculate the standard deviation of blood glucose concentration for each patient then report the median and interquartile range of standard deviations for the population |
| 3. Hypoglycemia - as a minimum, investigators should report the number and percentage of patients experiencing at least one episode of severe and moderate hypoglycemia (blood glucose concentration ?2.2 (?40 mg/dl) and 2.3 to 3.9 mmol/l (41 to 70 mg/dl) respectively). Report separately the number and percentage of patients experiencing hypoglycemia related to insulin treatment (iatrogenic) and unrelated to insulin treatment (spontaneous) |
| For severe hypoglycemia, report duration of hypoglycemia, associated symptoms, amount of glucose administered, and next blood glucose concentration |
| Blood sampling |
| 1. Patients whose severity of illness justifies invasive vascular monitoring |
| a. All blood samples should be drawn from an arterial line |
| b. If an arterial line is temporarily or permanently unavailable, sample from a venous line |
| c. Capillary (needle stick) samples are inaccurate and should not be used |
| 2. Patients whose severity of illness |
| a. Capillary (needle stick) samples may be used |
| Choice of blood glucose analyzer in clinical practice in critical care units |
| 1. Patients whose severity of illness justifies invasive vascular monitoring |
| a. Samples taken from arterial or central venous catheters should be analyzed in a central laboratory or blood gas analyzer; a blood gas analyzer should be the default analyzer, central laboratory measurements should only be used if results can be obtained without delay |
| 2. Patients whose severity of illness |
| a. Only when capillary samples are taken from patients considered well to need invasive vascular monitoring is analysis using a glucose meter acceptable |
| Accuracy of blood glucose analyzers used in clinical practice in critical care units |
| 1. Patients whose severity of illness justifies invasive vascular monitoring |
| a. Central laboratory analyzers and blood gas analyzers in the ICU should perform to currently acceptable international standards (for example, ±0.4 mmol/l (or ±8% above 5 mmol/l)) |
| 2. Patients whose severity of illness |
| a. The minimum standard for glucose meters to be used in critically ill patients should be that 98% of readings are within 12.5% of a reference standard (or within 0.55 mmol/l for readings <5.5 mmol/l). The remaining 2% of readings should be within 20% of a reference standard |
| Assessment of glycemic control when continuous or automated intermittent blood glucose monitoring used |
| 1. Currently there are few data to guide the choice of appropriate metrics for reporting glycemic control when continuous or automated intermittent blood glucose monitoring is used. There is a need to define measures that are associated with important patient-centered outcomes such as mortality and major morbidity |
| Comparing glycemic control with continuous versus intermittent measurement of blood glucose |
| 1. Comparison of the glycemic control achieved with continuous versus intermittent monitoring must be evaluated in randomized controlled trials with both groups of patients having a continuous monitor but the output from the continuous monitor masked in the control group where blood glucose is managed by intermittent monitoring |
| What are desirable performance standards for continuous glucose monitoring systems? |
| 1. Set-up, calibration and integration with standard ICU care |
| a. <20 minutes to set by a nurse, technician, or physician |
| b. <10 minutes required for initial and subsequent calibration |
| c. <20 minutes required for sensor insertion |
| d. Calibration against a reference standard to maintain point accuracy is required no more frequently than every 8 hours and preferably no more than twice per 24 hours |
| e. CGM requires few nursing interventions per 24 hours to produce a near-continuous datastream |
| f. CGM sensor inserted into an arterial or central venous catheter should not adversely affect blood sampling or monitoring of cardiovascular system or increase the frequency with which such monitoring lines occlued |
| g. Incidence of infection, hematoma, tissue ischemia, and/or thromboembolism should not exceed that which occurs in usual clinical practice in the absence of a CGM |
| 2. Reliability |
| a. CGM should continuously measure glucose and display in real time >95% of the time for the duration of time specified in the product label (2 to 7 days) |
| b. Skips in data acquisition should not exceed 30 minutes at a time |
| c. The CGM should have an internal mechanism that prevents the display or reporting of erroneous or spurious data |
| 3. Point accuracy - this should be the same as for intermittent monitors if the CGM is being used alone to guide clinical management and administration of insulin |
| a. 98% of readings are within 12.5% of a reference standard (or within 0.55 mmol/l for readings <5.5 mmol/l). The remaining 2% of readings should be within 20% of a reference standard |
| b. CGM sensors inserted in arterial and central venous catheters may transiently indicate false low readings when those catheters are flushed with saline or other glucose-free solutions. The CGM should alert the treating clinicians that rapid reductions in the measured blood glucose concentration may be due to flushing of the arterial or central venous catheter |
| c. CGM sensors inserted in peripheral and central veins may transiently indicate false high readings if venous blood in those veins is contaminated with glucose-containing solutions. The CGM should alert the treating clinicians to rapid increases in the measured blood glucose concentration that may be due to contamination by glucose-containing solutions |
| 4. Rate or trend accuracy - trend metrics have not been tested sufficiently to provide definitive guidance or recommendations; this is an area for future research |
| 5. Alarms and alerts for hypoglycemia and hyperglycemia alerts |
| a. Real-time blood glucose concentration and trend data should be displayed at the bedside; with visual and audible alerts and alarms for hypoglycemia, hyperglycemia, and rapid rates of change |
| b. The CGM data should be sufficiently frequent, reliable, and accurate for alarm algorithms to detect and/or predict hypoglycemia and hyperglycemia with high sensitivity and specificity |
Recommendations for conducting and reporting clinical trials and observational studies. CGM, continuous or automated intermittent glucose monitoring system.