Reliability of point-of-care testing for glucose measurement in critically ill adults.

Salmaan Kanji1, Jennifer Buffie, Brian Hutton, Peter S Bunting, Avinder Singh, Kevin McDonald, Dean Fergusson, Lauralyn A McIntyre, Paul C Hebert.   

Abstract

BACKGROUND: Glycemic control is increasingly being recognized as a priority in the treatment of critically ill patients. Titration and monitoring of insulin infusions involve frequent blood glucose measurement to achieve target glucose ranges and prevent adverse events related to hypoglycemia. Therefore, it is imperative that bedside glucose testing methods be safe and accurate.
OBJECTIVE: To determine the accuracy and clinical impact of three common methods of bedside point-of-care testing for glucose measurements in critically ill patients receiving insulin infusions.
DESIGN: Prospective observational study.
SETTING: A 21-bed mixed medical/surgical intensive care unit of a tertiary care teaching hospital. PATIENTS: Thirty consecutive critically ill patients who were vasopressor-dependent (n = 10), had significant peripheral edema (n = 10), or were admitted following major surgery (n = 10). MEASUREMENTS: Findings from three different methods of glucose measurement were compared with central laboratory measurements: (1) glucose meter analysis of capillary blood (fingerstick); (2) glucose meter analysis of arterial blood; and (3) blood gas/chemistry analysis of arterial blood. Patients were enrolled for a maximum of 3 days and had a maximum of nine sets of measurements determined during this time.
RESULTS: Clinical agreement with the central laboratory was significantly better with arterial blood analysis (69.9% and 76.5% for glucose meter and blood gas/chemistry analysis, respectively) than with capillary blood analysis (56.8%; p = .039 and .001, respectively). During hypoglycemia, clinical agreement was only 26.3% with capillary blood analysis and 55.6% and 64.9% for glucose meter and blood gas/chemistry analysis of arterial blood (p = .010 and <.001, respectively). Glucose meter analysis of both arterial and capillary blood tended to provide higher glucose values, whereas blood gas/chemistry analysis of arterial blood tended to yield lower glucose values.
CONCLUSIONS: The magnitude of the differences in the glucose values offered by the four different methods of glucose measurement led to frequent clinical disagreements regarding insulin dose titration in the context of an insulin infusion protocol for aggressive glucose control.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16352960     DOI: 10.1097/01.ccm.0000189939.10881.60

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  113 in total

Review 1.  Hypoglycemia in critically ill children.

Authors:  E Vincent S Faustino; Eliotte L Hirshberg; Clifford W Bogue
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

Review 2.  Blood glucose measurements in critically ill patients.

Authors:  Tom Van Herpe; Dieter Mesotten
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

3.  Assessing inpatient glycemic control: what are the next steps?

Authors:  Curtiss B Cook; Kay E Wellik; Gail L Kongable; Jianfen Shu
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

4.  Are tight glycemic targets achieved through intensive insulin infusion still applicable in the intensive care unit?

Authors: 
Journal:  Can J Hosp Pharm       Date:  2010-01

5.  Glycemic control in the pediatric intensive care unit of Leuven: two years of experience.

Authors:  Tom Van Herpe; Koen Vanhonsebrouck; Dieter Mesotten; Bart De Moor; Greet Van den Berghe
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

6.  Intraoperative accuracy of a point-of-care glucose meter compared with simultaneous central laboratory measurements.

Authors:  Boris Mraovic; Eric S Schwenk; Richard H Epstein
Journal:  J Diabetes Sci Technol       Date:  2012-05-01

7.  Regulatory controversies surround blood glucose monitoring devices.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

8.  Intensive versus conventional insulin therapy in critically ill neurologic patients: still searching for the sweet spot.

Authors:  Jennifer A Frontera
Journal:  Neurocrit Care       Date:  2010-12       Impact factor: 3.210

9.  Comparison of subcutaneous and intravenous continuous glucose monitoring accuracy in an operating room and an intensive care unit.

Authors:  Masaya Munekage; Tomoaki Yatabe; Masahiko Sakaguchi; Hiroyuki Kitagawa; Takahiko Tamura; Tsutomu Namikawa; Kazuhiro Hanazaki
Journal:  J Artif Organs       Date:  2015-12-31       Impact factor: 1.731

10.  Accuracy of bedside capillary blood glucose measurements in critically ill patients.

Authors:  C Dana Critchell; Vincent Savarese; Amy Callahan; Christine Aboud; Serge Jabbour; Paul Marik
Journal:  Intensive Care Med       Date:  2007-09-01       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.