A comparison of shimming techniques for optimizing fat suppression in MR mammography.

Yasuo Takatsu1, Kengo Nishiyama, Tosiaki Miyati, Hideto Miyano, Mariko Kajihara, Thai Akasaka.   

Abstract

We evaluated the degree of inhomogeneities of fat suppression using the fully automated three-dimensional breast shimming technique (Image Based-Smart: IB-Smart) and manual setting of a rectangular parallelepiped shim (volume shimming) in MR mammography. Information on breast shape was collected from 9 patients whose images were insufficiently fat-suppressed. A breast phantom made of a thermoplastic sheet was used. Shimming of the magnetic field was done with IB-Smart and various dimensions of volume shims: the anterior to posterior/right to left/head to foot directions were set to 75-150/150-350/50-150 mm. The volumes of inhomogeneously suppressed fat were measured. The calculated volume with inhomogeneous fat suppression with use of IB-Smart was 13.3 × 10(4) mm(3). The smallest volume of inhomogeneous fat suppression with volume shimming was 5.4 × 10(4) mm(3) when the anterior-posterior/right-left/head-foot directions were set to 75/350/50 mm. Our results show that using optimized dimensions of volume shims enables better fat suppression than does IB-Smart.

Entities:  

Mesh:

Year:  2013        PMID: 23728709     DOI: 10.1007/s12194-013-0224-3

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  14 in total

1.  Development and evaluation of TWIST Dixon for dynamic contrast-enhanced (DCE) MRI with improved acquisition efficiency and fat suppression.

Authors:  Yuan Le; Randall Kroeker; Hal D Kipfer; Chen Lin
Journal:  J Magn Reson Imaging       Date:  2012-04-27       Impact factor: 4.813

Review 2.  The current status of breast MR imaging. Part I. Choice of technique, image interpretation, diagnostic accuracy, and transfer to clinical practice.

Authors:  Christiane Kuhl
Journal:  Radiology       Date:  2007-08       Impact factor: 11.105

Review 3.  Current status of breast MR imaging. Part 2. Clinical applications.

Authors:  Christiane K Kuhl
Journal:  Radiology       Date:  2007-09       Impact factor: 11.105

Review 4.  The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds.

Authors:  J F Schenck
Journal:  Med Phys       Date:  1996-06       Impact factor: 4.071

5.  Simple proton spectroscopic imaging.

Authors:  W T Dixon
Journal:  Radiology       Date:  1984-10       Impact factor: 11.105

6.  Contrast-enhanced magnetic resonance angiography of the lower extremities: standard-dose vs. high-dose gadodiamide injection.

Authors:  Ulrich Krause; Thomas Kroencke; Erika Spielhaupter; Matthias Taupitz; Werner Kenn; Bernd Hamm; Dietbert Hahn
Journal:  J Magn Reson Imaging       Date:  2005-04       Impact factor: 4.813

7.  Use of a modified polysaccharide gel in developing a realistic breast phantom for MRI.

Authors:  G P Mazzara; R W Briggs; Z Wu; B G Steinbach
Journal:  Magn Reson Imaging       Date:  1996       Impact factor: 2.546

8.  Breast ptosis. Definition and treatment.

Authors:  P Regnault
Journal:  Clin Plast Surg       Date:  1976-04       Impact factor: 2.017

9.  Diffusion-weighted whole-body MR imaging with background body signal suppression: a feasibility study at 3.0 Tesla.

Authors:  Petra Mürtz; Carsten Krautmacher; Frank Träber; Jürgen Gieseke; Hans H Schild; Winfried A Willinek
Journal:  Eur Radiol       Date:  2007-07-24       Impact factor: 5.315

10.  Breast MRI: guidelines from the European Society of Breast Imaging.

Authors:  R M Mann; C K Kuhl; K Kinkel; C Boetes
Journal:  Eur Radiol       Date:  2008-04-04       Impact factor: 5.315

View more
  3 in total

1.  A mask method to assess the uniformity of fat suppression in phantom studies.

Authors:  Yasuo Takatsu; Masafumi Nakamura; Kenichiro Yamamura; Satoshi Sawa; Masaki Asahara; Michitaka Honda; Tosiaki Miyati
Journal:  Radiol Phys Technol       Date:  2019-08-17

2.  Bilateral breast MRI by use of dual-source parallel radiofrequency excitation and image-based shimming at 3 Tesla: improvement in homogeneity on fat-suppression imaging.

Authors:  Kinya Ishizaka; Fumi Kato; Satoshi Terae; Suzuko Mito; Noriko Oyama-Manabe; Tamotsu Kamishima; Mitsuhiro Nakanishi; Hiroyuki Sugimori; Hiroyuki Hamaguchi; Hiroki Shirato
Journal:  Radiol Phys Technol       Date:  2014-06-26

3.  Assessment of the Quality of Breast MR Imaging Using the Modified Dixon Method and Frequency-Selective Fat Suppression: A Phantom Study.

Authors:  Yasuo Takatsu; Katsusuke Kyotani; Tsuyoshi Ueyama; Tosiaki Miyati; Kenichirou Yamamura; Atsushi Andou
Journal:  Magn Reson Med Sci       Date:  2018-01-30       Impact factor: 2.471

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.