Evaluation of operator radioprotection using a new injection device during vertebroplasty.

L Nguyen-Kim1, C Fargeot, H Beaussier, S Payen, J Chiras.   

Abstract

This study aimed to evaluate the protection granted by a simple device (X'TENS(®), Thiebaud, France) and to provide operators with information on the performance of this new device, which has not yet been assessed. Our assumption is that this device efficiently reduces the radiation dose to the operator. In a prospective clinical study, the radiation dose the operator's hand receives has been assessed using a specific sensor (UNFOR Instrument). Each patient included in the study was to receive at least two injections of cement during the procedure. Exposure was measured with and without the range extender. The data collected were then processed using a Wilcoxon matched pairs test. During 14 interventions, 20 vertebrae were treated with both procedures. Eleven women and three men were included. Seven patients underwent vertebroplasty for metastatic lesions and seven for osteoporotic lesions, bone fractures or vertebral compressions. The average injection time was 1.35 minutes with the device and 1.20 without (p=0.75). The dose to the hand per ml injected was 111.37 vs. 166.91 (p<0.05). Theoretically, the protection granted by the range extender depends on the length of the device. Our results are consistent with the inverse-square law. However, the variations in our results indicate that a proper and rigorous use is mandatory for the device to be effective. Given that radioprotection during fluoroscopy procedures is a frequently raised issue, the need for information for a safer practice increases likewise.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23693040      PMCID: PMC3670055          DOI: 10.1177/159101991301900205

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  16 in total

Review 1.  Medical devices of the head, neck, and spine.

Authors:  Tim B Hunter; Mark T Yoshino; Robert B Dzioba; Rick A Light; William G Berger
Journal:  Radiographics       Date:  2004 Jan-Feb       Impact factor: 5.333

2.  Radiation dose to the operator during vertebroplasty: prospective comparison of the use of 1-cc syringes versus an injection device.

Authors:  David F Kallmes; Erwin O; Soma Sinha Roy; Richard G Piccolo; William F Marx; Jae K Lee; Mary E Jensen
Journal:  AJNR Am J Neuroradiol       Date:  2003 Jun-Jul       Impact factor: 3.825

3.  Radiation exposure to operators during vertebroplasty.

Authors:  Atsushi Komemushi; Noboru Tanigawa; Shuji Kariya; Hiroyuki Kojima; Yuzo Shomura; Satoshi Sawada
Journal:  J Vasc Interv Radiol       Date:  2005-10       Impact factor: 3.464

4.  Radiation exposure to the surgeon during fluoroscopically assisted percutaneous vertebroplasty: a prospective study.

Authors:  Roger Harstall; Paul F Heini; Roberto L Mini; Rene Orler
Journal:  Spine (Phila Pa 1976)       Date:  2005-08-15       Impact factor: 3.468

5.  Surgeon's radiation exposure during percutaneous vertebroplasty.

Authors:  Michael Synowitz; Juergen Kiwit
Journal:  J Neurosurg Spine       Date:  2006-02

6.  Radiation exposure to the hands of operators during angiographic procedures.

Authors:  J Damilakis; M Koukourakis; A Hatjidakis; S Karabekios; N Gourtsoyiannis
Journal:  Eur J Radiol       Date:  1995-11       Impact factor: 3.528

7.  Radiation exposure during standard and complex interventional procedures.

Authors:  T Vehmas
Journal:  Br J Radiol       Date:  1997-03       Impact factor: 3.039

8.  What factors influence radiologists' finger doses during percutaneous drainages under fluoroscopic guidance?

Authors:  T Vehmas
Journal:  Health Phys       Date:  1993-08       Impact factor: 1.316

9.  [Radiodermatitis in interventional radiology (hand dose measurement, screening and compensation)].

Authors:  S Artignan; F Conso; V Hazebroucq
Journal:  J Radiol       Date:  2003-03

Review 10.  Future directions in percutaneous vertebroplasty.

Authors:  S Masala; F Massari; R Fiori; M Mammucari; D A Bartolucci; G Simonetti
Journal:  Radiol Med       Date:  2009-06-25       Impact factor: 3.469

View more
  1 in total

1.  Significantly reduced radiation dose to operators during percutaneous vertebroplasty using a new cement delivery device.

Authors:  Guang-Quan Zhang; Yan-Zheng Gao; Shu-Lian Chen; Shuai Ding; Kun Gao
Journal:  BMC Musculoskelet Disord       Date:  2014-08-01       Impact factor: 2.362

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.