| Literature DB >> 23620787 |
Lidya Chaidir1, Ida Parwati, Jessi Annisa, Soni Muhsinin, Intan Meilana, Bachti Alisjahbana, Reinout van Crevel.
Abstract
BACKGROUND: Fluorescence microscopy (FM) has not been implemented widely in TB endemic settings and little evaluation has been done in HIV-infected patients. We evaluated diagnostic performance, time and costs of FM with light-emitting diodes technology (LED-FM), compared with conventional (Zieh-Neelsen) microscopy in a hospital in Indonesia which acts as referral centre for HIV-infected patients.Entities:
Mesh:
Year: 2013 PMID: 23620787 PMCID: PMC3631225 DOI: 10.1371/journal.pone.0061727
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Proportion of positive smears confirmed with culture.
| Positive smear grading scale | LED-FM | ZN | ||
| n = 94 | positive by culture | n = 40 | positive by culture | |
| Scanty | 69 | 19 (27,5%) | 10 | 6 (60%) |
| +1 | 6 | 5 (83,3%) | 11 | 6 (54,5%) |
| +2 | 4 | 3 (75%) | 3 | 2 (66,7%) |
| +3 | 15 | 14 (93,3%) | 16 | 14 (87,5%) |
smear results were reported by using the grading scales for ZN at 1000×magnification and FM at 400×magnification from WHO/IUATLD [18].
Figure 1ROC curve for LED-FM.
The optimal threshold for sensitivity and specificity of LED-FM was assessed using ROC analysis. Maximum area under the curve (AUC 0.85, 95% CI 0.78–0.92) was obtained at threshold of ≥2 AFB/length (sensitivity 75,5%, specificity 90,0%,). ROC = receiver operating characteristic, LED-FM = Light emitting diode-fluorescence microscopy, AFB = Acid Fast Bacilli, CI = confidence interval.
Accuracy of LED-FM and ZN method in direct and decontaminated smears.
| direct smears (n = 404) | decontaminated smears (n = 102) | |
| LED-FM | ||
| Sensitivity (95% CI) | 75,5 (71,3–79,7) | 81,8 (74,3–89,3) |
| Specificity (95% CI) | 90,0 (87,1–93,0) | 91,2 (85,7–96,7) |
| ZN | ||
| Sensitivity (95% CI) | 52,8 (47,9–57,7) | 72,7 (64,0–81,3) |
| Specificity (95% CI) | 96,6 (94,8–98,4) | 97,8 (95,0–100) |
LED-FM = Light emitting diode-fluorescence microscopy, ZN = Ziehl-Neelsen, CI = confidence interval.
Accuracy of LED-FM and ZN in two study groups.
| LED-FM | ZN | |
| Study group 1: Unknown HIV status (n = 404) | ||
| Sensitivity (95% CI) | 75,5 (71,3–79,7) | 52,8 (47,9–57,7) |
| Specificity (95% CI) | 90,0 (87,1–93,0) | 96,6 (94,8–98,4) |
| PPV (95% CI) | 53,3 (48,4–58,2) | 70,0 (65,5–74,5) |
| NPV (95% CI) | 96,0 (94,8–98,4) | 93,1 (90,6–95,6) |
| Study group 2: HIV-positive (n = 256) | ||
| Sensitivity (95% CI) | 65,2 (59,4–71,0) | 58,0 (52,0–64,0) |
| Specificity (95% CI) | 90,4 (86,8–94,0) | 96,3 (94,0–98,6) |
| PPV (95% CI) | 71,4 (65,9–77,0) | 85,1 (80,7–89,5) |
| NPV (95% CI) | 87,6 (83,6–91,6) | 86,1 (81,9–90,3) |
LED-FM = Light emitting diode-fluorescence microscopy, ZN = Ziehl-Neelsen, PPV = positive predictive value, NPV = negative predictive value, CI = confidence interval.