| Literature DB >> 23606907 |
Xiuqing Zheng1, Zhiwu Liao, Shaoxiang Hu, Ming Li, Jiliu Zhou.
Abstract
NLMs is a state-of-art image denoising method; however, it sometimes oversmoothes anatomical features in low-dose CT (LDCT) imaging. In this paper, we propose a simple way to improve the spatial adaptivity (SA) of NLMs using pointwise fractal dimension (PWFD). Unlike existing fractal image dimensions that are computed on the whole images or blocks of images, the new PWFD, named pointwise box-counting dimension (PWBCD), is computed for each image pixel. PWBCD uses a fixed size local window centered at the considered image pixel to fit the different local structures of images. Then based on PWBCD, a new method that uses PWBCD to improve SA of NLMs directly is proposed. That is, PWBCD is combined with the weight of the difference between local comparison windows for NLMs. Smoothing results for test images and real sinograms show that PWBCD-NLMs with well-chosen parameters can preserve anatomical features better while suppressing the noises efficiently. In addition, PWBCD-NLMs also has better performance both in visual quality and peak signal to noise ratio (PSNR) than NLMs in LDCT imaging.Entities:
Mesh:
Year: 2013 PMID: 23606907 PMCID: PMC3626180 DOI: 10.1155/2013/902143
Source DB: PubMed Journal: Comput Math Methods Med ISSN: 1748-670X Impact factor: 2.238
Figure 1Images and their pointwise box-counting dimension images: the first row shows images while the second row shows their pointwise box-counting dimension images. Here r = 32 and ε = 2, 4, 8, 16, 32.
PSNR for the test image.
| Noise | PSNR of | PSNR of | PSNR of |
|---|---|---|---|
| parameters | the noisy image | NLMs | PWBCD-NLMs |
|
| 23.29 | 34.19 |
|
|
| 21.88 | 33.79 |
|
|
| 21.30 | 33.45 |
|
Figure 2Noisy test images and reconstructed images.
Figure 3(b) Real LDCT reconstructed image, (a) related SDCT reconstructed images and (c)-(d) reconstructed images from LDCT sinogram using NLMs and the new method.