The relevance of image quality indices for dose optimization in abdominal multi-detector row CT in children: experimental assessment with pediatric phantoms.

H J Brisse1, J Brenot, N Pierrat, G Gaboriaud, A Savignoni, Y De Rycke, S Neuenschwander, B Aubert, J-C Rosenwald.   

Abstract

This study assessed and compared various image quality indices in order to manage the dose of pediatric abdominal MDCT protocols and to provide guidance on dose reduction. PMMA phantoms representing average body diameters at birth, 1 year, 5 years, 10 years and 15 years of age were scanned in a four-channel MDCT with a standard pediatric abdominal CT protocol. Image noise (SD, standard deviation of CT number), noise derivative (ND, derivative of the function of noise with respect to dose) and contrast-to-noise ratio (CNR) were measured. The 'relative' low-contrast detectability (rLCD) was introduced as a new quantity to adjust LCD to the various phantom diameters on the basis of the LCD(1%) assessed in a Catphan phantom and a constant central absorbed dose. The required variations of CTDIvol(16) with respect to phantom size were analyzed in order to maintain each image quality index constant. The use of a fixed SD or CNR level leads to major dose ratios between extreme patient sizes (factor 22.7 to 44 for SD, 31.7 to 51.5 for CNR(2.8%)), whereas fixed ND and rLCD result in acceptable dose ratios ranging between factors of 2.9 and 3.9 between extreme phantom diameters. For a 5-9 mm rLCD1(%), adjusted ND values range between -0.84 and -0.11 HU mGy(-1). Our data provide guidance on dose reduction on the basis of patient dimensions and the required rLCD (e.g., to get a constant 7 mm rLCD(1%) for abdominal diameters of 10, 13, 16, 20 and 25 cm, tube current-time product should be adjusted in order to obtain CTDIvol(16) values of 6.2, 7.2, 8.8, 11.6 and 17.7 mGy, respectively).

Entities:  

Mesh:

Year:  2009        PMID: 19265204     DOI: 10.1088/0031-9155/54/7/002

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  Can a revised paediatric radiation dose reduction CT protocol be applied and still maintain anatomical delineation, diagnostic confidence and overall imaging quality?

Authors:  S Kritsaneepaiboon; P Siriwanarangsun; P Tanaanantarak; A Krisanachinda
Journal:  Br J Radiol       Date:  2014-06-24       Impact factor: 3.039

2.  Pediatric thoracic CT angiography at 70 kV: a phantom study to investigate the effects on image quality and radiation dose.

Authors:  Robert D MacDougall; Patricia L Kleinman; Lifeng Yu; Edward Y Lee
Journal:  Pediatr Radiol       Date:  2016-03-17

3.  Effect of low tube voltage on image quality, radiation dose, and low-contrast detectability at abdominal multidetector CT: phantom study.

Authors:  Kun Tang; Ling Wang; Rui Li; Jie Lin; Xiangwu Zheng; Guoquan Cao
Journal:  J Biomed Biotechnol       Date:  2012-04-23

4.  A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre-A phantom study.

Authors:  Patrizio Barca; Fabio Paolicchi; Giacomo Aringhieri; Federica Palmas; Daniela Marfisi; Maria Evelina Fantacci; Davide Caramella; Marco Giannelli
Journal:  PLoS One       Date:  2021-01-14       Impact factor: 3.240

5.  Improving spatial adaptivity of nonlocal means in low-dosed CT imaging using pointwise fractal dimension.

Authors:  Xiuqing Zheng; Zhiwu Liao; Shaoxiang Hu; Ming Li; Jiliu Zhou
Journal:  Comput Math Methods Med       Date:  2013-03-31       Impact factor: 2.238

6.  Regularized multidirections and multiscales anisotropic diffusion for sinogram restoration of low-dosed computed tomography.

Authors:  Zhiwu Liao
Journal:  Comput Math Methods Med       Date:  2013-11-21       Impact factor: 2.238

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.