A vascular complication in computer navigated total knee arthroplasty.

Sandesh Gulhane1, Ian Holloway, Mathew Bartlett.   

Abstract

The use of computer navigation has the potential to improve implant position in total knee arthroplasty (TKA), but pin fixation of reference arrays introduces an additional potential source of complications. We report a case of vascular injury related to the insertion of a femoral pin during navigated TKA.

Entities:  

Keywords:  Navigated knee arthroplasty; total knee arthroplasty; vascular complications

Year:  2013        PMID: 23531499      PMCID: PMC3601244          DOI: 10.4103/0019-5413.106927

Source DB:  PubMed          Journal:  Indian J Orthop        ISSN: 0019-5413            Impact factor:   1.251


INTRODUCTION

Computer-assisted surgery has been popularized in orthopedic surgery in the last 10 years. Whilst a number of recent randomized control trials have shown some benefit in terms of implant position over conventional jig-based referencing,1 improved survivorship has not yet been demonstrated. Many navigation systems require the use of reference arrays, which are fixed to the tibia and femur with pins. Reported complications include pin breakage,2 superficial wound infection,3–4 prolonged operation time,5–8 and prolonged tourniquet time.8

CASE REPORT

A 58-year-old man with a diagnosis of varus osteoarthritis (OA) of the right knee underwent navigated total knee arthroplasty (TKA) using standard technique. Our current practice is to use computer navigation for all primary TKA. We use a passive reflector-based system (Brainlab, Feldkirchen, Germany) with fixation of the femoral reference array using two 3-mm-diameter threaded pins. These are inserted anteriorly through stab incisions, with the knee in flexion, as proximally as the tourniquet will allow. Tibial pins are inserted in the mid diaphysis into the subcutaneous surface of the tibia [Figure 1]. All pins are bicortical in order to maintain secure fixation for the duration of surgery.
Figure 1

Peroperative photographs (a) Insertion of tibial and femoral reference arrays (b) The styloid used to map out the tibia and femur

Peroperative photographs (a) Insertion of tibial and femoral reference arrays (b) The styloid used to map out the tibia and femur Computer registration is performed with digitalization of the hip and ankle centers and specific landmarks of the knee. A size 6 generation II cemented femoral implant and a size 6 generationII cemented tibial component were implanted with an 11-mm polyethylene insert. There was no obvious complication during surgery. The patient had some pain and thigh swelling postoperatively, but this was considered to be in keeping with recent surgery. He was discharged on the third postoperative day, having satisfied physiotherapy and occupational therapy goals. He reported 3 days later with increased right thigh pain, swelling, and a limited range of movement. He was given analgesia and placed onto a continuous passive motion machine. Deep vein thrombosis and fracture were excluded on ultrasound scan and plain radiographs. An ultrasound guided diagnosis of a quadriceps hematoma at the site of the femoral pins was made. This was confirmed by computerized topography (CT) with intravenous contrast [Figure 2], which showed a large quadriceps hematoma and active bleeding into the femoral canal from a branch of the superficial femoral artery as it entered linea aspera at the site of pin insertion.
Figure 2

Two CT scan images showing active bleeding into the femoral canal

Two CT scan images showing active bleeding into the femoral canal On day 10, the hematoma was evacuated and two 14-mm unicortical screws were inserted into the femur at the site of the bleed to tamponade the bleed. Unicortical screws were used to reduce the risk of further arterial injury with bicortical screws and the CT scan also showed unicortical injury. At 2 weeks post presentation, he underwent a manipulation under anesthesia for ongoing stiffness. He then had a flexion of 0°–90°, but also had some persistent swelling and was prescribed a short course of diazepam. He showed continued improvement since the second operation and had no further complications at the final followup 6 months later [Figure 3].
Figure 3

Final followup X-ray showing insertion of two unicortical screws

Final followup X-ray showing insertion of two unicortical screws

DISCUSSION

Computer navigated TKA has the potential advantage of improved implant position, which is known to reduce the rate of aseptic loosening. There have been a number of studies reporting complications of computer navigated TKA among these. Böhling et al.,9 Chin et al.,10 and Bäthis et al.11 demonstrated longer tourniquet time with computer navigation. Kalairajah et al.12 reported a longer tourniquet time as well as a statistically significant decrease in blood drainage. Chauhan et al.13 and Jenny et al.4 found that computerized navigation was associated with a longer operating time. Stöckl et al.14 reported one case of poor tibial tracker stability, related to osteoporotic bone, resulting in abandonment of computer navigation, from 64 patients. Victor et al.2 found pin breakage in 3 patients from a population of 100. Hernañdez-Vaquero15 reported four cases of failed pin fixation (three iliac crest and one femur) requiring abandonment of computer navigation from a total of 112 patients. Our report is the only case of arterial injury in the literature related to navigation pin usage. It highlights an important complication of computer navigated TKA. If the patient develops an unexpected postoperative swelling with increasing pain this diagnosis needs to be taken into consideration.
  14 in total

1.  Computer-assisted knee arthroplasty versus a conventional jig-based technique. A randomised, prospective trial.

Authors:  S K Chauhan; R G Scott; W Breidahl; R J Beaver
Journal:  J Bone Joint Surg Br       Date:  2004-04

2.  Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional technique.

Authors:  H Bäthis; L Perlick; M Tingart; C Lüring; D Zurakowski; J Grifka
Journal:  J Bone Joint Surg Br       Date:  2004-07

3.  Navigation improves accuracy of rotational alignment in total knee arthroplasty.

Authors:  Bernd Stöckl; Michael Nogler; Rafal Rosiek; Martin Fischer; Martin Krismer; Oliver Kessler
Journal:  Clin Orthop Relat Res       Date:  2004-09       Impact factor: 4.176

4.  Minimally invasive unicompartmental knee replacement with a nonimage-based navigation system.

Authors:  L Perlick; H Bäthis; M Tingart; C Perlick; C Lüring; J Grifka
Journal:  Int Orthop       Date:  2004-03-06       Impact factor: 3.075

5.  Image-based computer-assisted total knee arthroplasty leads to lower variability in coronal alignment.

Authors:  Jan Victor; Davy Hoste
Journal:  Clin Orthop Relat Res       Date:  2004-11       Impact factor: 4.176

6.  Blood loss after total knee replacement: effects of computer-assisted surgery.

Authors:  Y Kalairajah; D Simpson; A J Cossey; G M Verrall; A J Spriggins
Journal:  J Bone Joint Surg Br       Date:  2005-11

7.  Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a case-control study of 235 cases compared with 235 conventionally implanted prostheses.

Authors:  Jean-Yves Jenny; Ulrich Clemens; Steffen Kohler; Hartmuth Kiefer; Werner Konermann; Rolf K Miehlke
Journal:  J Arthroplasty       Date:  2005-10       Impact factor: 4.757

8.  Total knee arthroplasty implanted with and without kinematic navigation.

Authors:  R Hart; M Janecek; A Chaker; P Bucek
Journal:  Int Orthop       Date:  2003-08-26       Impact factor: 3.075

9.  Leg axis after computer-navigated total knee arthroplasty: a prospective randomized trial comparing computer-navigated and manual implantation.

Authors:  Ralf Decking; Yma Markmann; Johannes Fuchs; Wolfhart Puhl; Hanns-Peter Scharf
Journal:  J Arthroplasty       Date:  2005-04       Impact factor: 4.757

10.  Complications of fixed infrared emitters in computer-assisted total knee arthroplasties.

Authors:  Daniel Hernández-Vaquero; Abelardo Suárez-Vázquez
Journal:  BMC Musculoskelet Disord       Date:  2007-07-27       Impact factor: 2.362

View more
  2 in total

1.  Computer-assisted navigation is beneficial both in primary and revision surgery with modular rotating-hinge knee arthroplasty.

Authors:  Björn G Ochs; Anna J Schreiner; Peter M de Zwart; Ulrich Stöckle; Christoph Emanuel Gonser
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-09-20       Impact factor: 4.342

2.  Clinical value of CT-based patient-specific 3D preoperative design combined with conventional instruments in primary total knee arthroplasty: a propensity score-matched analysis.

Authors:  Kai Lei; Li Ming Liu; Yi Xiang; Xin Chen; Hua Quan Fan; Yang Peng; Jiang Ming Luo; Lin Guo
Journal:  J Orthop Surg Res       Date:  2020-12-09       Impact factor: 2.359

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.