Literature DB >> 23470928

A multi-institutional study for tolerance and action levels of IMRT dose quality assurance measurements in Korea.

Jung-in Kim1, Jin-Beom Chung, Yang-Kyun Park, Ju-Young Song, Sung Kyu Kim, Sung Hwan Ahn, Chang Heon Choi, Won Hoon Choi, Byungchul Cho, Sang Gyu Ju, Sung Jin Kim, Sung-Joon Ye.   

Abstract

The purpose of this study was to suggest tolerance levels for IMRT DQA measurements using confidence limits determined by a multi-institutional study in Korea. Ten institutions were grouped into LINAC (seven linear accelerators) and TOMO (three tomotherapy machines). The DQA processes consisted of point (high- and low-dose regions) and planar (per-field and composite-field) dose measurements using an ion chamber and films (or 2D detector array) inserted into a custom-made acryl phantom (LINAC) or a cheese phantom (TOMO). The five mock structures developed by AAPM TG-119 were employed, but the prostate as well as the H&N structures were modified according to Korean patients' anatomy. The point measurements were evaluated in a ratio of measured and planned doses, while the planar dose distributions were assessed using two gamma criteria of 2 mm/2% and 3 mm/3%. The confidence limit (|mean + 1.96 σ|) for point measurements was determined to be 3.0% in high-dose regions and 5.0% in low-dose regions. The average percentage of points passing the gamma criteria of 2 mm/2% and 3mm/3% for per-field measurements was 92.7 ± 6.5% and 98.2 ± 2.8%, respectively. Thus, the corresponding confidence limit was 79.1% and 92.7%, respectively. The gamma passing rate averaged over all mock tests and institutions for composite-field measurements was 86.1 ± 6.5% at 2 mm/2% and 95.3 ± 3.8% at 3 mm/3%, leading to the confidence limit of 73.3% and 87.9%, respectively. There was no significant difference in the tolerance levels of point dose measurements between LINAC and TOMO groups. In spite of the differences in mock structures and dosimetry tools, our tolerance levels were comparable to those of AAPM and ESTRO guidelines.

Entities:  

Mesh:

Year:  2013        PMID: 23470928      PMCID: PMC5714373          DOI: 10.1120/jacmp.v14i2.3964

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


I. INTRODUCTION

Intensity‐modulated radiation therapy (IMRT) has been rapidly implemented in Korea due to recent governmental policy.( ) Roughly three‐fourths of all radiation treatment facilities currently perform IMRT. IMRT is capable of generating extremely precise dose distributions that provide conformal coverage of complex target shapes and conformal avoidance of sensitive normal structures.( ) The AAPM (American Association of Physicists in Medicine) and ESTRO (European Society for Therapeutic Radiology and Oncology) emphasized a comprehensive quality assurance program for clinical implementation of IMRT.( , , ) Also, the guideline of IMRT use in clinical trials has been developed throughout the world.( , ) AAPM Task Group 119 carried out a multi‐institutional research to assess the overall accuracy of planning and delivery of IMRT, and produced quantitative confidence limits as baseline expectation values for IMRT commissioning.( ) The British group carried out a national dosimetry audit of IMRT to provide an independent check of safe implementation and to identify problems in the modeling and delivery of IMRT.( , – ) With a rapid implementation of IMRT in Korea, ten institutions reached consensus on a multi‐institutional study to provide tolerance levels of IMRT DQA measurements as a national safety guideline for the overall accuracy of IMRT planning and delivery. The confidence limit concept and test protocol of AAPM TG‐119 were used as a basis of this study. However, the mock structures were modified according to Korean patients' anatomy, and the dosimetry tools and phantoms used in this study were somewhat different from those of AAPM TG‐119. Furthermore, ten institutions were grouped into LINAC (seven linear accelerator‐based centers) and TOMO (three tomotherapy‐based centers), and the results of both groups were separately analyzed.

II. MATERIALS AND METHODS

This study was performed from October 2010 to September 2011. Ten participating institutions geometrically spread over Korea were divided into the LINAC and TOMO groups. The planning and delivery systems used at each institution are summarized in Table 1. Both groups used 6 MV photon beams. The institutions listed in subsequent tables were anonymously identified only by letter.
Table 1

List of participating institutions and the planning and delivery systems used.

Group LINAC TOMO
InstitutionSeoul Nat'l Univ. Bundang Hosp.Jeju Nat'l Univ. HospYeungnam Univ. Hosp.Dong‐A Univ. Hosp.Eulji Univ. Hosp.Seoul Nat'l Univ. HospAsan Medical CenterSeoul Samsung Medical CenterYonsei Cancer CenterChonnam Nat'l Univ. Hwasun Hosp.
AcceleratorVarian 21ExSVarian IXVarian 21ExSVarian NovalisElekta SynergyVarian IXVarian TrilogyTomotherapyTomotherapyTomotherapy
Delivery TechniqueDMLCDMLCDMLCDMLCSMLCDMLCDMLCBMLCBMLCBMLC
Planning SystemEclipse 6.5Eclipse 8.6Eclipse 8.6BrainLab iPlanCMS Monaco 2.0.3Eclipse 8.9Eclipse 8.9Tomotherapy TPS 3.1.4Tomotherapy TPS 4.0.2Tomotherapy TPS 3.2.3.2

; ; and .

List of participating institutions and the planning and delivery systems used. ; ; and .

A. Output and TPS audit

One of the institutions has been participating in a RTOG (Radiation Therapy Oncology Group) trial and passed the RPC's (Radiological Physics Center. MD Anderson Cancer Center, USA) credentialing requirements. The dose audit was performed with OSLD (Optically Stimulated Luminescence Dosimeter). RPC presented that the variance of the dose determined by a single OSLD was less than 3%. The OSLD chips were irradiated at of photon beams at 100 cm SSD (source‐to‐surface distance) in the institution and then shipped to RPC for analysis. As a result of RPC auditing, the absorbed dose measurements in good agreement were ranged from 0.98 to 1.00. This institution played the role as the reference site to indirectly verify the machine output of the other institutions through site visiting. The output was measured by skilled physicists of the reference site using a farmer type ionization chamber (0.6 cc, PTW TN30013; PTW, Freiburg, Germany) and an electrometer (PTW UNIDOS) calibrated by KRISS (Korea Research Institute of Standards and Science). The TPS (treatment planning system) commissioning audit was carried out using two test protocols described by Van Esch et al.( , ) These tests were designed with three consecutive rectangular volumes that had different specified doses. The first test called a “Dip” test, for which the specified dose to the middle volume and each outer volume was 0.7 and 0.0 Gy, respectively. This test was performed for dynamic delivery to verify that the leaves could adequately shield the central volume, and the TPS modeled the transmission of the leaves correctly. The second test is called a “Step” test, for which the specified dose to each volume was 0.7, 0.5 and 0.3 Gy, respectively. The Step test was for static delivery, to test the delivered accuracy of three relative dose levels. Each institution delineated the predefined volumes for both tests on a local solid water phantom and delivered the specified doses to films (EBT2, International Specialty Products, Wayne, NJ). The reference site centrally evaluated the films from all institutions.

B. Phantom

The LINAC group used the same custom‐made phantom of acryl to measure point dose and axial plane dose distributions. The phantom of cylinder, 265 mm in length and 180 mm in diameter, has two holes where an ion chamber can be inserted. It was cut into two pieces, as shown in Fig. 1. The two pieces of phantom (gray and green parts in Fig. 1) were tightened using the lever after inserting a film in order to reduce the air gap between them. One hole at 5 cm depth below the anterior surface was used to measure a conversion factor [nC/cGy] of chamber reading‐to‐dose. This standard measurement was also intended to exclude the daily variation of machine output. The other hole along the axis was used to measure a planned dose in a high‐ or low‐dose and low‐gradient region. The holes were designed especially for a 0.125 cc ion chamber (Semiflex, PTW, Freiburg, Germany). For the TOMO group, the commercial phantom (i.e., ‘cheese’ phantom; (Accuray, Sunnyvale, CA) was used to measure point doses and coronal plane dose distributions. Details of the measurements using this phantom are described elsewhere.( )
Figure 1

Developed IMRT DQA phantom for LINAC group.

Developed IMRT DQA phantom for LINAC group.

C. Mock structures

The AAPM TG‐119 produced the mock structures for the prostate, head and neck (H&N), C‐shape, and multitarget. Each test included the targets, normal structures, planning specifications of dose goals, and beam arrangements. This study followed the general guideline of the AAPM TG‐119 programs. However, the target volumes and OAR (organs at risk) locations of the mock prostate and H&N were based on Korean patients' anatomy.( ) The rest of the AAPM TG‐119 mock structures were identically applied for this study. All mock structures were segmented by one physicist from the reference site on DICOM CT images (DICOM, Rosslyn, VA) of both phantoms. Then these DICOM files were centrally distributed to all the institutions to eliminate any institutional variations during segmentation. It was recommended that a grid size of dose calculation used be less than 3 mm and that the calculation algorithm used be the convolution–superposition or equivalent for inhomogeneity correction.( , )

C.1 Structures

For the prostate test, the mean PTV (planning target volume) and CTV (clinical target volume) volumes of sample Korean patients were 141 cc and 50 cc, respectively. The PTV was defined to include a 1.0 cm margin around the prostate in all directions, except the posterior direction where a 0.5 cm margin was added. The rectum was a cylinder with a diameter of 1.5 cm (mean volume 10 cc) and the bladder was a semi‐ellipsoidal shape (mean volume 144 cc). The PTV included about one‐third of the rectum and bladder volumes. Unlike AAPM TG‐119, the femoral heads of spherical shape were added in this test suite. For the H&N test, the mean PTV volume of sample Korean patients was 534 cc. PTV included all anterior volume from the base of the skull to the upper neck and the posterior neck node. Both parotids were delineated with two “truncated cones” (i.e., a cone with the top cut off) of 1.5 cm diameter of a circular top and 2.4 cm diameter of a circular bottom opposing the circular bottom with 30 cc volume and located at the superior aspect of the PTV. The cord was a cylinder shape with a diameter of 1.5 cm. A gap between the cord and PTV was about 1.3 cm. For the C‐shape and multitarget tests, the AAPM TG‐119 structures were used as is. The mock structures are shown in Fig. 2.
Figure 2

Isodose lines for low‐dose point measurement with mock structures: (a) C‐shape easy, (b) C‐shape hard, (c) H&N, (d) superior of multitarget, (e) inferior of multitarget, and (f) prostate.

Isodose lines for low‐dose point measurement with mock structures: (a) C‐shape easy, (b) C‐shape hard, (c) H&N, (d) superior of multitarget, (e) inferior of multitarget, and (f) prostate.

C.2 Beam arrangements

For the prostate and multitarget tests, the plan had seven fields at 50° intervals from the vertical (e.g., 0°, 50°, 100°, 150°, 210°, 260°, and 310°). For H&N and C‐shape tests, the plan had nine fields at 40° intervals from the vertical (e.g., 0°, 40°, 80°, 120°, 160°, 200°, 240°, 280°, and 320°).

C.3 Dose goals

The total prescription dose was 80 Gy (daily dose of ) for the prostate test. The total prescription dose of 50 Gy (daily dose of ) was applied for H&N, C‐shape, and multitarget tests. The plan goals of C‐shape were divided into easy and hard versions. The specific planning goals are shown in Table 2.
Table 2

Treatment plan goals and results for all tests of both groups.

Test Group LINAC TOMO
Parameter Goal Mean STDV COV Mean STDV COV
ProstatePTV D95 >7600 7620.452.90.0077788.3166.30.021
PTV D5 <8400 8267.4188.10.0238111.0118.30.015
Rtm D30 <7000 6630.6392.40.0596270.0628.60.100
Rtm D10 <7500 7324.6208.70.0287694.0146.90.019
Bld D30 <7000 5452.7738.70.1355346.7660.30.124
Bld D10 <7500 7414.8144.80.0207729.3233.20.030
RtF D10 <5000 4127.4503.40.1224171.7926.40.222
LtF D10 <5000 4014.8504.50.1264112.7768.70.187
H&NPTV D9050005052.984.60.0174996.770.20.014
PTV D99 >4650 4784.194.80.0204883.398.70.020
PTV D20 <5500 5289.6147.20.0285215.7149.30.029
Cd max <4000 3915.0257.50.0663282.0499.20.152
RtPd D50 <2000 1916.7160.20.0841438.3192.40.134
LtPd D50 <2000 1887.3135.30.0721394.3144.70.104
C‐shape(E)PTV D9550004985.664.80.0134982.019.30.004
PTV D10 <5500 5463.3188.60.0355437.7207.70.038
Core D10 <2500 2446.3145.10.0591793.3583.20.325
C‐shape(H)PTV D9550004937.0116.60.0244790.7116.80.024
PTV D10 <5500 5639.4162.70.0295950.0578.20.097
Core D10 <1000 1552.9211.90.1361178.3361.40.307
Multiple‐targetCt D99 >5000 4975.654.00.0114918.074.60.015
Ct D13 <5300 5417.2117.10.0225852.71002.60.171
Sup D99 >2500 2676.2204.20.0762437.384.50.035
Sup D13 <3500 3521.3352.00.1003700.7149.90.041
Inf D99 >1250 1430.3353.20.2471220.060.80.050
Inf D13 <2500 2593.5607.20.2343262.7423.80.130

; ; ; ; ; ; ; ; ; ; ; ; .

Treatment plan goals and results for all tests of both groups. ; ; ; ; ; ; ; ; ; ; ; ; .

C.4 Measurement location

A point for high‐dose measurements was always the isocenter in the middle of PTV, where doses were high and uniform (see Fig. 3). A point for low‐dose measurements was located in the OAR structure such as 3 cm posterior to the rectum for the prostate, 4 cm posterior in midspinal cord for H&N, the center of cord for the C‐shape, and the center of either of two outer targets for the multitarget (see Fig. 2). For the per‐field measurements, a plane perpendicular to the beam axis was located at 5 cm depth in a water‐equivalent phantom with the SAD (source‐to‐axis distance) setup. The per‐field measurement was only performed for H&N test and limited to the LINAC group. A plane for the composite field measurement was across the isocenter. The plane selected for the composite field measurement often included the points for high‐ and low‐dose measurements.
Figure 3

Isodose lines for high‐dose point measurements and composite field plane with mock structures: (a) C‐shape easy, (b) C‐shape hard, (c) H&N, (d) multitarget center, and (e) prostate.

Isodose lines for high‐dose point measurements and composite field plane with mock structures: (a) C‐shape easy, (b) C‐shape hard, (c) H&N, (d) multitarget center, and (e) prostate.

D. Point measurements

The LINAC group used the phantom of Fig. 1 using a 0.125 cc ion chamber (Semiflex, PTW, Freiburg, Germany). Two sets of phantom CT images (i.e., vertical and horizontal positions) were acquired with a slice thickness of 2.5 mm. The vertical phantom images were used to calculate MUs in TPS for a conversion factor, where it was assumed that the vertical phantom at 100 cm SSD be irradiated by a field (so‐called “standard measurements”). The horizontal phantom images were used to calculate point dose values for all tests by assuming the horizontal phantom was irradiated by all fields as planned in TPS. These planned values were compared with the measured doses. In the TOMO group, the cheese phantom was used to measure a point dose using a 0.05 cc ion chamber (Exradin A1SL, Standard Imaging Inc., Middleton, WI). The TOMO group measured an absolute dose for a standard measurement using and values. The rest of the procedure for comparison was the same as those in the LINAC group. In the low‐dose measurement, the C‐shape plan had the lowest dose in the OAR, which was expected to be at least 30 cGy. Therefore, the issue regarding very low‐dose measurements would not arise in this study.( )

E. Planar dose measurements

The per‐field measurements were done by the LINAC group only for the H&N test. Each institution in the LINAC group used either a detector array or film; two institutions used MapCHECK (Sun Nuclear Corporation, Melbourne, FL); one institution used the ion chamber array (PTW, Freiburg, Germany); two institutions used MatriXX (IBA Dosimetry GmbH, Schwarzenbruck, Germany); and two institutions used EBT2 films (International Specialty Products, Wayne, NJ). The individual fields were delivered at gantry 0°. This avoided the angular dependency of detectors during per‐field measurements.( , ) All institutions were asked to perform the composite field measurements using films and the custom‐made (LINAC) or cheese (TOMO) phantom. The multitarget test was excluded from this measurement. The films for composite field measurement were evaluated using two gamma criteria of 2 mm DTA/2% dose difference and 3 mm DTA/3% dose difference. The planar dose distributions were normalized at a reference (dose to the isocenter) or maximum dose in a low‐gradient region. The region of interest (ROI) was first specified as a maximum size of rectangle on the film. Then any pixels that received less than 10% of the maximum dose in the dose map were excluded from the gamma evaluation. Thus, the excluded points were outside of ROI.

III. RESULTS

The results of output audit for all participating institutions ranged from to , which indicated that all of them passed the study criteria of less than 3%. Six institutions undertook Dip and Step tests for the TPS audit, and all test plans met the required dose constraints. Four of the institutions did not perform these tests, but were requested to present the results of local TPS commission tests with film measurements. The average percentage of points passing the gamma criteria of 3 mm/3% in the Dip and Step tests were 98.2% (97.4% to 99.2%) and 97.8% (97.2% to 98.6%), respectively.

B. Planning results

The statistics of the mock plans for both groups are presented in Table 2. In this table, Dxx means the dose covering xx% of the volume. The coefficient of variation (COV) was a normalized measurement of the dispersion of a probability distribution that was defined as a ratio of the standard deviation (STDV) to the mean.

C. Point dose measurements

In this study, the dose difference was expressed as a ratio of the difference between measured and planned doses to the planned dose, instead of the prescription dose used in AAPM TG‐119. The dose difference is planned according to the following equation: The planned dose from TPS was a value at a point of measurement for comparison. In high‐dose measurements, the average difference between measured and planned doses averaged over all tests was for LINAC, for TOMO, and for all institutions. It ranged from to 1.9% for the LINAC group and from to 2.9% for the TOMO group. The maximum dose difference occurred in the hard C‐shape structure for both groups. The average confidence limit was 3.1% for both groups. The results of these measurements for both groups are summarized in Table 3.
Table 3

Results of high‐dose point measurement, averaged over the institutions for LINAC/TOMO/Total groups.

Test Mean (%) STDV(%) CL(%) No.
Multitarget 0.3/0.2/0.2 1.0/1.8/1.22.3/3.6/2.57/3/10
Prostate 0.0/0.6/0.3 1.0/1.2/1.11.9/3.0/2.47/6/13
H&N 1.0/1.3/1.1 0.7/0.8/0.72.4/2.8/2.67/6/13
C‐shape(E) 1.3/0.2/0.9 1.1/1.4/1.43.5/3.0/3.57/3/10
C‐shape(H) 0.8/0.5/0.4 1.9/2.2/2.04.5/4.7/4.27/3/10
Overall combined 0.7/0.5/0.6 1.2/1.4/1.33.1/3.2/3.135/21/56

; ; and .

Results of high‐dose point measurement, averaged over the institutions for LINAC/TOMO/Total groups. ; ; and . During the high‐dose measurement analysis, two institutions showed abnormal values larger than . Both institutions were informed of these large discrepancies and they then repeated the measurements. They carefully checked the whole process and found that the conversion factor from the standard measurement was incorrect. After the correction, the results from both institutions were then less than 3%. In low‐dose measurements, the average difference between measured and planned doses averaged over all tests was for LINAC group, for TOMO group, and for all institutions. It ranged from to 3.5% for the LINAC group and from to 6.6% for TOMO group. The maximum dose difference also occurred in the hard C‐shape structure for both groups. The average confidence limit was 4.9% for both groups. The results of these measurements are shown in Table 4. The local confidence limit of each institution, averaged over all the test plans, is listed in Table 5.
Table 4

Results of low‐dose point measurement, averaged over the institutions for LINAC/TOMO/Total groups.

Test Mean (%) STDV(%) CL(%) No.
MT (suf) 0.3/1.0/0.1 1.8/2.7/2.13.9/6.3/4.17/3/10
MT (inf) 0.5/0.1/0.4 1.6/3.0/1.93.3/6.0/4.07/3/10
Prostate 1.5/1.0/1.2 1.3/1.2/1.23.9/3.4/3.67/6/13
H&N 1.7/0.5/1.1 0.8/2.1/1.63.3/4.5/4.27/6/13
C‐shape(E) 0.2/0.7/0.1 2.3/2.6/2.34.8/5.7/4.67/3/10
C‐shape(H) 2.3/3.7/0.5 2.6/2.6/3.87.4/8.7/7.97/3/10
Overall 1.0/0.1/0.6 1.9/2.5/2.24.8/5.0/4.942/24/66

; ; ; and .

Table 5

Results of dose point measurement, averaged over all the test plans measured at each.

Measurement Group LINAC TOMO
Institution A B C D E F G H I J
High DoseMean(%) 1.0 1.1 0.4 0.3 1.8 0.3 0.5 1.1 0.1 0.1
TDV(%)0.81.61.20.81.40.51.51.31.71.1
LCL(%) a 2.64.32.71.94.51.23.43.63.42.3
No.5555555777
Low DoseMean(%) 1.5 1.7 0.4 0.7 2.2 1.7 0.4 0.6 1.1 0.1
STDV(%)1.02.41.71.82.01.31.82.13.31.6
LCL(%) a 3.46.43.74.16.24.34.04.77.73.3
No.6666666888

Results of low‐dose point measurement, averaged over the institutions for LINAC/TOMO/Total groups. ; ; ; and . Results of dose point measurement, averaged over all the test plans measured at each.

D. Planar dose measurements

The planar dose distributions were assessed using two gamma criteria of 2 mm/2% and 3 mm/3%. Recently, several studies reported the statistical correlation between planar IMRT QA passing rates and clinical relevance using a DVH‐based method.( – ) However, the current study was based on the mock structures and didn't intend to predict any clinical relevance with DQA results. Thus, the method for per‐field and composite field analysis coincided with the AAPM TG‐119 tolerance limit metrics using the concept of confidence limit.

D.1 Per‐field measurements

The per‐field measurements were done by seven institutions of the LINAC group only. The H&N test was selected for the per‐field measurements since it had the most complicated dose distributions. Each institution used an available tool such as a detector array or films. The average percentage of points passing the gamma criteria of 2 mm/2% and 3 mm/3% was and , respectively. The corresponding confidence limit was 79.1% and 92.7% (see Table 6). The local confidence limit ranged from 71.4% to 95.3% for 2 mm/2% criteria and from 88.3% to 100% for 3 mm/3% criteria (Table 7).
Table 6

Per‐field measurement: averaged percentage of points passing gamma criteria of 2 mm/2% and 3 mm/3% over the institutions.

Criteria Mean(%) STDV(%) No. CL(%)
2 mm/2%92.76.56379.1
3 mm/3%98.22.86392.7
Table 7

Per‐field measurement: local averaged percentage of points passing gamma criteria of 2 mm/2% and 3 mm/3%, with associated confidence limits.

Criteria Institution A B C D E F G
Device MapCHECK MapCHECK EBT2 MatriXX EBT2 2D‐Array MatriXX
2 mm/2%Mean(%)95.293.090.596.379.196.898.1
STDV(%)2.91.75.41.04.02.71.5
LCL(%)89.489.780.094.471.491.595.3
3 mm/3%Mean(%)99.499.398.099.092.499.0100.0
STDV(%)1.00.52.10.92.11.30.0
LCL(%)97.498.393.997.288.396.5100
No.9999999
Per‐field measurement: averaged percentage of points passing gamma criteria of 2 mm/2% and 3 mm/3% over the institutions. Per‐field measurement: local averaged percentage of points passing gamma criteria of 2 mm/2% and 3 mm/3%, with associated confidence limits.

D.2 Composite film measurements

The composite film measurements for all mock tests, except for the multitarget, were performed by all institutions of both groups. The TOMO group measured one additional set of tests for the prostate and H&N structures. Table 8 summarizes the percentage of points passing the gamma criteria that was averaged over all institutions of both groups and the associated confidence limits. The gamma passing rate averaged over all mock test plans in the LINAC group was for 2 mm/2% criteria and for 3 mm/3% criteria. The gamma passing rate averaged over all mock test plans in the TOMO group was for 2 mm/2% criteria and for 3 mm/3% criteria. The gamma passing rate averaged over all mock test plans in all institutions was for 2 mm/2% criteria and for 3 mm/3% criteria. The TOMO group showed a higher passing rate and a lower standard deviation than the LINAC group. The local confidence limit is summarized in Table 9.
Table 8

Composite film: percentage of points passing gamma criteria of 2 mm/2% and 3 mm/3%, with associated confidence limits for LINAC/TOMO/Total groups.

Criteria Test Prostate H&N C‐shape(E) C‐shape(H) Overall
2 mm/2%Mean(%)86.5/88.1/87.285.2/86.6/85.785.4/92.1/88.881.7/88.6/83.284.7/88.4/86.1
STDV(%)6.1/2.9/4.87.3/0.9/5.78.3/7.4/8.29.1/3.2/8.57.5/3.7/6.5
CL(%)74.6/82.4/77.970.9/84.8/74.569.1/77.6/72.763.9/82.4/66.669.9/81.1/73.3
3 mm/3%Mean(%)95.4/96.4/95.894.6/96.1/95.395.1/98.0/96.093.3/95.5/94.094.6/96.4/95.3
STDV(%)3.4/3.6/3.43.4/4.2/3.74.1/1.5/3.75.3/2.1/4.64.0/3.2/3.8
CL(%)88.7/89.3/89.287.9/87.8/88.087.1/95.1/88.782.9/91.3/85.086.8/90.0/87.9
No.7/6/137/6/137/3/107/3/1028/18/46
Table 9

Composite film: percentage of points passing gamma criteria of 2 mm/2% and 3 mm/3%, averaged over the test plans, with associated confidence limits.

Criteria Group LINAC TOMO
Institution A B C D E F G H I J
2 mm/2%Mean91.181.077.284.178.995.485.086.389.089.8
LCL82.678.567.770.662.690.483.282.080.881.8
3 mm/3%Mean98.093.190.195.490.799.395.796.693.898.8
STDV1.30.73.23.44.30.61.23.02.91.7
LCL95.591.783.988.782.298.193.490.788.195.5
No.4444444666
Composite film: percentage of points passing gamma criteria of 2 mm/2% and 3 mm/3%, with associated confidence limits for LINAC/TOMO/Total groups. Composite film: percentage of points passing gamma criteria of 2 mm/2% and 3 mm/3%, averaged over the test plans, with associated confidence limits.

D.3 Correlation between the magnitude of point dose error and gamma passing rates of composite field measurements

The point dose error is an absolute value of dose difference as defined below: Figure 4 shows the magnitude of the point dose errors (low‐ and high‐dose measurements) versus composite field gamma passing rates (2 mm/2% and 3 mm/3%) for each test. The gamma passing rates for 2 mm/2% criteria ranged much broader than those for 3 mm/3%, while there was no significant difference in point dose errors between both criteria. However, data points for C‐shape hard cases were the most broadly dispersed, while data points for prostate case were the most closely confined. There was no stringent correlation between the magnitude of point dose errors and gamma passing rates for composite field measurements in this study.
Figure 4

Magnitude of the point dose errors (low dose and high dose) vs. composite filed gamma passing rates (2 mm/2% and 3 mm/3%) for each test: (a) C‐shape easy, (b) C‐shape hard, (c) H&N, and (d) prostate.

Magnitude of the point dose errors (low dose and high dose) vs. composite filed gamma passing rates (2 mm/2% and 3 mm/3%) for each test: (a) C‐shape easy, (b) C‐shape hard, (c) H&N, and (d) prostate.

IV. DISCUSSION & CONCLUSION

The multi‐institutional joint research was first performed in Korea to suggest tolerance levels for the IMRT DQA measurements. We followed the confidence limit concept presented in the AAPM TG‐119 report. Venselaar et al.( ) first suggested that the confidence limit concept was expressed with the mean value and the standard deviation (SD) multiplied by 1.5 to quantify the dose accuracy of photon beam calculations of 3D treatment planning. In AAPM TG‐119 and this study, a confidence probability () was applied. In this study, a concept of confidence limit for the point dose measurements was appropriate metrics because of a small standard deviation and a large number of samples.( ) However, for the planar dose gamma evaluation with 2 mm/2% criteria, the concept of confidence limit was not appropriate metrics because the data including the local measurements have not only a large local deviation, but were also from a small number of samples. With 3 mm/3% criteria, it was also difficult to define the confidence limit because of a small number of samples. There was no significant difference in the tolerance levels of point dose measurements between LINAC and TOMO groups. In spite of the differences in mock structures and dosimetry tools, our tolerance levels agreed with those of AAPM and ESTRO guidelines. The result can be used for a comparison guide of other institutions in Korea as they evaluate their IMRT commissioning and DQA results. In the near future, the domestic audit of IMRT DQA will be carried out by comparing institutional local values with the confidence limits determined by this study.

ACKNOWLEDGMENTS

This work was in part supported by the National Research Foundation of Korea (800‐20110212 and 490‐20120026) grant funded by the Korea government (MEST).
  18 in total

1.  Tolerances for the accuracy of photon beam dose calculations of treatment planning systems.

Authors:  J Venselaar; H Welleweerd; B Mijnheer
Journal:  Radiother Oncol       Date:  2001-08       Impact factor: 6.280

2.  Evaluation of the accuracy of 3DVH software estimates of dose to virtual ion chamber and film in composite IMRT QA.

Authors:  Arthur J Olch
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

3.  Pre-trial quality assurance processes for an intensity-modulated radiation therapy (IMRT) trial: PARSPORT, a UK multicentre Phase III trial comparing conventional radiotherapy and parotid-sparing IMRT for locally advanced head and neck cancer.

Authors:  C H Clark; E A Miles; M T Guerrero Urbano; S A Bhide; A M Bidmead; K J Harrington; C M Nutting
Journal:  Br J Radiol       Date:  2009-03-30       Impact factor: 3.039

4.  Variations of lung density and geometry on inhomogeneity correction algorithms: a Monte Carlo dosimetric evaluation.

Authors:  James C L Chow; Michael K K Leung; Jake Van Dyk
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

5.  A national dosimetric audit of IMRT.

Authors:  Geoff Budgell; Joe Berresford; Michael Trainer; Ellie Bradshaw; Peter Sharpe; Peter Williams
Journal:  Radiother Oncol       Date:  2011-05-05       Impact factor: 6.280

6.  Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors.

Authors:  Benjamin E Nelms; Heming Zhen; Wolfgang A Tomé
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

7.  Dosimetry audit for a multi-centre IMRT head and neck trial.

Authors:  Catharine H Clark; Vibeke Nordmark Hansen; Hannah Chantler; Craig Edwards; Hayley V James; Gareth Webster; Elizabeth A Miles; M Teresa Guerrero Urbano; Shree A Bhide; A Margaret Bidmead; Christoper M Nutting
Journal:  Radiother Oncol       Date:  2009-10       Impact factor: 6.280

8.  Acceptance tests and quality control (QC) procedures for the clinical implementation of intensity modulated radiotherapy (IMRT) using inverse planning and the sliding window technique: experience from five radiotherapy departments.

Authors:  Ann Van Esch; Jörg Bohsung; Pekka Sorvari; Mikko Tenhunen; Marta Paiusco; Mauro Iori; Per Engström; Håkan Nyström; Dominique Pierre Huyskens
Journal:  Radiother Oncol       Date:  2002-10       Impact factor: 6.280

9.  Accuracy of patient dose calculation for lung IMRT: A comparison of Monte Carlo, convolution/superposition, and pencil beam computations.

Authors:  Barbara Vanderstraeten; Nick Reynaert; Leen Paelinck; Indira Madani; Carlos De Wagter; Werner De Gersem; Wilfried De Neve; Hubert Thierens
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

10.  Commissioning experience and quality assurance of helical tomotherapy machines.

Authors:  Amarjit Sen; Matthew K West
Journal:  J Med Phys       Date:  2009-10
View more
  4 in total

1.  Multi-institutional evaluation using the end-to-end test for implementation of dynamic techniques of radiation therapy in Thailand.

Authors:  Lalida Tuntipumiamorn; Puangpen Tangboonduangjit; Taweap Sanghangthum; Rattapol Rangseevijitprapa; Chirasak Khamfongkhruea; Thida Niyomthai; Boonrut Vuttiprasertpong; Supranee Supanant; Nongnuch Chatchaipaiboon; Porntip Iampongpaiboon; Pitchayut Nakkrasae; Tanwiwat Jaikuna
Journal:  Rep Pract Oncol Radiother       Date:  2018-11-28

2.  Practical approach for pretreatment verification of IMRT with flattening filter free(FFF) beams using Varian Portal Dosimetry.

Authors:  Soonki Min; Young Eun Choi; Jungwon Kwak; Byungchul Cho
Journal:  J Appl Clin Med Phys       Date:  2014-01-08       Impact factor: 2.102

3.  Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea.

Authors:  Jung-In Kim; Jin-Beom Chung; Ju-Young Song; Sung Kyu Kim; Yunseok Choi; Chang Heon Choi; Won Hoon Choi; Byungchul Cho; Jin Sung Kim; Sung Jin Kim; Sung-Joon Ye
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

4.  Comparison of VMAT-SABR treatment plans with flattening filter (FF) and flattening filter-free (FFF) beam for localized prostate cancer.

Authors:  Jin-Beom Chung; Jae-Sung Kim; Keun-Yong Eom; In-Ah Kim; Sang-Won Kang; Jeong-Woo Lee; Jin-Young Kim; Tae-Suk Suh
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.