Optimization of a polarized source for in vivo x-ray fluorescence analysis of platinum and other heavy metals.

D G Lewis1.   

Abstract

The Monte Carlo method was used to optimize a polarized photon source for the x-ray fluorescence analysis of platinum and other heavy metals in vivo. The source consisted of a 140 kVp, 25 mA x-ray tube with the photons plane-polarized by 90 degrees scattering. The use of plane-polarized photons results in a significant reduction in background when the fluorescent radiation is measured along the direction of polarization. A Monte Carlo computer programme was written to simulate the production and interaction of polarized photons in order to determine the optimal polarizing material and dimensions, together with beam width and geometrical arrangement of source, polarizer and beam collimators. Calculated photon energy distributions are compared with experimental data to test the validity of the model. The best configuration of the polarization system for the in vivo analysis of platinum consisted of a 20 mm Cu polarizing block with a secondary collimator subtending a 0.1 radian angle at the polarizer.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7651997     DOI: 10.1088/0031-9155/39/1/012