Absolute quantification of left ventricular global and regional function at nuclear MPI using ultrafast CZT SPECT: initial validation versus cardiac MR.

Hubert Cochet1, Eric Bullier, Edouard Gerbaud, Muriel Durieux, Yann Godbert, Mathieu Lederlin, Pierre Coste, Jean-Louis Barat, François Laurent, Michel Montaudon.   

Abstract

UNLABELLED: We sought to evaluate the accuracy of myocardial perfusion imaging (MPI) using cadmium-zinc-telluride (CZT) SPECT cameras for the measurement of left ventricular (LV) global and regional function, and the performance of absolute wall motion (WM) and wall thickness (WT) measurements for the detection of myocardial scarring, using cardiac MR as a reference.
METHODS: Sixty patients with known or suspected coronary artery disease underwent MPI using a CZT SPECT camera, as well as cine and delayed enhanced cardiac MR. Gated MPI data were processed using quantitative gated SPECT software. LV global function was assessed by measuring LV end-diastolic volume (EDV), end-systolic volume (ESV), and ejection fraction (EF). Regional function was assessed by quantifying segmental WM and WT according to the 17-segment model. Myocardial scarring was quantified on each segment using delayed enhanced cardiac MR. Agreement between SPECT and cardiac MR was assessed using Pearson correlation and Bland-Altman analyses. The influence of measurement magnitude on the agreement was analyzed using Spearman correlation coefficients between the mean and SD of measurements. The performance of WM and WT for the detection of segments with more than 25% scar transmurality was assessed using receiver-operating-characteristic analysis.
RESULTS: Correlation between methods was excellent for EF (R = 0.81, P < 0.0001) and ESV (R = 0.88, P < 0.0001) and was good for EDV (R = 0.71, P < 0.0001). Agreement was good for EF (bias, -2.7%; limits of agreement, -15.5 to +10.1) but was lower for EDV (bias, -29.7 mL; limits of agreement, -68.3 to +8.9) and ESV (bias, -9.9 mL; limits of agreement, -30.7 to +10.9). Correlation between methods was fair for WM (R = 0.49, P < 0.0001) and WT (R = 0.48, P < 0.0001). SPECT underestimated WT (bias, -41%; limits of agreement, -108 to +26), with an error depending on thickening magnitude (ρ = 0.70, P < 0.0001). The agreement in WM measurement was higher (bias, -1.4 mm; limits of agreement, -6.9 to +4) and independent of motion magnitude (ρ = 0.006, P = 0.86). WM and WT were equally able to identify scarred segments (area under the receiver-operating-characteristic curve, 0.74 ± 0.03 and 0.74 ± 0.03, respectively).
CONCLUSION: MPI using CZT SPECT cameras and quantitative gated SPECT analysis accurately quantifies EF but still underestimates LV volumes. WM shows a higher agreement with cardiac MR than does WT, with errors in WT measurement increasing at greater thicknesses. Absolute quantification of segmental WM and WT can equally be used to identify myocardial scarring.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23385955     DOI: 10.2967/jnumed.112.110577

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  16 in total

1.  The time for radionuclide ventriculography resurrection is coming.

Authors:  Raffaele Giubbini; Elisa Milan
Journal:  J Nucl Cardiol       Date:  2015-08-08       Impact factor: 5.952

2.  Validation of cadmium-zinc-telluride camera for measurement of left ventricular systolic performance.

Authors:  Elisabeth Coupez; Charles Merlin; Viateur Tuyisenge; Laurent Sarry; Bruno Pereira; Jean René Lusson; Louis Boyer; Lucie Cassagnes
Journal:  J Nucl Cardiol       Date:  2017-02-13       Impact factor: 5.952

3.  Prospective diagnostic performance of semiconductor SPECT myocardial perfusion imaging: wall thickening analysis reduces the need for an additional prone acquisition.

Authors:  Loïc Djaileb; Benjamin Dubois; Nicolas de Leiris; Julien Leenhardt; Marjorie Canu; Olivier Phan Sy; Adrien Carabelli; Bastien Boussat; Laurent Dumas; Alexis Broisat; Gérald Vanzetto; Daniel Fagret; Catherine Ghezzi; Gilles Barone-Rochette; Laurent M Riou
Journal:  Eur J Nucl Med Mol Imaging       Date:  2019-07-18       Impact factor: 9.236

4.  First assessment of simultaneous dual isotope (123I/99mTc) cardiac SPECT on two different CZT cameras: A phantom study.

Authors:  Tanguy Blaire; Alban Bailliez; Fayçal Ben Bouallegue; Dimitri Bellevre; Denis Agostini; Alain Manrique
Journal:  J Nucl Cardiol       Date:  2017-03-08       Impact factor: 5.952

5.  Comparison of CZT SPECT and conventional SPECT for assessment of contractile function, mechanical synchrony and myocardial scar in patients with heart failure.

Authors:  Dayong Wu; Zongyao Zhang; Rongzheng Ma; Feng Guo; Lei Wang; Wei Fang
Journal:  J Nucl Cardiol       Date:  2017-06-16       Impact factor: 5.952

6.  Real-time gated-SPECT myocardial perfusion imaging with CZT detectors: A promising tool for monitoring left ventricular function.

Authors:  Roberta Assante; Alberto Cuocolo
Journal:  J Nucl Cardiol       Date:  2018-03-15       Impact factor: 5.952

7.  What is the value of motion and thickening in gated myocardial perfusion SPECT?

Authors:  Tali Sharir
Journal:  J Nucl Cardiol       Date:  2018-02-23       Impact factor: 5.952

8.  Ultrafast bone scintigraphy scan for detecting bone metastasis using a CZT whole-body gamma camera.

Authors:  Tomohiko Yamane; Atsushi Kondo; Masafumi Takahashi; Yuuki Miyazaki; Toshihiko Ehara; Kenji Koga; Ichiei Kuji; Ichiro Matsunari
Journal:  Eur J Nucl Med Mol Imaging       Date:  2019-05-01       Impact factor: 9.236

9.  Segmental and global left ventricular function assessment using gated SPECT with a semiconductor Cadmium Zinc Telluride (CZT) camera: phantom study and clinical validation vs cardiac magnetic resonance.

Authors:  Alban Bailliez; Tanguy Blaire; Frédéric Mouquet; R Legghe; B Etienne; Damien Legallois; Denis Agostini; Alain Manrique
Journal:  J Nucl Cardiol       Date:  2014-05-09       Impact factor: 5.952

10.  Pre-chemotherapy values for left and right ventricular volumes and ejection fraction by gated tomographic radionuclide angiography using a cadmium-zinc-telluride detector gamma camera.

Authors:  Christian Haarmark; Christine Haase; Maria Maj Jensen; Bo Zerahn
Journal:  J Nucl Cardiol       Date:  2015-09-04       Impact factor: 5.952

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.