On the implementation of an automated acoustic output optimization algorithm for subharmonic aided pressure estimation.

J K Dave1, V G Halldorsdottir, J R Eisenbrey, D A Merton, J B Liu, P Machado, H Zhao, S Park, S Dianis, C L Chalek, K E Thomenius, D B Brown, F Forsberg.   

Abstract

Incident acoustic output (IAO) dependent subharmonic signal amplitudes from ultrasound contrast agents can be categorized into occurrence, growth or saturation stages. Subharmonic aided pressure estimation (SHAPE) is a technique that utilizes growth stage subharmonic signal amplitudes for hydrostatic pressure estimation. In this study, we developed an automated IAO optimization algorithm to identify the IAO level eliciting growth stage subharmonic signals and also studied the effect of pulse length on SHAPE. This approach may help eliminate the problems of acquiring and analyzing the data offline at all IAO levels as was done in previous studies and thus, pave the way for real-time clinical pressure monitoring applications. The IAO optimization algorithm was implemented on a Logiq 9 (GE Healthcare, Milwaukee, WI) scanner interfaced with a computer. The optimization algorithm stepped the ultrasound scanner from 0% to 100% IAO. A logistic equation fitting function was applied with the criterion of minimum least squared error between the fitted subharmonic amplitudes and the measured subharmonic amplitudes as a function of the IAO levels and the optimum IAO level was chosen corresponding to the inflection point calculated from the fitted data. The efficacy of the optimum IAO level was investigated for in vivo SHAPE to monitor portal vein (PV) pressures in 5 canines and was compared with the performance of IAO levels, below and above the optimum IAO level, for 4, 8 and 16 transmit cycles. The canines received a continuous infusion of Sonazoid microbubbles (1.5 μl/kg/min; GE Healthcare, Oslo, Norway). PV pressures were obtained using a surgically introduced pressure catheter (Millar Instruments, Inc., Houston, TX) and were recorded before and after increasing PV pressures. The experiments showed that optimum IAO levels for SHAPE in the canines ranged from 6% to 40%. The best correlation between changes in PV pressures and in subharmonic amplitudes (r=-0.76; p=0.24), and between the absolute PV pressures and the subharmonic amplitudes (r=-0.89; p<0.01) were obtained for the optimized IAO and 4 transmit cycles. Only for the optimized IAO and 4 transmit cycles did the subharmonic amplitudes differ significantly (p<0.01) before and after increasing PV pressures. A new algorithm to identify optimum IAO levels for SHAPE has been developed and validated with the best results being obtained for 4 transmit cycles. The work presented in this study may pave the way for real-time clinical applications of estimating pressures using the subharmonic signals from ultrasound contrast agents.
Copyright © 2012 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23347593      PMCID: PMC3595343          DOI: 10.1016/j.ultras.2012.12.010

Source DB:  PubMed          Journal:  Ultrasonics        ISSN: 0041-624X            Impact factor:   2.890


  45 in total

1.  Pressure dependence of subharmonic signals from contrast microbubbles.

Authors:  W T Shi; F Forsberg; J S Raichlen; L Needleman; B B Goldberg
Journal:  Ultrasound Med Biol       Date:  1999-02       Impact factor: 2.998

2.  Noninvasive estimation of dynamic pressures in vitro and in vivo using the subharmonic response from microbubbles.

Authors:  Jaydev K Dave; Valgerdur G Halldorsdottir; John R Eisenbrey; Ji-Bin Liu; Maureen E McDonald; Kris Dickie; Corina Leung; Flemming Forsberg
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2011-10       Impact factor: 2.725

3.  Time and pressure dependence of acoustic signals radiated from microbubbles.

Authors:  Yaniv Ganor; Dan Adam; Eitan Kimmel
Journal:  Ultrasound Med Biol       Date:  2005-10       Impact factor: 2.998

Review 4.  Animal models of portal hypertension.

Authors:  Juan-G Abraldes; Marcos Pasarín; Juan-Carlos García-Pagán
Journal:  World J Gastroenterol       Date:  2006-11-07       Impact factor: 5.742

5.  Liver stiffness measurement predicts severe portal hypertension in patients with HCV-related cirrhosis.

Authors:  Francesco Vizzutti; Umberto Arena; Roberto G Romanelli; Luigi Rega; Marco Foschi; Stefano Colagrande; Antonio Petrarca; Stefania Moscarella; Giacomo Belli; Anna Linda Zignego; Fabio Marra; Giacomo Laffi; Massimo Pinzani
Journal:  Hepatology       Date:  2007-05       Impact factor: 17.425

6.  In vivo pressure estimation using subharmonic contrast microbubble signals: proof of concept.

Authors:  Flemming Forsberg; Ji-Bin Liu; William T Shi; Junji Furuse; Masafumi Shimizu; Barry B Goldberg
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2005-04       Impact factor: 2.725

7.  Doppler sonography of the portal vein and hepatic artery: measurement of a prandial effect in healthy subjects.

Authors:  A J Fisher; E K Paulson; M A Kliewer; D M DeLong; R C Nelson
Journal:  Radiology       Date:  1998-06       Impact factor: 11.105

8.  A new noninvasive technique for cardiac pressure measurement: resonant scattering of ultrasound from bubbles.

Authors:  W M Fairbank; M O Scully
Journal:  IEEE Trans Biomed Eng       Date:  1977-03       Impact factor: 4.538

9.  Subharmonic contrast microbubble signals for noninvasive pressure estimation under static and dynamic flow conditions.

Authors:  Valgerdur G Halldorsdottir; Jaydev K Dave; Lauren M Leodore; John R Eisenbrey; Suhyun Park; Anne L Hall; Kai Thomenius; Flemming Forsberg
Journal:  Ultrason Imaging       Date:  2011-07       Impact factor: 1.578

10.  Damping index of Doppler hepatic vein waveform to assess the severity of portal hypertension and response to propranolol in liver cirrhosis: a prospective nonrandomized study.

Authors:  Moon Young Kim; Soon Koo Baik; Dong Hun Park; Dae Wook Lim; Jae Woo Kim; Hyun Soo Kim; Sang Ok Kwon; Young Ju Kim; Sei Jin Chang; Samuel S Lee
Journal:  Liver Int       Date:  2007-10       Impact factor: 5.828

View more
  11 in total

1.  Three-Dimensional Subharmonic Aided Pressure Estimation for Assessing Arterial Plaques in a Rabbit Model.

Authors:  Kibo Nam; Ji-Bin Liu; John R Eisenbrey; Maria Stanczak; Priscilla Machado; Jingzhi Li; Zhaojun Li; Ying Wei; Flemming Forsberg
Journal:  J Ultrasound Med       Date:  2018-12-17       Impact factor: 2.153

2.  Cavitation thresholds of contrast agents in an in vitro human clot model exposed to 120-kHz ultrasound.

Authors:  Matthew J Gruber; Kenneth B Bader; Christy K Holland
Journal:  J Acoust Soc Am       Date:  2014-02       Impact factor: 1.840

Review 3.  Portal pressure monitoring-state-of-the-art and future perspective.

Authors:  Gang Xu; Fei Li; Yilei Mao
Journal:  Ann Transl Med       Date:  2019-10

4.  Relationship between cavitation and loss of echogenicity from ultrasound contrast agents.

Authors:  Kirthi Radhakrishnan; Kenneth B Bader; Kevin J Haworth; Jonathan A Kopechek; Jason L Raymond; Shao-Ling Huang; David D McPherson; Christy K Holland
Journal:  Phys Med Biol       Date:  2013-09-04       Impact factor: 3.609

5.  Monitoring Neoadjuvant Chemotherapy for Breast Cancer by Using Three-dimensional Subharmonic Aided Pressure Estimation and Imaging with US Contrast Agents: Preliminary Experience.

Authors:  Kibo Nam; John R Eisenbrey; Maria Stanczak; Anush Sridharan; Adam C Berger; Tiffany Avery; Juan P Palazzo; Flemming Forsberg
Journal:  Radiology       Date:  2017-05-03       Impact factor: 11.105

6.  Subharmonic-Aided Pressure Estimation for Monitoring Interstitial Fluid Pressure in Tumors: Calibration and Treatment with Paclitaxel in Breast Cancer Xenografts.

Authors:  Valgerdur G Halldorsdottir; Jaydev K Dave; Andrew Marshall; Anya I Forsberg; Traci B Fox; John R Eisenbrey; Priscilla Machado; Ji-Bin Liu; Daniel A Merton; Flemming Forsberg
Journal:  Ultrasound Med Biol       Date:  2017-04-19       Impact factor: 2.998

7.  Effect of Pulse Shaping on Subharmonic Aided Pressure Estimation In Vitro and In Vivo.

Authors:  Ipshita Gupta; John Eisenbrey; Maria Stanczak; Anush Sridharan; Jaydev K Dave; Ji-Bin Liu; Christopher Hazard; Xinghua Wang; Ping Wang; Huiwen Li; Kirk Wallace; Flemming Forsberg
Journal:  J Ultrasound Med       Date:  2016-12-10       Impact factor: 2.153

8.  Developing an Interface and Investigating Optimal Parameters for Real-Time Intracardiac Subharmonic-Aided Pressure Estimation.

Authors:  Cara Esposito; Kris Dickie; Flemming Forsberg; Jaydev K Dave
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2021-02-25       Impact factor: 2.725

9.  Optimal Control of SonoVue Microbubbles to Estimate Hydrostatic Pressure.

Authors:  Amanda Q X Nio; Alessandro Faraci; Kirsten Christensen-Jeffries; Jason L Raymond; Mark J Monaghan; Daniel Fuster; Flemming Forsberg; Robert J Eckersley; Pablo Lamata
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2019-10-21       Impact factor: 2.725

10.  Hepatic Vein Contrast-Enhanced Ultrasound Subharmonic Imaging Signal as a Screening Test for Portal Hypertension.

Authors:  Priscilla Machado; Ipshita Gupta; Sriharsha Gummadi; Maria Stanczak; Corinne E Wessner; Jonathan M Fenkel; Colette M Shaw; Susan Shamini-Noori; Susan Schultz; Michael C Soulen; Chandra M Sehgal; Kirk Wallace; John R Eisenbrey; Flemming Forsberg
Journal:  Dig Dis Sci       Date:  2021-01-04       Impact factor: 3.487

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.