Ahmed Assaf1, Maged Maher Roshdy.
Abstract
PURPOSE: This paper compares and evaluates the corneal morphological changes occurring after cataract surgery through a 2.2 mm corneal incision. We use two platforms for comparison and evaluation, transversal and torsional phacoemulsification. PATIENTS AND METHODS: This study includes 139 consecutive cataractous eyes (nuclear color 2-4, according to the Lens Opacities Classification System III [LOCSIII]) of 82 patients undergoing cataract surgery through a 2.2 mm corneal incision. Two different phacoemulsification platforms were used and assigned randomly: we used the WhiteStar Signature(®) system with the Ellips™ FX transversal continuous ultrasound (US) mode for group I (mean age: 65.33 ± 6.97 years), and we used the Infiniti(®) system with the OZil(®) Intelligent Phaco (IP) torsional US mode for group II (mean age: 64.02 ± 7.55 years). The corneal endothelium and pachymetry were evaluated preoperatively and at 1 month postoperatively. Incision size changes were also evaluated.Entities:
Keywords: Ellips™ FX; OZil® IP; cornea; endothelium; torsional phacoemulsification; transversal phacoemulsification
Year: 2013 PMID: 23326184 PMCID: PMC3544352 DOI: 10.2147/OPTH.S39019
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Summary of the corneal endothelial changes analyzed by specular microscopy in the two groups analyzed in the current study: group I, eyes undergoing phacoemulsification using the WhiteStar Signature® platform and the Ellips™ FX (Abbott Medical Optics, Santa Ana, CA, USA) transversal continuous ultrasound mode; and group II, eyes undergoing phacoemulsification using the Infiniti® OZil® Intelligent Phaco (Bausch and Lomb, Rochester, NY, USA) torsional ultrasound mode
| Endothelial parameters | Mean (SD), median (range) | ||
|---|---|---|---|
|
| |||
| Preoperative | 1 month postoperative | ||
| ECD (cell/mm2) | 2304.1 (122.5) | 2264.1 (129.3) | <0.01 |
| 2303.0 (1950–2588) | 2279.0 (1924–2567) | ||
| CV | 38.2 (3.5) | 38.7 (3.2) | 0.02 |
| 38.0 (33–45) | 39.0 (31–46) | ||
| HEX (%) | 39.7 (3.1) | 38.7 (3.3) | <0.01 |
| 39.0 (33–47) | 39.0 (33–45) | ||
| ECD (cell/mm2) | 2315.6 (83.1) | 2270.3 (89.9) | <0.01 |
| 2318.0 (2067–2466) | 2270.0 (1990–2450) | ||
| CV | 38.0 (3.2) | 38.9 (2.6) | <0.01 |
| 38.0 (33–45) | 39.0 (33–44) | ||
| HEX (%) | 39.1 (3.2) | 38.6 (3.1) | 0.07 |
| 39.0 (33–45) | 39.0 (32–45) | ||
Abbreviations: SD, standard deviation; ECD, endothelial cell density; CV, coefficient of variation; HEX, hexagonality.
Figure 1Scatter diagram of postoperative endothelial cells in relation to preoperative values in the two groups analyzed in the current study: group I, transversal phaco with Ellips™ FX (Abbott Medical Optics, Santa Ana, CA, USA); and group II, torsional phaco with OZil® Intelligent Phaco (Bausch and Lomb, Rochester, NY, USA).