Combined occlusion-triggered longitudinal and torsional phacoemulsification during coaxial microincision cataract surgery: effect on 30-degree mini-flared tip behavior.

Gokulan Ratnarajan1, Richard Packard, Mina Ward.   

Abstract

PURPOSE: To study the impact of a new power-delivery option using occlusion-triggered bursts of longitudinal phaco energy in addition to torsional tip movement in the performance of the 30-degree mini-flared tip to minimize tip clogging with dense nuclei.
SETTING: Prince Charles Eye Unit, King Edward VII Hospital, Windsor, United Kingdom.
DESIGN: Comparative case series.
METHOD: The cumulative dissipated energy (CDE) and time in foot pedal position 3 (FP3) to phacoemulsify each half of the nucleus with and without use of the new power-delivery option (Intelligent Phaco) were compared.
RESULTS: Fifty consecutive cataract operations were reviewed. The mean CDE was 1.79 with the new power-delivery option and 2.03 without the option (P=.08). For grades I, II, III, and IV nuclei, the corresponding values were 0.41 and 0.51 (P=.16), 1.43 and 1.23 (P=.31), 2.32 and 2.92 (P=.12), 3.82 and 5.03 (P=.005). The mean FP 3 time was 5.40 seconds with the new power-delivery option and 6.35 seconds without it (P=.03). For grades I, II, III, and IV nuclei, the corresponding values were 1.53 seconds and 1.98 seconds (P=.13), 4.47 seconds and 4.26 seconds (P=.69), 6.51 seconds and 8.77 seconds (P=.05), 11.81 seconds and 16.49 seconds (P=.008).
CONCLUSIONS: The new power-delivery option appeared to reduce the CDE and FP 3 time when a 30-degree mini-flared tip was used. The benefit accrued by preventing clogging of the needle was significant only in cases with a denser nucleus.
Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21440411     DOI: 10.1016/j.jcrs.2010.11.038

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  4 in total

1.  The Effect of Cumulative Dissipated Energy on Changes in Intraocular Pressure After Uncomplicated Cataract Surgery by Phacoemulsification.

Authors:  Roland Y Lee; Rebecca I Chen; Toshimitsu Kasuga; Qi N Cui; Travis C Porco; Shan C Lin
Journal:  J Glaucoma       Date:  2016-07       Impact factor: 2.503

2.  Comparison of 45-degree Kelman and 45-degree balanced phaco tip designs in torsional microcoaxial phacoemulsification.

Authors:  Süleyman Demircan; Mustafa Ataş; Emre Göktaş; Burhan Başkan
Journal:  Int J Ophthalmol       Date:  2015-12-18       Impact factor: 1.779

3.  Impact of intelligent phacoemulsification software on torsional phacoemulsification surgery.

Authors:  Silay Canturk Ugurbas; Sinan Caliskan; Atilla Alpay; Suat Hayri Ugurbas
Journal:  Clin Ophthalmol       Date:  2012-09-12

4.  Comparative analysis of corneal morphological changes after transversal and torsional phacoemulsification through 2.2 mm corneal incision.

Authors:  Ahmed Assaf; Maged Maher Roshdy
Journal:  Clin Ophthalmol       Date:  2013-01-08
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.