Literature DB >> 20457377

Thermal study of longitudinal and torsional ultrasound phacoemulsification: tracking the temperature of the corneal surface, incision, and handpiece.

Bokkwan Jun1, John P Berdahl, Terry Kim.   

Abstract

PURPOSE: To evaluate the change and difference in the corneal surface, incision, and handpiece temperatures during longitudinal and torsional ultrasound (US) phacoemulsification with standard incisions (2.75 mm) and microincisions (2.20 mm) and the thermal effect on wounds.
SETTING: Department of Ophthalmology, Duke University, Durham, North Carolina, USA.
METHODS: In this prospective study, human cadaver eyes had simulated phacoemulsification. Group 1 had a 2.75 mm incision with 100% longitudinal US; Group 2, a 2.20 mm incision with 100% longitudinal US; Group 3, a 2.75 mm incision with 100% torsional US; and Group 4, a 2.20 mm incision with 100% torsional US. During phacoemulsification, the corneal incision was evaluated by surgical microscopy and scanning electron microscopy (SEM) and images of the corneal surface, incision, and handpiece were captured with an infrared camera.
RESULTS: Twelve eyes (3 each group) were evaluated. The maximum incision temperature was higher in the longitudinal groups than in the torsional groups. With the same US modality, the maximum microincision temperature was higher than the maximum standard incision temperature. After application of full power for 40 seconds, wound burn was observed in all eyes in the longitudinal groups and no eyes in the torsional groups. On SEM, there was more extensive loss of Descemet membrane in the longitudinal groups than in the torsional groups.
CONCLUSION: Incision temperature was influenced by US modality and was significantly lower with torsional US than with longitudinal US. Using torsional US with smaller incisions may decrease the risk for wound burn in eyes with denser cataracts. (c) 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20457377     DOI: 10.1016/j.jcrs.2009.11.015

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  Investigating the ocular temperature rise during femtosecond laser lens fragmentation: an in vitro study.

Authors:  Rita Mencucci; Sara Matteoli; Andrea Corvi; Luca Terracciano; Eleonora Favuzza; Stefano Gherardini; Filippo Caruso; Roberto Bellucci
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-10-21       Impact factor: 3.117

2.  Comparison of clear corneal incision injuries between torsional and conventional phacoemulsification.

Authors:  Xu Chen; Yinghong Ji; Yi Lu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-07-20       Impact factor: 3.117

3.  Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines.

Authors:  Mustafa Ataş; Süleyman Demircan; Arzu Seyhan Karatepe Haşhaş; Ahmet Gülhan; Gökmen Zararsız
Journal:  Int J Ophthalmol       Date:  2014-10-18       Impact factor: 1.779

4.  A novel method to compare phacoemulsification parameters in vivo: two halves of one nucleus.

Authors:  Elena Tomilova; Sergey Shukhaev
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-06-21       Impact factor: 3.117

5.  Laboratory assessment of thermal characteristics of three phacoemulsification tip designs operated using torsional ultrasound.

Authors:  Jaime Zacharias
Journal:  Clin Ophthalmol       Date:  2016-06-13

6.  Wound and surface temperatures in vivo in torsional and longitudinal modalities of ultrasound in coaxial microincisional cataract surgery.

Authors:  Swapna Nair; Ramachandran Unnikrishnan Nair
Journal:  Clin Ophthalmol       Date:  2017-01-27

7.  Comparisons of the energy efficiency and intraocular safety of two torsional phacoemulsification tips.

Authors:  Yan-Xiu Sun; Rong Cao; Zi-Yuan Liu; Hua-Qin Xia; Yu-Jie Cen; Lu Gao; Dan-Dan Shi
Journal:  BMC Ophthalmol       Date:  2022-10-03       Impact factor: 2.086

8.  Comparative analysis of corneal morphological changes after transversal and torsional phacoemulsification through 2.2 mm corneal incision.

Authors:  Ahmed Assaf; Maged Maher Roshdy
Journal:  Clin Ophthalmol       Date:  2013-01-08
  8 in total