Scatter correction in cone-beam CT via a half beam blocker technique allowing simultaneous acquisition of scatter and image information.

Ho Lee1, Lei Xing, Rena Lee, Benjamin P Fahimian.   

Abstract

PURPOSE: X-ray scatter incurred to detectors degrades the quality of cone-beam computed tomography (CBCT) and represents a problem in volumetric image guided and adaptive radiation therapy. Several methods using a beam blocker for the estimation and subtraction of scatter have been proposed. However, due to missing information resulting from the obstruction of the blocker, such methods require dual scanning or dynamically moving blocker to obtain a complete volumetric image. Here, we propose a half beam blocker-based approach, in conjunction with a total variation (TV) regularized Feldkamp-Davis-Kress (FDK) algorithm, to correct scatter-induced artifacts by simultaneously acquiring image and scatter information from a single-rotation CBCT scan.
METHODS: A half beam blocker, comprising lead strips, is used to simultaneously acquire image data on one side of the projection data and scatter data on the other half side. One-dimensional cubic B-Spline interpolation/extrapolation is applied to derive patient specific scatter information by using the scatter distributions on strips. The estimated scatter is subtracted from the projection image acquired at the opposite view. With scatter-corrected projections where this subtraction is completed, the FDK algorithm based on a cosine weighting function is performed to reconstruct CBCT volume. To suppress the noise in the reconstructed CBCT images produced by geometric errors between two opposed projections and interpolated scatter information, total variation regularization is applied by a minimization using a steepest gradient descent optimization method. The experimental studies using Catphan504 and anthropomorphic phantoms were carried out to evaluate the performance of the proposed scheme.
RESULTS: The scatter-induced shading artifacts were markedly suppressed in CBCT using the proposed scheme. Compared with CBCT without a blocker, the nonuniformity value was reduced from 39.3% to 3.1%. The root mean square error relative to values inside the regions of interest selected from a benchmark scatter free image was reduced from 50 to 11.3. The TV regularization also led to a better contrast-to-noise ratio.
CONCLUSIONS: An asymmetric half beam blocker-based FDK acquisition and reconstruction technique has been established. The proposed scheme enables simultaneous detection of patient specific scatter and complete volumetric CBCT reconstruction without additional requirements such as prior images, dual scans, or moving strips.

Entities:  

Mesh:

Year:  2012        PMID: 22559608     DOI: 10.1118/1.3691901

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Single-scan patient-specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval.

Authors:  Bowen Meng; Ho Lee; Lei Xing; Benjamin P Fahimian
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

2.  A model-based scatter artifacts correction for cone beam CT.

Authors:  Wei Zhao; Don Vernekohl; Jun Zhu; Luyao Wang; Lei Xing
Journal:  Med Phys       Date:  2016-04       Impact factor: 4.071

3.  Optimization of the geometry and speed of a moving blocker system for cone-beam computed tomography scatter correction.

Authors:  Xi Chen; Luo Ouyang; Hao Yan; Xun Jia; Bin Li; Qingwen Lyu; You Zhang; Jing Wang
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

4.  Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter - Part I: Core algorithms and validation.

Authors:  Alexander Maslowski; Adam Wang; Mingshan Sun; Todd Wareing; Ian Davis; Josh Star-Lack
Journal:  Med Phys       Date:  2018-04-06       Impact factor: 4.071

5.  Image reconstruction and scan configurations enabled by optimization-based algorithms in multispectral CT.

Authors:  Buxin Chen; Zheng Zhang; Emil Y Sidky; Dan Xia; Xiaochuan Pan
Journal:  Phys Med Biol       Date:  2017-11-02       Impact factor: 3.609

6.  4D cone-beam computed tomography (CBCT) using a moving blocker for simultaneous radiation dose reduction and scatter correction.

Authors:  Cong Zhao; Yuncheng Zhong; Xinhui Duan; You Zhang; Xiaokun Huang; Jing Wang; Mingwu Jin
Journal:  Phys Med Biol       Date:  2018-05-29       Impact factor: 3.609

7.  Radiation dose to procedural personnel and patients from an X-ray volume imaging system.

Authors:  Jijo Paul; Emmanuel C Mbalisike; Thomas J Vogl
Journal:  Eur Radiol       Date:  2013-06-29       Impact factor: 5.315

Review 8.  Modelling the physics in the iterative reconstruction for transmission computed tomography.

Authors:  Johan Nuyts; Bruno De Man; Jeffrey A Fessler; Wojciech Zbijewski; Freek J Beekman
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

9.  Shading correction for volumetric CT using deep convolutional neural network and adaptive filter.

Authors:  Xiaokun Liang; Na Li; Zhicheng Zhang; Shaode Yu; Wenjian Qin; Yafen Li; Shupeng Chen; Huailing Zhang; Yaoqin Xie
Journal:  Quant Imaging Med Surg       Date:  2019-07

10.  Robust moving-blocker scatter correction for cone-beam computed tomography using multiple-view information.

Authors:  Cong Zhao; Xi Chen; Luo Ouyang; Jing Wang; Mingwu Jin
Journal:  PLoS One       Date:  2017-12-21       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.