| Literature DB >> 36267985 |
Yan Zhang1, Han Zhou1, Kaiyue Chu2, Chuanfeng Wu3, Yun Ge1, Guoping Shan1,4, Jundong Zhou3, Jing Cai2, Jianhua Jin2, Weiyu Sun1, Ying Chen1, Xiaolin Huang1.
Abstract
Background: The setup accuracy plays an extremely important role in the local control of tumors. The purpose of this study is to verify the feasibility of "Sphere-Mask" Optical Positioning System (S-M_OPS) for fast and accurate setup.Entities:
Keywords: CBCT; S-M_OPS; laser line; radiotherapy; setup
Year: 2022 PMID: 36267985 PMCID: PMC9577199 DOI: 10.3389/fonc.2022.918296
Source DB: PubMed Journal: Front Oncol ISSN: 2234-943X Impact factor: 5.738
Figure 1The workflow of S-M_OPS.
Figure 2Reference positions of the positioning spheres in different parts (Upper left: top view; Upper right: skeletal diagram; Lower left: left view; Lower right: right view).
Figure 3The workflow of clinical co-setup experiment.
Figure 4Schematic diagram of CBCT tumor center, laser line tumor center, S-M_OPS tumor center, D and D.
Figure 5(A) Histogram of setup time of laser line; (B) Histogram of mask deformation; (C) Histogram of setup error in the LAT direction; (D) Histogram of setup error in the VRT direction; (E) Histogram of setup error in the LNG direction.
Figure 6Distributions of D and D in the LAT (left/right), VRT (superior/inferior) and LNG (anterior/posterior) and D ( ) directions. (A) D -LAT; (B) D -LAT; (C) D -VRT; (D) D -VRT ; (E) D -LNG; (F) D -LNG; (G) D -D; (H) D -D.
Comparisons among laser line-guided setup errors and S-M_OPS-guided setup errors.
| LAT | VRT | LNG | D | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Mean ± SD |
| Mean ± SD |
| Mean ± SD |
| Mean ± SD |
| ||
| All | Laser | -0.21 ± 3.13 | <10-15 | 1.02 ± 2.76 | <10-11 | 2.22 ± 4.26 | <10-24 | 5.36 ± 3.58 | <10-17 |
| S-M_OPS | 0.12 ± 1.91 | 1.02 ± 1.81 | -0.10 ± 2.25 | 2.94 ± 2.09 | |||||
| Intracalvarium | Laser | -1.24 ± 1.52 | 0.01 | -0.22 ± 1.52 | 0.21 | 2.00 ± 2.56 | <10-4 | 3.67 ± 1.72 | <10-3 |
| S-M_OPS | -0.66 ± 0.94 | 0.32 ± 1.29 | -0.03 ± 1.09 | 1.85 ± 0.84 | |||||
| Nasopharynx | Laser | -1.20 ± 1.40 | 0.02 | 0.13 ± 1.91 | 0.03 | 2.09 ± 3.00 | <10-6 | 3.92 ± 2.15 | <10-2 |
| S-M_OPS | -0.64 ± 0.88 | 1.25 ± 1.26 | -0.06 ± 0.89 | 1.93 ± 1.15 | |||||
| Esophagus | Laser | -0.30 ± 3.29 | <10-2 | 1.05 ± 2.36 | 0.29 | 2.18 ± 3.46 | <10-8 | 5.10 ± 2.83 | <10-2 |
| S-M_OPS | 0.23 ± 2.32 | 0.95 ± 2.22 | -0.54 ± 1.78 | 3.15 ± 2.17 | |||||
| Lung | Laser | -1.65 ± 3.83 | <10-5 | 1.39 ± 2.21 | 0.06 | 2.91 ± 3.34 | 0.44 | 5.84 ± 3.02 | 0.45 |
| S-M_OPS | -0.40 ± 1.86 | 1.21 ± 1.68 | 0.03 ± 3.43 | 3.15 ± 3.08 | |||||
| Liver | Laser | 2.79 ± 2.10 | 0.06 | 2.02 ± 1.91 | 0.05 | 6.43 ± 5.15 | <10-7 | 7.76 ± 5.22 | <10-4 |
| S-M_OPS | 1.87 ± 1.51 | 1.67 ± 1.37 | 2.04 ± 1.55 | 3.54 ± 2.09 | |||||
| Abdomen | Laser | -0.28 ± 2.52 | 0.12 | 1.40 ± 3.98 | <10-3 | 0.82 ± 3.28 | 0.43 | 4.93 ± 3.25 | 0.02 |
| S-M_OPS | -0.29 ± 2.01 | -0.04 ± 2.01 | -0.26 ± 3.17 | 3.64 ± 2.14 | |||||
| Cervix | Laser | -0.25 ± 3.16 | <10-10 | 0.98 ± 3.49 | <10-10 | 1.00 ± 5.38 | <10-13 | 5.79 ± 4.37 | <10-13 |
| S-M_OPS | -0.73 ± 1.35 | 1.39 ± 1.45 | -0.41 ± 1.94 | 2.78 ± 1.58 | |||||
Values are shown in mean ± SD and p-value (F-test).
Figure 795% Confidence intervals of D and D in different parts in the LAT, VRT, LNG and D directions. (A) Intracalvarium; (B) Nasopharynx; (C) Esophagus; (D) Lung; (E) Liver (F) Abdomen; (G) Cervix.
Figure 8Comparison between laser line-guide setup consistency and S-M_OPS-guide setup consistency in different parts in the LAT, VRT, and LNG directions (A) Intracalvarium; (B) Nasopharynx; (C) Esophagus; (D) Lung; (E) Liver (F) Abdomen; (G) Cervix.
Comparisons among S-M_OPS-guided setup errors, Sentinel-guided setup errors, Catalyst-guided setup errors and ExacTrac-guided setup errors.
| Head and neck/mm | Thorax/mm | Pelvis/mm | Overall/mm | ||
|---|---|---|---|---|---|
| LAT | S-M_OPS | 0.1 ± 2.1 | -0.1 ± 1.9 | -0.6 ± 1.6 | -0.2 ± 1.9 |
| Sentinel | 0.9 ± 1.8 | 1.2 ± 3.6 | -2.5 ± 4.1 | -1.0 ± 3.6 | |
| Catalyst | 0.3 ± 2.4 | 0.7 ± 3.3 | 1.2 ± 2.5 | -0.7 ± 2.8 | |
| ExacTrac | 4.1 ± 2.7 | 0.1 ± 1.8 | -0.6 ± 2.7 | N/A | |
| VRT | S-M_OPS | 1.0 ± 2.1 | 1.1 ± 1.8 | 1.0 ± 1.8 | 1.0 ± 1.9 |
| Sentinel | -2.7 ± 3.8 | 0.8 ± 5.1 | -4.6 ± 7.3 | 1.0 ± 6.3 | |
| Catalyst | -3.7 ± 3.4 | -0.7 ± 3.8 | 0.2 ± 3.7 | -1.3 ± 4.0 | |
| ExacTrac | 1.2 ± 0.5 | 0.2 ± 2.4 | -0.3 ± 2.3 | N/A | |
| LNG | S-M_OPS | -0.4 ± 1.7 | 0.2 ± 3.2 | -0.4 ± 2.4 | -0.3 ± 2.3 |
| Sentinel | -0.8 ± 3.6 | 0.8 ± 4.3 | -5.1 ± 7.4 | -1.8 ± 5.9 | |
| Catalyst | -0.2 ± 3.4 | 2.4 ± 3.2 | 1.8 ± 3.7 | 1.5 ± 3.6 | |
| ExacTrac | 1.1 ± 0.7 | -0.6 ± 1.8 | 1.6 ± 3.5 | N/A | |
Values are shown in mean ± SD with best shown with shading.
NA, not applicable.