Iqbal Al Amri1, Ramamoorthy Ravichandran, Somangili Satyamoorthi Sivakumar, Johnson Pichi Binukumar, Chirayathmanjiyil Antony Davis, Zakia Al Rahbi, Khalsa Al Shukeili, Fatima Al Kindi.
Abstract
Inversely planned intensity-modulated radiotherapy (IMRT) and stereotactic small field radiotherapy should be verified before treatment execution. A second verification is carried out for planned treatments in IMRT and 3D conformal radiotherapy (3D-CRT) using a monitor verification commercial dose calculation management software (DCMS). For the same reference point the ion-chamber measured doses are compared for IMRT plans. DCMS (Diamond) computes dose based on modified Clarkson integration, accounting for multi-leaf collimators (MLC) transmission and measured collimator scatter factors. DCMS was validated with treatment planning system (TPS) (Eclipse 6.5 Version, Varian, USA) separately. Treatment plans computed from TPS are exported to DCMS using DICOM interface. Doses are re-calculated at selected points for fields delivered to IMRT phantom (IBA Scanditronix Wellhofer) in high-energy linac (Clinac 2300 CD, Varian). Doses measured at central axis, for the same points using CC13 (0.13 cc) ion chamber with Dose 1 Electrometer (Scanditronix Wellhofer) are compared with calculated data on DCMS and TPS. The data of 53 IMRT patients with fields ranging from 5 to 9 are reported. The computed dose for selected monitor units (MU) by Diamond showed good agreement with planned doses by TPS. DCMS dose prediction matched well in 3D-CRT forward plans (0.8 ± 1.3%, n = 37) and in IMRT inverse plans (-0.1 ± 2.2%, n = 37). Ion chamber measurements agreed well with Eclipse planned doses (-2.1 ± 2.0%, n = 53) and re-calculated DCMS doses (-1.5 ± 2.6%, n = 37) in phantom. DCMS dose validation is in reasonable agreement with TPS. DCMS calculations corroborate well with ionometric measured doses in most of the treatment plans.Entities:
Keywords: Diamond software; IMRT verification; dose accuracy in RT; ionometry
Year: 2012 PMID: 23293456 PMCID: PMC3532753 DOI: 10.4103/0971-6203.103610
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Comparison of calculated dose by DCMS (open fields)
Comparison of calculated doses for irregular shaped fields (blocks and MLC shaping)
Comparison of calculated doses (Eclipse and DCMS) and measured doses
Comparison of calculated doses on other MU check software[13]
Comparison of calculated dose for 6 MV by DCMS (physical wedge fields)
Comparison of calculated dose for 15 MV by DCMS (physical wedge fields)
Comparison of calculated doses for changed source skin distances