Comparison of monitor units calculated by radiotherapy treatment planning system and an independent monitor unit verification software.

P Sellakumar1, C Arun, S S Sanjay, S B Ramesh.   

Abstract

In radiation therapy, the monitor units (MU) needed to deliver a treatment plan are calculated by treatment planning systems (TPS). The essential part of quality assurance is to verify the MU with independent monitor unit calculation to correct any potential errors prior to the start of treatment. In this study, we have compared the MU calculated by TPS and by independent MU verification software. The MU verification software was commissioned and tested for the data integrity to ensure that the correct beam data was considered for MU calculations. The accuracy of the calculations was tested by creating a series of test plans and comparing them with ion chamber measurements. The results show that there is good agreement between the two. The MU difference (MUdiff) between the monitor unit calculations of TPS and independent MU verification system was calculated for 623 fields from 245 patients and was analyzed by treatment site for head &amp; neck, thorax, breast, abdomen and pelvis. The mean MUdiff of -0.838% with a standard deviation of 3.04% was observed for all 623 fields. The site specific standard deviation of MUdiff was as follows: abdomen and pelvis (<1.75%), head &amp; neck (2.5%), thorax (2.32%) and breast (6.01%). The disparities were analyzed and different correction methods were used to reduce the disparity.
© 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20172756     DOI: 10.1016/j.ejmp.2010.01.006

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

1.  Validation and clinical implementation of commercial secondary check software with heterogeneity corrections.

Authors:  Vijayalakshmi Kuppusamy; Vivekanandan Nagarajan; Lavanya Murugan
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-16

2.  Radiotherapy pre-treatment dose validation: A second verification of monitor units (MU) with a commercial software.

Authors:  Iqbal Al Amri; Ramamoorthy Ravichandran; Somangili Satyamoorthi Sivakumar; Johnson Pichi Binukumar; Chirayathmanjiyil Antony Davis; Zakia Al Rahbi; Khalsa Al Shukeili; Fatima Al Kindi
Journal:  J Med Phys       Date:  2012-10

3.  A clinical database to assess action levels and tolerances for the ongoing use of Mobius3D.

Authors:  David Jolly; Leon Dunn; John Kenny
Journal:  J Appl Clin Med Phys       Date:  2016-11-30       Impact factor: 2.102

4.  A Novel Algorithm in Radiation Dosimetry of Regular and Irregular Treatment Fields.

Authors:  Fatemeh Seif; Nayyer Mostafavi; Mohammadreza Bayatiyani; Hossein Taheri
Journal:  Adv Biomed Res       Date:  2019-07-25

5.  Effect of isocenter deviation on volume modulated arc therapy plan results and gamma passing rate in the treatment of cervical cancer.

Authors:  Yuan Yuan; Xiao-Bin Chang; Li-Juan Hu; Guo-Qing Wang; Yan-Jun Xue
Journal:  Transl Cancer Res       Date:  2021-10       Impact factor: 1.241

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.