Dosimetric validation of a redundant independent calculation software for VMAT fields.

F Mata Colodro1, A Serna Berná, V Puchades Puchades.   

Abstract

A redundant independent dosimetric calculation (RIDC) prior to treatment has become a basic part of the QA process for 3D conventional radiotherapy, and is strongly recommended in several international publications. On the other hand, the rapid growth in the number of intensity modulated treatments has led to a significant increase in the workflow associated with QA treatments. Diamond ("K&S Associates") is RIDC software which is capable of calculating VMAT (Volumetric Modulated Arc Therapy) fields. Modeling, validation and commissioning are necessary steps thereby making it a useful tool for VMAT QA. In this paper, a procedure for the validation of the calculation algorithm is demonstrated. A set 3D conventional field was verified in two ways: firstly, a comparison was made between Diamond calculations and experimental measures obtaining an average deviation of -0.1 ± 0.7%(1SD), and secondly, a comparison made between Diamond and the treatment planning system (TPS) Eclipse, obtaining an average deviation of 0.4 ± 0.8%(1SD). For both steps, a plastic slab phantom was used. VMAT validation was carried out by analyzing 59 VMAT plans in two ways: first, Diamond calculation versus experimental measurement with an average deviation of -0.2 ± 1.7%(1SD), and second, Diamond calculation versus TPS calculation with an average deviation of 0.0 ± 1.6%(1SD). In this phase a homogeneous cylindrical phantom was used. These results led us to consider this calculation algorithm validated for use in VMAT verifications.
Copyright © 2012 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Mesh:

Year:  2012        PMID: 22658464     DOI: 10.1016/j.ejmp.2012.05.003

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

1.  Validation and clinical implementation of commercial secondary check software with heterogeneity corrections.

Authors:  Vijayalakshmi Kuppusamy; Vivekanandan Nagarajan; Lavanya Murugan
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-16

2.  Radiotherapy pre-treatment dose validation: A second verification of monitor units (MU) with a commercial software.

Authors:  Iqbal Al Amri; Ramamoorthy Ravichandran; Somangili Satyamoorthi Sivakumar; Johnson Pichi Binukumar; Chirayathmanjiyil Antony Davis; Zakia Al Rahbi; Khalsa Al Shukeili; Fatima Al Kindi
Journal:  J Med Phys       Date:  2012-10

3.  Results of 1 year of clinical experience with independent dose calculation software for VMAT fields.

Authors:  Juan Fernando Mata Colodro; Alfredo Serna Berna; Vicente Puchades Puchades; David Ramos Amores; Miguel Alcaraz Baños
Journal:  J Med Phys       Date:  2014-10

4.  Dosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques.

Authors:  Francisco Clemente-Gutiérrez; Consuelo Pérez-Vara
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

5.  Monte Carlo modeling of the Elekta Versa HD and patient dose calculation with EGSnrc/BEAMnrc.

Authors:  Holly M Parenica Paschal; Christopher N Kabat; Pavlos Papaconstadopoulos; Neil A Kirby; Pamela A Myers; Timothy D Wagner; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2022-08-19       Impact factor: 2.243

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.