Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use.

Takahiro Komori1, Kazuo Eguchi, Satoshi Hoshide, Bryan Williams, Kazuomi Kario.   

Abstract

OBJECTIVE(S): We compared a convenient, cuffless, wrist-type ambulatory blood pressure monitoring (ABPM) device (BPro) and a standard arm-type ABPM monitor under ambulatory conditions.
METHODS: Fifty normotensive and prehypertensive volunteers who had no interarm differences in BP were enrolled. The wrist and arm monitors were attached to the left wrist and the right arm, respectively, and provided readouts at 15 and 30-min intervals, respectively. Ambulatory BP levels and the extent of agreement by intraclass correlation were evaluated. In 15 of the 50 participants, we also examined values at different arm positions: heart level, above the head, and hanging at the sides.
RESULTS: The awake mean systolic blood pressure (SBP) values (122±13 vs. 127±11 mmHg, P<0.01) were significantly lower with the wrist monitor than the arm monitor, and the mean sleep diastolic blood pressure (DBP) (71±8 vs. 64±8 mmHg, P<0.01) was significantly higher with the wrist monitor than the arm monitor. The intraclass correlation values between the monitors were 0.54 for 24 h SBP and 0.52 for awake SBP, considered to indicate a moderate agreement. The BP values in the arm-raised position were significantly higher for the wrist monitor than for the arm monitor (SBP: 129±14 vs. 108±14 mmHg, P<0.01; DBP: 83±13 vs. 64±11 mmHg, P<0.01). However, the SBP and DBP values in the other arm positions were similar between the monitors.
CONCLUSION: The wrist monitor showed fair agreement with the arm monitor in the ambulatory condition, and was stable irrespective of arm positions.

Entities:  

Mesh:

Year:  2013        PMID: 23263536     DOI: 10.1097/MBP.0b013e32835d124f

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  13 in total

1.  Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement.

Authors:  Jarkko Harju; Antti Vehkaoja; Pekka Kumpulainen; Stefano Campadello; Ville Lindroos; Arvi Yli-Hankala; Niku Oksala
Journal:  J Clin Monit Comput       Date:  2017-01-19       Impact factor: 2.502

Review 2.  Cuffless Blood Pressure Devices.

Authors:  Corey K Bradley; Daichi Shimbo; David Alexander Colburn; Daniel N Pugliese; Raj Padwal; Samuel K Sia; D Edmund Anstey
Journal:  Am J Hypertens       Date:  2022-05-10       Impact factor: 3.080

3.  Latest hypertension research to inform clinical practice in Asia.

Authors:  Kazuomi Kario; Masaki Mogi; Satoshi Hoshide
Journal:  Hypertens Res       Date:  2022-04-05       Impact factor: 5.528

4.  The first study comparing a wearable watch-type blood pressure monitor with a conventional ambulatory blood pressure monitor on in-office and out-of-office settings.

Authors:  Kazuomi Kario; Daichi Shimbo; Naoko Tomitani; Hiroshi Kanegae; Joseph E Schwartz; Bryan Williams
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-01-19       Impact factor: 3.738

5.  24-hour central aortic systolic pressure and 24-hour central pulse pressure are related to diabetic complications in type 1 diabetes - a cross-sectional study.

Authors:  Simone Theilade; Maria Lajer; Tine Willum Hansen; Christel Joergensen; Frederik Persson; Gudbjörg Andrésdottir; Henrik Reinhard; Stine Elkjær Nielsen; Peter Lacy; Bryan Williams; Peter Rossing
Journal:  Cardiovasc Diabetol       Date:  2013-08-27       Impact factor: 9.951

6.  Influence of ethnicity on acceptability of method of blood pressure monitoring: a cross-sectional study in primary care.

Authors:  Sally Wood; Sheila M Greenfield; M Sayeed Haque; Una Martin; Paramjit S Gill; Jonathan Mant; Mohammed A Mohammed; Gurdip Heer; Amanpreet Johal; Ramandeep Kaur; Claire Schwartz; Richard J McManus
Journal:  Br J Gen Pract       Date:  2016-06-06       Impact factor: 5.386

7.  Management of Hypertension in the Digital Era: Small Wearable Monitoring Devices for Remote Blood Pressure Monitoring.

Authors:  Kazuomi Kario
Journal:  Hypertension       Date:  2020-08-03       Impact factor: 10.190

8.  Comparison of Ambulatory Tonometric and Oscillometric Blood Pressure Monitoring in Hypertensive Patients.

Authors:  Bodil Gade Hornstrup; Jeppe Bakkestrøm Rosenbæk; Jesper Nørgaard Bech
Journal:  Integr Blood Press Control       Date:  2020-03-30

9.  Response to exercise-induced blood pressure elevation is blunted in wrist-cuff automated oscillometric measurement in healthy young college students.

Authors:  Tatsuya Sato; Nobutoshi Ichise; Yoshinori Terashima; Aoi Kato; Hiroya Yamazaki; Shunsuke Jimbo; Noritsugu Tohse
Journal:  Physiol Rep       Date:  2020-09

10.  Assessment of nocturnal hypertension by ambulatory blood pressure monitoring at the forearm in people with morbid obesity.

Authors:  Giacomo Pucci; Marco D'Abbondanza; Matteo Camilli; Valeria Bisogni; Fabio Anastasio; Vito Gandolfo; Riccardo Alcidi; Natasa Mojovic; Stefano Ministrini; Graziana Lupattelli; Gaetano Vaudo
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-08-13       Impact factor: 3.738

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.