The first study comparing a wearable watch-type blood pressure monitor with a conventional ambulatory blood pressure monitor on in-office and out-of-office settings.

Kazuomi Kario1, Daichi Shimbo2, Naoko Tomitani1, Hiroshi Kanegae1,3, Joseph E Schwartz4,5, Bryan Williams6.   

Abstract

Wearable blood pressure (BP) monitoring devices which measure BP levels accurately both in and out of the office are valuable for hypertension management using digital technology. The authors have conducted the first comparison study of BPs measured by a recently developed wrist-worn watch-type oscillometric BP monitoring (WBPM) device, the "HeartGuide," versus BPs measured by an ambulatory BP monitoring (ABPM) device, A&D TM-2441, in the office (total of 4 readings alternately measured in the sitting position) and outside the office (30-minutes interval measurements during daytime) in 50 consecutive patients (mean age 66.1 ± 10.8 years). The 2 BP monitoring devices were simultaneously worn on the same non-dominant arm throughout the monitoring period. The mean difference (±SD) in systolic BPs (average of 2 readings) between WBPM and ABPM was 0.8 ± 12.8 mm Hg (P = .564) in the office and 3.2 ± 17.0 mm Hg (P < .001) outside the office. The proportion of differences that were within ±10 mm Hg was 58.7% in the office and 47.2% outside the office. In a mixed-effects model analysis, the temporal trend in the difference between the out-of-office BPs measured by the two devices was not statistically significant. In conclusion, the difference between the WBPM and ABPM device was acceptable both in and out of the office.
© 2020 Wiley Periodicals, Inc.

Entities:  

Keywords:  HeartGuide; ambulatory blood pressure; comparison study; wearable watch-type wrist blood pressure monitoring

Mesh:

Year:  2020        PMID: 31955505      PMCID: PMC8029679          DOI: 10.1111/jch.13799

Source DB:  PubMed          Journal:  J Clin Hypertens (Greenwich)        ISSN: 1524-6175            Impact factor:   3.738


  19 in total

1.  Validation of the TM-2441 ambulatory blood pressure measurement device according to the ISO 81060-2: 2013 standard.

Authors:  Kazuomi Kario; Satoshi Hoshide; Kimiyo Saito; Keiko Sato; Haruna Hamasaki; Hiromi Suwa; Naoko Tomitani
Journal:  Blood Press Monit       Date:  2019-02       Impact factor: 1.444

2.  Nighttime home blood pressure and the risk of hypertensive target organ damage.

Authors:  Joji Ishikawa; Satoshi Hoshide; Kazuo Eguchi; Shizukiyo Ishikawa; Kazuyuki Shimada; Kazuomi Kario
Journal:  Hypertension       Date:  2012-08-14       Impact factor: 10.190

3.  2018 ESC/ESH Guidelines for the management of arterial hypertension.

Authors:  Bryan Williams; Giuseppe Mancia; Wilko Spiering; Enrico Agabiti Rosei; Michel Azizi; Michel Burnier; Denis L Clement; Antonio Coca; Giovanni de Simone; Anna Dominiczak; Thomas Kahan; Felix Mahfoud; Josep Redon; Luis Ruilope; Alberto Zanchetti; Mary Kerins; Sverre E Kjeldsen; Reinhold Kreutz; Stephane Laurent; Gregory Y H Lip; Richard McManus; Krzysztof Narkiewicz; Frank Ruschitzka; Roland E Schmieder; Evgeny Shlyakhto; Costas Tsioufis; Victor Aboyans; Ileana Desormais
Journal:  Eur Heart J       Date:  2018-09-01       Impact factor: 29.983

4.  Non-invasive monitoring of pulmonary artery pressure from timing information by EIT: experimental evaluation during induced hypoxia.

Authors:  Martin Proença; Fabian Braun; Josep Solà; Andy Adler; Mathieu Lemay; Jean-Philippe Thiran; Stefano F Rimoldi
Journal:  Physiol Meas       Date:  2016-05-23       Impact factor: 2.833

5.  Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method.

Authors:  Heiko Gesche; Detlef Grosskurth; Gert Küchler; Andreas Patzak
Journal:  Eur J Appl Physiol       Date:  2011-05-10       Impact factor: 3.078

6.  The further development of out-of-office BP monitoring: Japan's ImPACT Program Project's achievements, impact, and direction.

Authors:  Kazuomi Kario; Naoko Tomitani; Hiroshi Kanegae; Nobuhiko Yasui; Ryozo Nagai; Hiroshi Harada
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-02-27       Impact factor: 3.738

7.  Expert panel consensus recommendations for ambulatory blood pressure monitoring in Asia: The HOPE Asia Network.

Authors:  Kazuomi Kario; Jinho Shin; Chen-Huan Chen; Peera Buranakitjaroen; Yook-Chin Chia; Romeo Divinagracia; Jennifer Nailes; Satoshi Hoshide; Saulat Siddique; Jorge Sison; Arieska Ann Soenarta; Guru Prasad Sogunuru; Jam Chin Tay; Boon Wee Teo; Yuda Turana; Yuqing Zhang; Sungha Park; Huynh Van Minh; Ji-Guang Wang
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-09       Impact factor: 3.738

8.  Development and Validation of a Novel Cuff-Less Blood Pressure Monitoring Device.

Authors:  Naoki Watanabe; Yasuko K Bando; Taiji Kawachi; Hiroshi Yamakita; Kouki Futatsuyama; Yoshikazu Honda; Hisae Yasui; Kazuyuki Nishimura; Takahiro Kamihara; Takahiro Okumura; Hideki Ishii; Takahisa Kondo; Toyoaki Murohara
Journal:  JACC Basic Transl Sci       Date:  2017-12-25

9.  Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-02-25       Impact factor: 3.738

10.  Sleep Blood Pressure Self-Measured at Home as a Novel Determinant of Organ Damage: Japan Morning Surge Home Blood Pressure (J-HOP) Study.

Authors:  Kazuomi Kario; Satoshi Hoshide; Hajime Haimoto; Kayo Yamagiwa; Kiyoshi Uchiba; Shoichiro Nagasaka; Yuichiro Yano; Kazuo Eguchi; Yoshio Matsui; Motohiro Shimizu; Joji Ishikawa; Shizukiyo Ishikawa
Journal:  J Clin Hypertens (Greenwich)       Date:  2015-02-16       Impact factor: 3.738

View more
  19 in total

Review 1.  Remote Patient Monitoring in Heart Failure: Factors for Clinical Efficacy.

Authors:  Ankit Bhatia; Thomas M Maddox
Journal:  Int J Heart Fail       Date:  2020-11-30

2.  Enabling respiratory control after severe chronic tetraplegia: an exploratory case study.

Authors:  Parag Gad; Evgeniy Kreydin; Hui Zhong; V Reggie Edgerton
Journal:  J Neurophysiol       Date:  2020-08-05       Impact factor: 2.714

3.  How to assess the agreement between different methods of blood pressure measurement?

Authors:  Piotr Sobieraj; Zbigniew Gaciong
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-03-13       Impact factor: 3.738

4.  The first study comparing a wearable watch-type blood pressure monitor with a conventional ambulatory blood pressure monitor on in-office and out-of-office settings.

Authors:  Kazuomi Kario; Daichi Shimbo; Naoko Tomitani; Hiroshi Kanegae; Joseph E Schwartz; Bryan Williams
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-01-19       Impact factor: 3.738

5.  Afternoon blood pressure increase on home blood pressure measurement: A forgotten entity?

Authors:  Shiniya Taguchi; Kouichi Tamura
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-10-15       Impact factor: 3.738

Review 6.  Smart wearable devices in cardiovascular care: where we are and how to move forward.

Authors:  Karim Bayoumy; Mohammed Gaber; Abdallah Elshafeey; Omar Mhaimeed; Elizabeth H Dineen; Francoise A Marvel; Seth S Martin; Evan D Muse; Mintu P Turakhia; Khaldoun G Tarakji; Mohamed B Elshazly
Journal:  Nat Rev Cardiol       Date:  2021-03-04       Impact factor: 32.419

Review 7.  Telemedicine in the management of hypertension: Evolving technological platforms for blood pressure telemonitoring.

Authors:  Ji-Guang Wang; Yan Li; Yook-Chin Chia; Hao-Min Cheng; Huynh Van Minh; Saulat Siddique; Guru Prasad Sogunuru; Jam Chin Tay; Boon Wee Teo; Kelvin Tsoi; Yuda Turana; Tzung-Dau Wang; Yu-Qing Zhang; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2021-01-23       Impact factor: 3.738

8.  Management of Hypertension in the Digital Era: Small Wearable Monitoring Devices for Remote Blood Pressure Monitoring.

Authors:  Kazuomi Kario
Journal:  Hypertension       Date:  2020-08-03       Impact factor: 10.190

9.  Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device.

Authors:  Joon Ho Moon; Myung-Kyun Kang; Chang-Eun Choi; Jeonghee Min; Hae-Young Lee; Soo Lim
Journal:  Sci Rep       Date:  2020-11-04       Impact factor: 4.379

10.  Design and Fabrication of a New Wearable Pressure Sensor for Blood Pressure Monitoring.

Authors:  Marian Ion; Silviu Dinulescu; Bogdan Firtat; Mihaela Savin; Octavian N Ionescu; Carmen Moldovan
Journal:  Sensors (Basel)       Date:  2021-03-16       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.