| Literature DB >> 2325455 |
F L Drijber1, J B Finlay, T K Moroz, C H Rorabeck.
Abstract
An investigation was conducted to determine what improvements in the resistance to slippage could be obtained in selected interfaces (rod/clip torsional, cheek/bowl and cheek/clip) of the Hoffmann external fixator. The modification involved changing the standard wing-nut clamp for a bolt with a thread of 7 mm and a 1 mm pitch and placing an FAG 28-303 thrust-bearing (needle roller and cage assembly) between the bolt and the cheek. The results showed a significant improvement in the slippage values of all interfaces; increases of approximately six times were obtained at all torque values of the wing-nut clamp or fastener tested. Such improvements would markedly increase the reliability of external fixation systems and thus reduce the incidence of loss-of-reduction of fracture due to slippage of the universal joint.Entities:
Mesh:
Year: 1990 PMID: 2325455 DOI: 10.1007/bf02441671
Source DB: PubMed Journal: Med Biol Eng Comput ISSN: 0140-0118 Impact factor: 2.602