Hoffmann half-frame external fixation rigidity and its relationship to universal joint slippage.

F L Drijber1, J B Finlay.   

Abstract

Hoffmann half-frame external fixation device configurations often fail under minimal loads secondary to joint slippage. In these experiments improved universal joints that were developed in an earlier study were tested on Hoffmann half-frame assemblies. The rigidity of selected half-frame configurations was tested in four modes (axial compression, torsion, medial-lateral and anterior-posterior four-point bending). These results were compared to those of an earlier, similar study looking at the standard Hoffmann half-frame. No changes in overall rigidity were noted, but significant increases in yield loads and loads to frame failure were achieved. Such improvements will increase the reliability and usefulness of the Hoffmann device to the orthopaedic community.

Entities:  

Mesh:

Year:  1992        PMID: 1416289     DOI: 10.1007/bf02368172

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  20 in total

1.  Rigidity of immobilization of transverse fractures.

Authors:  O LINDAHL
Journal:  Acta Orthop Scand       Date:  1962

2.  Comparative study of fracture gap motion in external fixation.

Authors:  T Kristiansen; B Fleming; G Neale; S Reinecke; M H Pope
Journal:  Clin Biomech (Bristol, Avon)       Date:  1987-11       Impact factor: 2.063

3.  Fatigue properties of the hoffmann-vidal external fixation apparatus.

Authors:  R Shiba; E Y Chao; R Kasman
Journal:  Orthopedics       Date:  1984-03-01       Impact factor: 1.390

4.  External skeletal fixation: choosing a system based on biomechanical stability.

Authors:  T K Moroz; J B Finlay; C H Rorabeck; R B Bourne
Journal:  J Orthop Trauma       Date:  1988       Impact factor: 2.512

5.  Unilateral external fixation. Methods to increase and reduce frame stiffness.

Authors:  F Behrens; W Johnson
Journal:  Clin Orthop Relat Res       Date:  1989-04       Impact factor: 4.176

6.  An evaluation of the bending stiffness of various tibial fixation methods.

Authors:  M D Carter; J A Gilbert; L E Dahners
Journal:  Clin Orthop Relat Res       Date:  1987-11       Impact factor: 4.176

7.  Stability of ten configurations of the Hoffmann external-fixation frame.

Authors:  J B Finlay; T K Moroz; C H Rorabeck; J R Davey; R B Bourne
Journal:  J Bone Joint Surg Am       Date:  1987-06       Impact factor: 5.284

8.  The deformation of external fixation devices during loading.

Authors:  E O Karaharju; K Aalto
Journal:  Int Orthop       Date:  1983       Impact factor: 3.075

9.  Rigidity and stress analyses of external fracture fixation devices--a theoretical approach.

Authors:  E Y Chao; R A Kasman; K N An
Journal:  J Biomech       Date:  1982       Impact factor: 2.712

10.  The influence of induced micromovement upon the healing of experimental tibial fractures.

Authors:  A E Goodship; J Kenwright
Journal:  J Bone Joint Surg Br       Date:  1985-08
View more