Literature DB >> 21457385

Insulation defects of thin high-voltage ICD leads: an underestimated problem?

Damir Erkapic1, Gabor Z Duray, Tamas Bauernfeind, Salvatore De Rosa, Stefan H Hohnloser.   

Abstract

BACKGROUND: Long-term lead failure is a known complication of ICD therapy. The precise incidence and sequelae of insulation defects at the tricuspid level, however, are not well characterized.
OBJECTIVE: This study determined the risk of lead failure, with particular emphasis on insulation defects at the level of the tricuspid valve, in a large series of consecutive ICD recipients.
METHODS: Data from 357 consecutive patients, who had received transvenous 7 and 8 French ICD-leads (St. Jude Medical, Riata family) and were followed at our center, formed the basis of this study.
RESULTS: During a mean follow-up of 42 ± 24 months, 30 of 357 (8%) patients required surgical intervention due to lead failure. For overall lead defects, lead access via the subclavian vein and subpectoral device placement were independent predictors of overall lead failure (OR 3.47, 95% CI 1.38-8.72, P = 0.013 and OR 3.83, 95% CI 1.77-8.27, P = 0.001, respectively). Lead insulation defects at the level of the tricuspid valve accounted for 20% of all lead failures. Diagnosis of this specific insulation defect could only be established by fluoroscopy, while electrical parameters were within normal limits in all of these patients. On univariate but not on multivariate analysis the presence of nonischemic cardiomyopathy was a predictor of this lead complication (OR 8.2, CI 1.5-46.1, P = 0.02).
CONCLUSION: Insulation defects of 7 and 8 French ICD leads at the tricuspid level represent an important complication of device therapy. Even moderate changes in lead impedance within the normal limits at follow-up should prompt careful fluoroscopic evaluation to avoid spurious shocks.
© 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21457385     DOI: 10.1111/j.1540-8167.2011.02055.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  10 in total

1.  Axillary sub-pectoral pulse generator pocket for lowering defibrillation threshold.

Authors:  Ernest W Lau
Journal:  J Interv Card Electrophysiol       Date:  2012-02-23       Impact factor: 1.900

2.  [Lead survival and complications (except infections). Are we doing better nowadays?].

Authors:  Martin Seifert; Michael Neuss; Maren Schöpp; Cornel Koban; Christian Butter
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-08-06

Review 3.  [ICD leads].

Authors:  Carsten W Israel; Mohamed Karim Sheta
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06

4.  Prospective evaluation of cinefluoroscopy and chest radiography for Riata lead defects: implications for future lead screening.

Authors:  Peem Lorvidhaya; Ivan Mendoza; Sharmila Sehli; Michael K Atalay; Michael H Kim
Journal:  J Interv Card Electrophysiol       Date:  2013-09-08       Impact factor: 1.900

5.  What did we learn from Riata™?

Authors:  Avi Fischer; C Jenney
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06-09

6.  Increasing lead burden correlates with externalized cables during systematic fluoroscopic screening of Riata leads.

Authors:  Liza P Moorman; J Randall Moorman; John P DiMarco; Rohit Malhotra; Andrew Darby; Kenneth Bilchick; John D Ferguson; J Michael Mangrum; Sandeep Kamath; Pamela K Mason
Journal:  J Interv Card Electrophysiol       Date:  2012-12-20       Impact factor: 1.900

7.  Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

Authors:  S D A Valk; D A M J Theuns; L Jordaens
Journal:  Neth Heart J       Date:  2013-03       Impact factor: 2.380

8.  [Recommendations of the Working Group of Arrhythmias of the German Society of Cardiology on the approach to patients with Riata® and Riata ST® leads (St. Jude Medical). Nucleus of the Working Group of Arrhythmias of the German Society of Cardiology].

Authors:  C W Israel; D Bänsch; D Böcker; C Butter; J Chun; I Deisenhofer; L Eckardt; J-C Geller; T Hanke; T Klingenheben; C Piorkowski; B Schumacher
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2012-06

9.  Introduction of new industry standards for cardiac implantable electronic devices: balancing benefits and unexpected risks.

Authors:  Christian Sticherling; Haran Burri
Journal:  Europace       Date:  2012-03-01       Impact factor: 5.214

10.  A case of riata® dual coil defibrillator lead failure in a patient with ventricular fibrillation.

Authors:  Kyu Choi; Jung Hee Kim; Hyo Jin Kim; Se-Ok Lee; Eun-Young Jang; June Soo Kim
Journal:  Korean Circ J       Date:  2013-05-31       Impact factor: 3.243

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.