An ICD lead with failure of outer insulation goes undetected by regular measurements.

Chi-Woh Chan1, Chung-Seung Chiang.   

Abstract

We report a patient with ischemic cardiomyopathy status post implantable cardioverter-defibrillator (ICD) implantation in June 2009. Outer insulation failure of her Riata ST ventricular ICD lead goes undetected by routine interrogation and provocative tests as recommended from its manufacturer.(1) Another domain of assessment, fluoroscopy, to facilitate early detection of structural abnormality in this family of ICD lead is discussed. ©2011, The Authors. Journal compilation ©2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21745226     DOI: 10.1111/j.1540-8159.2011.03164.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

1.  Prospective evaluation of cinefluoroscopy and chest radiography for Riata lead defects: implications for future lead screening.

Authors:  Peem Lorvidhaya; Ivan Mendoza; Sharmila Sehli; Michael K Atalay; Michael H Kim
Journal:  J Interv Card Electrophysiol       Date:  2013-09-08       Impact factor: 1.900

2.  Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

Authors:  S D A Valk; D A M J Theuns; L Jordaens
Journal:  Neth Heart J       Date:  2013-03       Impact factor: 2.380

3.  [Recommendations of the Working Group of Arrhythmias of the German Society of Cardiology on the approach to patients with Riata® and Riata ST® leads (St. Jude Medical). Nucleus of the Working Group of Arrhythmias of the German Society of Cardiology].

Authors:  C W Israel; D Bänsch; D Böcker; C Butter; J Chun; I Deisenhofer; L Eckardt; J-C Geller; T Hanke; T Klingenheben; C Piorkowski; B Schumacher
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2012-06

4.  Contribution of ethylenetetrafluoroethylene (ETFE) insulation to the electrical performance of Riata® silicone leads having externalized conductors.

Authors:  Avi Fischer; Russell Klehn
Journal:  J Interv Card Electrophysiol       Date:  2013-03-26       Impact factor: 1.900

5.  A case of riata® dual coil defibrillator lead failure in a patient with ventricular fibrillation.

Authors:  Kyu Choi; Jung Hee Kim; Hyo Jin Kim; Se-Ok Lee; Eun-Young Jang; June Soo Kim
Journal:  Korean Circ J       Date:  2013-05-31       Impact factor: 3.243

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.