Comparison of patient-specific instruments with standard surgical instruments in determining glenoid component position: a randomized prospective clinical trial.

Michael D Hendel1, Jason A Bryan, Wael K Barsoum, Eric J Rodriguez, John J Brems, Peter J Evans, Joseph P Iannotti.   

Abstract

BACKGROUND: Glenoid component malposition for anatomic shoulder replacement may result in complications. The purpose of this study was to define the efficacy of a new surgical method to place the glenoid component.
METHODS: Thirty-one patients were randomized for glenoid component placement with use of either novel three-dimensional computed tomographic scan planning software combined with patient-specific instrumentation (the glenoid positioning system group), or conventional computed tomographic scan, preoperative planning, and surgical technique, utilizing instruments provided by the implant manufacturer (the standard surgical group). The desired position of the component was determined preoperatively. Postoperatively, a computed tomographic scan was used to define and compare the actual implant location with the preoperative plan.
RESULTS: In the standard surgical group, the average preoperative glenoid retroversion was -11.3° (range, -39° to 17°). In the glenoid positioning system group, the average glenoid retroversion was -14.8° (range, -27° to 7°). When the standard surgical group was compared with the glenoid positioning system group, patient-specific instrumentation technology significantly decreased (p < 0.05) the average deviation of implant position for inclination and medial-lateral offset. Overall, the average deviation in version was 6.9° in the standard surgical group and 4.3° in the glenoid positioning system group. The average deviation in inclination was 11.6° in the standard surgical group and 2.9° in the glenoid positioning system group. The greatest benefit of patient-specific instrumentation was observed in patients with retroversion in excess of 16°; the average deviation was 10° in the standard surgical group and 1.2° in the glenoid positioning system group (p < 0.001). Preoperative planning and patient-specific instrumentation use resulted in a significant improvement in the selection and use of the optimal type of implant and a significant reduction in the frequency of malpositioned glenoid implants.
CONCLUSIONS: Novel three-dimensional preoperative planning, coupled with patient and implant-specific instrumentation, allows the surgeon to better define the preoperative pathology, select the optimal implant design and location, and then accurately execute the plan at the time of surgery.

Entities:  

Mesh:

Year:  2012        PMID: 23224387     DOI: 10.2106/JBJS.K.01209

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  45 in total

1.  The arthritic glenoid: anatomy and arthroplasty designs.

Authors:  Nikolas K Knowles; Louis M Ferreira; George S Athwal
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

2.  Surgical management of the biconcave (B2) glenoid.

Authors:  Kenneth W Donohue; Eric T Ricchetti; Joseph P Iannotti
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

3.  Management of complications after revision shoulder arthroplasty.

Authors:  Hithem Rahmi; Andrew Jawa
Journal:  Curr Rev Musculoskelet Med       Date:  2015-03

Review 4.  [Shoulder prosthesis replacement options : New implants, treatment algorithms and clinical results].

Authors:  D Seybold; T A Schildhauer; J Geßmann
Journal:  Orthopade       Date:  2018-05       Impact factor: 1.087

5.  Glenoid bone loss in primary and revision shoulder arthroplasty.

Authors:  Amar Malhas; Abbas Rashid; Dave Copas; Steve Bale; Ian Trail
Journal:  Shoulder Elbow       Date:  2016-05-06

6.  Is premorbid glenoid anatomy altered in patients with glenohumeral osteoarthritis?

Authors:  Eric T Ricchetti; Michael D Hendel; David N Collins; Joseph P Iannotti
Journal:  Clin Orthop Relat Res       Date:  2013-05-18       Impact factor: 4.176

7.  Planning software and patient-specific instruments in shoulder arthroplasty.

Authors:  James D Wylie; Robert Z Tashjian
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

8.  Proper benefit of a three dimensional pre-operative planning software for glenoid component positioning in total shoulder arthroplasty.

Authors:  Adrien Jacquot; Marc-Olivier Gauci; Jean Chaoui; Mohammed Baba; Pierric Deransart; Pascal Boileau; Daniel Mole; Gilles Walch
Journal:  Int Orthop       Date:  2018-07-02       Impact factor: 3.075

Review 9.  Glenoid component positioning and guidance techniques in anatomic and reverse total shoulder arthroplasty: A systematic review and meta-analysis.

Authors:  David M Burns; Tym Frank; Cari M Whyne; Patrick Dg Henry
Journal:  Shoulder Elbow       Date:  2018-10-26

10.  The fulcrum axis: an accurate measure of glenoid version on radiographs and computed tomography.

Authors:  Jennifer Mutch; Martin Sidler; Claudia Sidler-Maier; Terry Axelrod; Diane Nam
Journal:  Shoulder Elbow       Date:  2017-08-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.