Literature DB >> 23223668

Shock or no shock - a question of philosophy or should intraoperative implantable cardioverter defibrillator testing be recommended?

Andreas Keyser1, Michael K Hilker, Sebastian Schmidt, Christian von Bary, Wolfgang Zink, Michael Ried, Christof Schmid, Claudius Diez.   

Abstract

OBJECTIVES: Implantation of implantable cardioverter defibrillators (ICDs) in patients with a high risk for life-threatening ventricular arrhythmias is a standard therapy. The development of new ICD leads, shock algorithms, high-energy defibrillators and rapid energy supply has improved the devices. Nevertheless, the discussion regarding 'shock or no shock' to test the system intraoperatively has not silenced yet.
METHODS: In this study, all 718 patients (60.0 ± 14.2 years old, 570 male) who were treated with a first ICD at our institution since 2005 were analysed. The indication for implantation was primarily prophylactic in 511 patients (71.3%). Underlying diseases included ischaemic cardiomyopathy (358 patients, 50%), dilated cardiomyopathy (270 patients, 37.7%) and others (12.3%). Mean ejection fraction was 27.4 ± 11.8%. Intraoperative ventricular fibrillation was induced with a T-wave shock or burst stimulation. The primary end-point was failing the initial intraoperative testing.
RESULTS: During the initial testing, 28 patients (3.9%) had a defibrillation threshold (DFT) >21 J. The mean age of these patients was 51 ± 14 years, ranging from 22 to 71 years, 20 were male, and the ejection fraction was 23.8 ± 11.8%. The indication for ICD implantation was prophylactic in 13 patients. Twenty-one of the 28 patients suffered from dilated cardiomyopathy, whereas seven patients had ischaemic cardiomyopathy. Twenty-four ICDs were implanted on the left side and four on the right side. None of the patients had been treated with amiodarone at the time of implantation. All patients achieved a sufficient DFT ≤ 21 J by changing the ICD leads, device repositioning and/or optimizing the shock configuration.
CONCLUSIONS: The standard of care intraoperative ICD testing remains necessary.

Entities:  

Mesh:

Year:  2012        PMID: 23223668      PMCID: PMC3568803          DOI: 10.1093/icvts/ivs479

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  23 in total

1.  Effects of repeated electrical defibrillations on cardiac troponin I levels.

Authors:  J A Joglar; D J Kessler; P J Welch; J H Keffer; M E Jessen; M H Hamdan; R L Page
Journal:  Am J Cardiol       Date:  1999-01-15       Impact factor: 2.778

2.  The top 10 reasons to avoid defibrillation threshold testing during ICD implantation.

Authors:  Sami Viskin; Raphael Rosso
Journal:  Heart Rhythm       Date:  2008-01-11       Impact factor: 6.343

3.  The effect of induction method on defibrillation threshold and ventricular fibrillation cycle length.

Authors:  Endre Zima; Mihály Gergely; Pál Soós; László Alajos Gellér; Attila Nemes; György Acsády; Béla Merkely
Journal:  J Cardiovasc Electrophysiol       Date:  2006-04

4.  Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients.

Authors:  Peter Alter; Stefan Waldhans; Eveline Plachta; Rainer Moosdorf; Wolfram Grimm
Journal:  Pacing Clin Electrophysiol       Date:  2005-09       Impact factor: 1.976

5.  Differences in the characteristics of induced and spontaneous episodes of ventricular fibrillation.

Authors:  Nigel A Lever; Emma G Newall; Peter D Larsen
Journal:  Europace       Date:  2007-09-14       Impact factor: 5.214

6.  Inductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations).

Authors:  John D Day; Rahul N Doshi; Peter Belott; Ulrika Birgersdotter-Green; Mahnaz Behboodikhah; Peter Ott; Kathryn A Glatter; Serge Tobias; Howard Frumin; Byron K Lee; John Merillat; Isaac Wiener; Samuel Wang; Harlan Grogin; Sung Chun; Rob Patrawalla; Brian Crandall; Jeffrey S Osborn; J Peter Weiss; Donald L Lappe; Stacey Neuman
Journal:  Circulation       Date:  2007-04-30       Impact factor: 29.690

7.  No benefit from defibrillation threshold testing in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial).

Authors:  Joseph A Blatt; Jeanne E Poole; George W Johnson; David J Callans; Merritt H Raitt; Ramakota K Reddy; Francis E Marchlinski; Raymond Yee; Thomas Guarnieri; Mario Talajic; David J Wilber; Jill Anderson; Kiyon Chung; Wai Shun Wong; Daniel B Mark; Kerry L Lee; Gust H Bardy
Journal:  J Am Coll Cardiol       Date:  2008-08-12       Impact factor: 24.094

8.  Indications for predismissal testing with arrhythmia-induction in patients receiving an implantable cardioverter defibrillator.

Authors:  G Christ; R Becker; F Voss; K Kelemen; J Senges-Becker; M Hauck; W Schoels; I Bald; H A Katus; A Bauer
Journal:  Clin Res Cardiol       Date:  2007-06-27       Impact factor: 5.460

Review 9.  ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure 2008: the Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2008 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association of the ESC (HFA) and endorsed by the European Society of Intensive Care Medicine (ESICM).

Authors:  Kenneth Dickstein; Alain Cohen-Solal; Gerasimos Filippatos; John J V McMurray; Piotr Ponikowski; Philip Alexander Poole-Wilson; Anna Strömberg; Dirk J van Veldhuisen; Dan Atar; Arno W Hoes; Andre Keren; Alexandre Mebazaa; Markku Nieminen; Silvia Giuliana Priori; Karl Swedberg
Journal:  Eur Heart J       Date:  2008-09-17       Impact factor: 29.983

10.  Complications associated with defibrillation threshold testing: the Canadian experience.

Authors:  David Birnie; Stanley Tung; Christopher Simpson; Eugene Crystal; Derek Exner; Felix-Alejandro Ayala Paredes; Andrew Krahn; Ratika Parkash; Yaariv Khaykin; Francois Philippon; Peter Guerra; Shane Kimber; Douglas Cameron; Jeffrey S Healey
Journal:  Heart Rhythm       Date:  2007-11-28       Impact factor: 6.343

View more
  2 in total

1.  eComment. Intraoperative implantable cardioverter defibrillator testing.

Authors:  Murat Unlu; Sevket Balta; Mustafa Cakar; Sait Demirkol
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-03

2.  eComment. Shock or no shock: individual option for some patients.

Authors:  Leo Bokeria; Andrey Filatov; Alexey Kovalev
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-03
  2 in total