Literature DB >> 18687249

No benefit from defibrillation threshold testing in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial).

Joseph A Blatt1, Jeanne E Poole, George W Johnson, David J Callans, Merritt H Raitt, Ramakota K Reddy, Francis E Marchlinski, Raymond Yee, Thomas Guarnieri, Mario Talajic, David J Wilber, Jill Anderson, Kiyon Chung, Wai Shun Wong, Daniel B Mark, Kerry L Lee, Gust H Bardy.   

Abstract

OBJECTIVES: This study investigated whether defibrillation threshold (DFT) testing during implantable cardioverter-defibrillator (ICD) implantation predicts clinical outcomes.
BACKGROUND: Defibrillation testing is often performed during insertion of ICDs to confirm shock efficacy. There are no prospective data to suggest that this procedure improves outcomes when modern ICDs are implanted for primary prevention of sudden death.
METHODS: The analysis included the 811 patients who were randomized to the ICD arm of the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial) and had the device implanted. The DFT testing protocol in SCD-HeFT was designed to limit shock testing in a primary prevention heart failure population.
RESULTS: Baseline DFT data were available for 717 patients (88.4%). All 717 patients had a DFT of < or =30 J, the maximum output of the device in this study. The DFT was < or =20 J in 97.8% of patients. There was no survival difference between patients with a lower DFT (< or =10 J, n = 547) and a higher DFT (>10 J, n = 170) (p = 0.41). First shock efficacy was 83.0% for the first clinical ventricular tachyarrhythmia event; there were no differences in shock efficacies when the cohort was subdivided by baseline DFT.
CONCLUSIONS: Low baseline DFTs were obtained in patients with stable, optimally treated heart failure during ICD implantation for primary prevention of sudden death. First shock efficacy for ventricular tachyarrhythmias was high regardless of baseline DFT testing results. Baseline DFT testing did not predict long-term mortality or shock efficacy in this study.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18687249     DOI: 10.1016/j.jacc.2008.04.051

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  30 in total

1.  [Experienced anesthetists indispensable. Anesthesia management for implantation of cardioverter/defibrillators].

Authors:  P Tassani-Prell; G Juchem
Journal:  Anaesthesist       Date:  2010-06       Impact factor: 1.041

2.  [Is intraoperative ICD-testing still necessary?].

Authors:  C Mewis; H-R Neuberger; A Buob
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2010-06

Review 3.  Complications of defibrillation testing. A review of the literature.

Authors:  Bishoy Deif; Jeff S Healey
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-11-30

4.  Shock or no shock - a question of philosophy or should intraoperative implantable cardioverter defibrillator testing be recommended?

Authors:  Andreas Keyser; Michael K Hilker; Sebastian Schmidt; Christian von Bary; Wolfgang Zink; Michael Ried; Christof Schmid; Claudius Diez
Journal:  Interact Cardiovasc Thorac Surg       Date:  2012-12-07

5.  Effect of defibrillation threshold testing-induced ventricular fibrillation on renal function.

Authors:  John H Shin; Chotikorn Khunnawat; Jose Baez-Escudero; Bradley P Knight; John F Beshai
Journal:  J Interv Card Electrophysiol       Date:  2013-10-11       Impact factor: 1.900

6.  Evaluation of defibrillation safety margin in modern implantable cardioverter defibrillators after administration of amiodarone.

Authors:  Julia Köbe; Florian Reinke; Dirk G Dechering; Günter Breithardt; Lars Eckardt
Journal:  Clin Res Cardiol       Date:  2011-11-05       Impact factor: 5.460

7.  Prophylactic implantable defibrillators in dilated cardiomyopathy.

Authors:  W Grimm
Journal:  Herz       Date:  2012-12       Impact factor: 1.443

Review 8.  The Saga of Defibrillation Testing: When Less Is More.

Authors:  Marye J Gleva; Melissa Robinson; Jeanne Poole
Journal:  Curr Cardiol Rep       Date:  2018-05-05       Impact factor: 2.931

9.  Incidence of very high defibrillation thresholds (DFT) and efficacy of subcutaneous (SQ) array insertion during implantable cardioverter defibrillator (ICD) implantation.

Authors:  Atul Verma; Alexander J Kaplan; Bradley Sarak; Richard Oosthuizen; Marianne Beardsall; Jan Higgenbottam; Zaev Wulffhart; Yaariv Khaykin
Journal:  J Interv Card Electrophysiol       Date:  2010-09-24       Impact factor: 1.900

10.  [Is the determination of the defibrillation threshold in patients with an implantable cardioverter-defibrillator still required?].

Authors:  M Licka; L Jahn; K Kelemen; F Voss; K Trappe; R Becker; O Bikou; M Hauck; M Koch; H A Katus; A Bauer
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2011-11-13
View more