Interfraction and intrafraction performance of the Gamma Knife Extend system for patient positioning and immobilization.

David Schlesinger1, Zhiyuan Xu, Frances Taylor, Chun-Po Yen, Jason Sheehan.   

Abstract

OBJECT: The Extend system for the Gamma Knife Perfexion makes possible multifractional Gamma Knife treatments. The Extend system consists of a vacuum-monitored immobilization frame and a positioning measurement system used to determine the location of the patient's head within the frame at the time of simulation imaging and before each treatment fraction. The measurement system consists of a repositioning check tool (RCT), which attaches to the Extend frame, and associated digital measuring gauges. The purpose of this study is to evaluate the performance of the Extend system for patient repositioning before each treatment session (fraction) and patient immobilization between (interfraction) and during (intrafraction) each session in the first 10 patients (36 fractional treatments) treated at the University of Virginia.
METHODS: The RCT was used to acquire a set of reference measurements for each patient position at the time of CT simulation. Repositioning measurements were acquired before each fraction, and the patient position was adjusted until the residual radial difference from the reference position measurements was less than 1 mm. After treatment, patient position measurements were acquired, and the difference between those measurements and the ones obtained for patient position before the fraction was calculated as a measure of immobilization capability. Analysis of patient setup and immobilization performance included calculation of the group mean, standard deviation (SD), and distribution of systematic (components affecting all fractions) and random (per fraction) uncertainty components.
RESULTS: Across all patients and fractions, the mean radial setup difference from the reference measurements was 0.64 mm, with an SD of 0.24 mm. The distribution of systematic uncertainty (Σ) was 0.17 mm, and the distribution of random uncertainty (σ) was 0.16 mm. The root mean square (RMS) differences for each plate of the RCT were as follows: right = 0.35 mm; left = 0.41 mm; superior = 0.28 mm; and anterior = 0.20 mm. The mean intrafractional positional difference across all treatments was 0.47 mm, with an SD of 0.30 mm. The distribution of systematic uncertainty was 0.18 mm, and the distribution of random uncertainty was 0.22 mm. The RMS differences for each plate of the RCT were 0.24 mm for the right plate, 0.22 mm for the left plate, 0.24 mm for the superior plate, and 0.34 mm for the anterior plate. Data from 1 fraction were excluded from the analysis because the vacuum-monitoring interlock detected patient motion, which in turn required repositioning in the middle of the fraction.
CONCLUSIONS: The Extend system can be used to reposition and immobilize patients in a radiosurgical setting. However, care should be taken to acquire measurements that can implicitly account for rotations of the patient's head. Further work is required to determine the sensitivity of the vacuum interlock to detect patient motion.

Entities:  

Mesh:

Year:  2012        PMID: 23205813     DOI: 10.3171/2012.6.GKS12989

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  12 in total

Review 1.  Cranial stereotactic radiosurgery: current status of the initial paradigm shifter.

Authors:  Jason P Sheehan; Chun-Po Yen; Cheng-Chia Lee; Jay S Loeffler
Journal:  J Clin Oncol       Date:  2014-08-11       Impact factor: 44.544

Review 2.  Unyielding progress: recent advances in the treatment of central nervous system neoplasms with radiosurgery and radiation therapy.

Authors:  Dale Ding; Chun-Po Yen; Robert M Starke; Cheng-Chia Lee; Jason P Sheehan
Journal:  J Neurooncol       Date:  2014-08-14       Impact factor: 4.130

3.  Evaluation of a magnetic resonance guided linear accelerator for stereotactic radiosurgery treatment.

Authors:  Ning Wen; Joshua Kim; Anthony Doemer; Carri Glide-Hurst; Indrin J Chetty; Chang Liu; Eric Laugeman; Ilma Xhaferllari; Akila Kumarasiri; James Victoria; Maria Bellon; Steve Kalkanis; M Salim Siddiqui; Benjamin Movsas
Journal:  Radiother Oncol       Date:  2018-05-25       Impact factor: 6.280

4.  Quantifying the trigger level of the vacuum surveillance system of the Gamma-Knife eXtend™ positioning system and evaluating the potential impact on dose delivery.

Authors:  Béatrice Reiner; Peter Bownes; David L Buckley; David I Thwaites
Journal:  J Radiosurg SBRT       Date:  2016

5.  Validity of the use of nose tip motion as a surrogate for intracranial motion in mask-fixated frameless Gamma Knife® Icon™ therapy.

Authors:  Gavin Wright; Natalie Harrold; Paul Hatfield; Peter Bownes
Journal:  J Radiosurg SBRT       Date:  2017

6.  Whole-procedural radiological accuracy for delivering multi-session gamma knife radiosurgery with a relocatable frame system.

Authors:  Lijun Ma; Dilini Pinnaduwage; Michael McDermott; Penny K Sneed
Journal:  Technol Cancer Res Treat       Date:  2013-08-31

7.  Simulational study of a dosimetric comparison between a Gamma Knife treatment plan and an intensity-modulated radiotherapy plan for skull base tumors.

Authors:  Hisato Nakazawa; Yoshimasa Mori; Masataka Komori; Takahiko Tsugawa; Yuta Shibamoto; Tatsuya Kobayashi; Chisa Hashizume; Yukio Uchiyama; Masahiro Hagiwara
Journal:  J Radiat Res       Date:  2013-12-17       Impact factor: 2.724

8.  Preliminary Evaluation of a Novel Thermoplastic Mask System with Intra-fraction Motion Monitoring for Future Use with Image-Guided Gamma Knife.

Authors:  Winnie Li; Gregory Bootsma; Oscar Von Schultz; Per Carlsson; Normand Laperriere; Barbara-Ann Millar; David Jaffray; Caroline Chung
Journal:  Cureus       Date:  2016-03-13

Review 9.  Re-irradiation of Recurrent Pineal Germ Cell Tumors with Radiosurgery: Report of Two Cases and Review of Literature.

Authors:  Kenneth Wong; Anthony B Opimo; Arthur J Olch; Sean All; Jonathan F Waxer; Desirae Clark; Justine Cheng; Alisha Chlebik; Anat Erdreich-Epstein; Mark D Krieger; Benita Tamrazi; Girish Dhall; Jonathan L Finlay; Eric L Chang
Journal:  Cureus       Date:  2016-04-25

10.  Clinical Positioning Accuracy for Multisession Stereotactic Radiotherapy With the Gamma Knife Perfexion.

Authors:  Wade P Smith; Lori A Young; Mark H Phillips; Michael Cheung; Lia M Halasz; Jason K Rockhill
Journal:  Technol Cancer Res Treat       Date:  2017-05-17